Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K162010

    Validate with FDA (Live)

    Device Name
    HPM-6000U
    Date Cleared
    2016-12-13

    (146 days)

    Product Code
    Regulation Number
    876.5320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HPM-6000U is intended to provide entirely non-invasive electromagnetic stimulation of pelvic floor musculature for the purpose of rehabilitation of weak pelvic muscles and restoration of neuromuscular control for the treatment of urinary incontinence in women.

    Device Description

    The HPM-6000U is a non-invasive therapeutic device. The device produces electromagnetic field that interacts with the tissues of the human body. The electromagnetic field delivered in the muscular or neuronal tissue area is triggering the stimulation and tonisation.

    The HPM-6000U is equipped with a color touch screen with wide view angle that significantly facilitates the use of the device. The on-screen information guides the user step-by-step through the entire therapy procedure. The therapeutic parameters are easily set using the touch screen, buttons and knob on the device. During the therapy the device keeps information about the applied therapy type, remaining therapy time and main therapy parameters on the screen.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for the HPM-6000U device, which is an electromagnetic stimulator for treating urinary incontinence in women. The document explicitly states that the substantial equivalence determination for the HPM-6000U is NOT based on clinical testing. Instead, its safety and efficacy were demonstrated by comparing its technical characteristics with a predicate device, the Neotonus Model 1000 Muscle Stimulator System (K973096).

    Therefore, there is no study described in this document that proves the device meets specific acceptance criteria related to its clinical performance (e.g., sensitivity, specificity, accuracy) using a test set against a ground truth.

    However, the document does contain information about non-clinical testing for compliance with safety standards and a comparison table of technical specifications between the HPM-6000U and the predicate device, which serves as the basis for its substantial equivalence claim.

    Given the information provided in the document:

    1. A table of acceptance criteria and the reported device performance:

    Since no clinical efficacy study with specific performance acceptance criteria is detailed, the "acceptance criteria" here relate to technical specifications and safety standards compliance.

    Acceptance Criteria (Technical/Safety Standard)Reported Device Performance (HPM-6000U)
    Technical Specifications (Compared to Predicate Device)
    Primary Function: Stimulation of pelvic floor musculatureStimulation of pelvic floor musculature
    Principle of Action: Initiating action potential of nerves results in muscle contractionInitiating action potential of nerves results in muscle contraction
    Type of Energy: Magnetic fieldMagnetic field
    Energy Source100 - 240 V AC, 50 - 60 Hz, max 14 A
    Type of Coil: Single magnetic coilSingle magnetic coil
    Number of Magnetic Coils in the Applicator: 11
    Type of Applicator: ChairChair
    Number of Applicators: 11
    Patient Position during Therapy: Sitting positionSitting position
    Position of Coil: Center of applicator seatCenter of applicator seat
    Type of Operation: ContinuousContinuous
    Magnetic Field Intensity0.7 - 2.5 T
    Pulse Repetition Rate1 - 150 Hz
    Step of Frequency Setting: 1 Hz1 Hz
    Pulse Duration280 μs (± 20%)
    Pulse Amplitude0 – 100 %
    Shape of Stimulation Pulse: Sine, biphasicSine, biphasic
    Therapy Time: 30 min30 min
    Operating Temperature+10 to +30 °C (50 - 86 °F)
    Interface: Touch-screenTouch-screen
    Firmware controlled: YesYes
    Environmental Specifications: For indoor use onlyFor indoor use only
    Clinical Use: Prescription UsePrescription Use
    External Exchangeable Fuse: YesYes
    Main Unit Dimensions: 500x970x580 mm (20x38x23 in)500x970x580 mm (20x38x23 in)
    Applicator Dimensions: 730x730x730 mm (29x29x29 in)730x730x730 mm (29x29x29 in)
    System Weight46 kg (101 lb)
    Position: Vertical - On castorsVertical - On castors
    Electrical ProtectionClass II, BF
    Compliance with Safety Standards
    ISO 14971 (Risk Management)Compliant
    IEC 62304 (Software Life Cycle Processes)Compliant
    IEC 60601-1 (Basic Safety and Essential Performance)Compliant
    IEC 60601-1-2 (EMC Requirements and Tests)Compliant
    IEC 60601-1-6 (Usability)Compliant
    ISO 10993-1 (Biological Evaluation)Compliant
    ISO 10993-5 (In vitro Cytotoxicity)Compliant
    ISO 10993-10 (Irritation and Skin Sensitization)Compliant

    2. Sample size used for the test set and the data provenance:

    • Not applicable. No clinical test set data is provided as clinical testing was not performed for substantial equivalence. The comparison is based on the technical specifications of the device and its predicate.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. No clinical test set or ground truth established by experts is described.

    4. Adjudication method for the test set:

    • Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a therapeutic device, not an AI-assisted diagnostic tool for human readers. No MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. The device is a physical therapeutic device, not an algorithm. Its performance is based on its physical/electromagnetic output, not an algorithm's classification or detection capabilities.

    7. The type of ground truth used:

    • Not applicable. For the substantial equivalence, the "ground truth" is implied to be the established safety and effectiveness of the legally marketed predicate device, as demonstrated through its technical specifications and intended use.

    8. The sample size for the training set:

    • Not applicable. No training set for an algorithm is mentioned as this is a physical medical device.

    9. How the ground truth for the training set was established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1