510(k) Data Aggregation

    K Number
    K161849

    Validate with FDA (Live)

    Device Name
    Bright
    Date Cleared
    2016-07-29

    (24 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bright device is intended to be used for the treatment of wrinkles, rhytides and fine lines in the periorbital region.

    Device Description

    The Bright is an over-the-counter device that uses light emitting diodes to emit red light (660 nm) during treatment. The device includes a control unit, a treatment surface and a protective cap that fits over the treatment surface area of the device when not in use. The device controls consist of a single on/off activation button. The rechargeable battery included in the device can be charged using a USB cable that is supplied.

    AI/ML Overview

    The provided document is a 510(k) summary for the Bright device, a light-emitting diode (LED) therapy device intended for the treatment of wrinkles, rhytides, and fine lines in the periorbital region. The purpose of this 510(k) notice is to demonstrate substantial equivalence to a previously cleared device, the Tanda Max OTC (K110735), after modifications to create a smaller version.

    Based on the provided text, the device is not an AI product, but rather a physical medical device (LED therapy device). Therefore, the specific questions related to AI device performance, sample sizes for test sets and training sets, expert adjudication, MRMC studies, standalone algorithm performance, and ground truth establishment for AI models are not directly applicable or discussed in this document.

    The document focuses on demonstrating that the modified Bright device maintains the same safety and effectiveness as its predicate device despite physical changes.

    Here's a breakdown of the requested information based on the provided text, with an emphasis that this is for a physical device, not an AI model:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Test Methods/Standards)Reported Device Performance
    IEC 60601-1:2005 (General Safety)Functioned as intended
    IEC 60601-1-2 Ed 4.0 2014-02 (EMC)Functioned as intended
    IEC 62304 Ed 1.0 (Software Life Cycle)Functioned as intended
    ISO 10993-1 (Biocompatibility)Functioned as intended
    IEC 62471 (Photobiological Safety)Functioned as intended
    ISO 14971 (Risk Management)Functioned as intended
    Output Power (mW/cm²)Identical to predicate
    Treatment Cycle TimeIdentical to predicate
    Treatment RegimenIdentical to predicate
    Skin Contact SensorPreserved
    Temperature SensorPreserved

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not applicable or provided for a physical device's performance testing in this context. The document describes engineering and safety standard compliance testing for hardware and software, not a clinical trial with a "test set" of patients in the way an AI model would be evaluated. The testing is focused on the device's physical and electrical characteristics and adherence to recognized standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable or provided. The "ground truth" for a physical device's safety and performance is established by its compliance with recognized engineering, safety, and biological standards, as well as its functional specifications. It does not involve expert consensus on medical images or diagnoses in the way an AI medical device would.

    4. Adjudication Method for the Test Set

    This information is not applicable or provided. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving interpretation (e.g., radiology reads) where there can be disagreement among experts. This document describes compliance testing against established standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of AI on human reader performance, which is not pertinent to this physical LED therapy device.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    No, a standalone algorithm performance study was not done. This device is a physical LED therapy device, not an AI algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's safety and performance is based on compliance with established international and national engineering, safety, and biocompatibility standards (e.g., IEC 60601-1, ISO 10993-1). Additionally, its functional characteristics (e.g., output power, treatment cycle, sensor operation) were compared to the cleared predicate device.

    8. The Sample Size for the Training Set

    This information is not applicable or provided. "Training set" refers to data used to train an AI model. This document describes the manufacturing and testing of a physical medical device.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable or provided as it relates to AI model training data, not a physical device.