Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K161979

    Validate with FDA (Live)

    Date Cleared
    2017-03-22

    (247 days)

    Product Code
    Regulation Number
    876.5980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The proposed device is indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syringes are intended to be used in clinical or home care setting by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups.

    Device Description

    The proposed device is a disposable enteral feeding syringe provided in a variety of sizes from 1ml~60ml. This device incorporates a female ENFit connection to an enteral access device with male ENFit connector.

    The propsoed syringe is sterilized by Ethylene Oxide Gas to achieve a SAL of 10 and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of five years.

    AI/ML Overview

    This document describes the non-clinical testing performed to demonstrate that the ENFit Enteral Syringe meets its acceptance criteria and is substantially equivalent to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The device performance is demonstrated by compliance with several international standards. The specific acceptance criteria are defined within these standards. Below is a table summarizing the tests described and the reported outcome of compliance.

    Individual Test (Standard Clause)Standard Defining RequirementRequirement / Acceptance Criteria (Derived from Standard)Reported Device Performance
    Connector Tests (from ISO 80369-3:2016)AAMI/CN3 (PS):2014Specified in Clause 6.2 (Fluid Leakage)Complies with ISO 80369-3:2016 requirements
    Fluid LeakageAAMI/CN3 (PS):2014Specified in Clause 6.3 (Stress Cracking)Complies with ISO 80369-3:2016 requirements
    Stress CrackingAAMI/CN3 (PS):2014Specified in Clause 6.4 (Resistance to separation from axial load)Complies with ISO 80369-3:2016 requirements
    Resistance to separation from axial loadAAMI/CN3 (PS):2014Specified in Clause 6.5 (Resistance to separation from unscrewing)Complies with ISO 80369-3:2016 requirements
    Resistance to separation from unscrewingAAMI/CN3 (PS):2014Specified in Clause 6.6 (Resistance to overriding)Complies with ISO 80369-3:2016 requirements
    Resistance to overridingAAMI/CN3 (PS):2014Specified in Clause 6.7 (Disconnection by unscrewing)Complies with ISO 80369-3:2016 requirements
    Disconnection by unscrewing
    Syringe Tests (from ISO 7886-1:1993)ISO 7886-1:1993Specified in Clause 5 (Cleanliness)Complies with ISO 7886-1:1993 requirements
    CleanlinessISO 7886-1:1993Specified in Clause 6 (Limits for acidity or alkalinity)Complies with ISO 7886-1:1993 requirements
    Limits for acidity or alkalinityISO 7886-1:1993Specified in Clause 7 (Limits for extractable metals)Complies with ISO 7886-1:1993 requirements
    Limits for extractable metalsISO 7886-1:1993Specified in Clause 8 (Lubricant)Complies with ISO 7886-1:1993 requirements
    LubricantISO 7886-1:1993Specified in Clause 9 (Tolerance on graduated capacity)Complies with ISO 7886-1:1993 requirements
    Tolerance on graduated capacityISO 7886-1:1993Specified in Clause 10 (Graduated scale)Complies with ISO 7886-1:1993 requirements
    Graduated scaleISO 7886-1:1993Specified in Clause 11 (Barrel)Complies with ISO 7886-1:1993 requirements
    BarrelISO 7886-1:1993Specified in Clause 12 (Piston/plunger assembly)Complies with ISO 7886-1:1993 requirements
    Piston/plunger assemblyISO 7886-1:1993Specified in Clause 14.1 (Dead space)Complies with ISO 7886-1:1993 requirements
    Dead spaceISO 7886-1:1993Specified in Clause 14.2 (Freedom from air and liquid leakage past piston)Complies with ISO 7886-1:1993 requirements
    Freedom from air and liquid leakage past piston
    Biocompatibility Tests (from ISO 10993 series)ISO 10993-5:2009No CytoxicityNo Cytoxicity
    Tests for Vitro CytotoxicityISO 10993-10:2010No Irritation to SkinNo Irritation to Skin
    Tests for irritation and skin sensitizationISO 10993-10:2010No significant evidence of sensitizationNo significant evidence of sensitization
    ISO 10993 seriesConforms to requirements of ISO 10993 series StandardsConforms to requirements of ISO 10993 series Standards
    Sterilization & Packaging TestsISO 10993-7:2008Acceptable Ethylene oxide sterilization residualsComplies with ISO 10993-7:2008
    Ethylene oxide sterilization residualsASTM F 88/F88M-09Seal strength of flexible barrier materials meets standardComplies with ASTM F 88/F88M-09
    Seal strength (packaging)ASTM F1140/F1140-13Internal pressurization failure resistance meets standardComplies with ASTM F1140/F1140-13
    Internal pressurization failure resistance (packaging)USP37-NF32 <85>Bacterial Endotoxins within limitsComplies with USP37-NF32 <85>
    Bacterial Endotoxins Limit-Achieves SAL of 10^-6 for sterilizationAchieves SAL of 10^-6
    Sterilization Assurance Level (SAL)-Maintained sterility during shelf life of five yearsSterility maintained after accelerated aging tests

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state the specific sample sizes (number of units tested) for each non-clinical test. However, it indicates that "Non clinical tests were conducted to verify that the proposed device met all design specifications" and that "Connector testing performed on the proposed device included the items listed below, in accordance with ISO 80369-3:2016... using the test methods provided in ISO 80369-20..." and similarly for syringe testing. These standards often specify minimum sample sizes for different tests.

    The data provenance is from non-clinical (bench) testing performed on the proposed device. The device manufacturer is Jiangsu Caina Medical Co., Ltd. located in Jiangyin, Jiangsu, China. The report is a summary of these tests. This is retrospective in the sense that the tests were completed prior to this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This section is not applicable as the information provided pertains to non-clinical (bench) testing of a medical device (enteral syringe). "Ground truth" and "experts" typically refer to human interpretation or pathology results in diagnostic or clinical studies. For physical device testing, the "ground truth" is defined by the technical specifications and requirements within the cited international standards (e.g., ISO, ASTM, USP).

    4. Adjudication method for the test set

    This section is not applicable as adjacency methods are used in clinical studies with human readers, not in non-clinical bench testing. The compliance is determined by whether the device's physical and performance characteristics meet the quantitative and qualitative requirements of the listed standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is not applicable. An MRMC comparative effectiveness study is used for evaluating diagnostic or screening tools, often involving artificial intelligence. This document describes the non-clinical testing of a physical medical device (enteral syringe).

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This section is not applicable. "Standalone performance" in this context refers to the performance of an algorithm without human intervention, typically in AI/software as a medical device. This document is about a physical medical device.

    7. The type of ground truth used

    For the non-clinical tests, the "ground truth" is compliance with established international and national standards and their specified test methods and acceptance criteria. This includes:

    • Performance Specifications: Defined by standards like ISO 80369-3, ISO 7886-1 for connector and syringe function (e.g., fluid leakage, tolerance on graduated capacity).
    • Biocompatibility Standards: Defined by ISO 10993 series for biological safety (e.g., cytotoxicity, irritation, sensitization).
    • Sterilization and Packaging Standards: Defined by ISO 10993-7, ASTM F 88/F88M-09, ASTM F1140/F1140-13, USP37-NF32 <85>.

    8. The sample size for the training set

    This section is not applicable. "Training set" refers to data used to train machine learning models. This document describes the non-clinical testing of a physical medical device.

    9. How the ground truth for the training set was established

    This section is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1