    K Number
    K161746

    Validate with FDA (Live)

    Device Name
    ARIX Hand System
    Date Cleared
    2016-09-21

    (89 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ARIX Hand System is intended for use in internal fixation of thand and wrist. Examples of these procedures may include but are not limited to replantation, lag screw techniques, joint fusions, corrective osteotomies, and the treatment of fractures.

    Device Description

    The ARIX Hand System is designed to fixation of the bones of hand and wrist. Examples of these procedures may include but are not limited to replantation, lag screw techniques, joint fusions, corrective osteotomies, and the treatment of fractures. The ARIX Hand System consists of the ARIX Hand System Plate and the ARIX Hand System Screw. The ARIX Hand System Plate is made of Pure Titanium (ASTM F67) and the ARIX Hand System Screw is made of the Titanium Alloy (ASTM F136). This device performance is not adversely affected by aging or storage conditions since this device is to be sterilized at the hospital before use and also to be single use.

    AI/ML Overview

    The document describes a 510(k) premarket notification for the ARIX Hand System, a bone plate and screw system. The submission aims to demonstrate substantial equivalence to a predicate device (K131566) also named ARIX Hand System. The information provided is primarily focused on the non-clinical testing performed to establish this equivalence.

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state acceptance criteria in terms of specific numerical thresholds that the device must meet for each test. Instead, it states that the tests were performed to "verify that the proposed device met all design specifications" and that the results "indicate that the ARIX Hand System is equivalent to predicate device." Therefore, the implied acceptance criterion is that the device's performance in these non-clinical tests is comparable or superior to the predicate device.

    Test NameAcceptance Criteria (Implied)Reported Device Performance
    Plate Tensile Strength TestComparable to or better than the predicate device (ASTM F382-99)Results indicate equivalence to the predicate device.
    Plate Bending Strength TestComparable to or better than the predicate device (ASTM F382-99)Results indicate equivalence to the predicate device.
    Screw Driving Torque TestComparable to or better than the predicate device (ASTM F543-07)Results indicate equivalence to the predicate device.
    Screw Axial Pull-out TestComparable to or better than the predicate device (ASTM F543-07)Results indicate equivalence to the predicate device.
    Screw Torsion TestComparable to or better than the predicate device (ASTM F543-07)Results indicate equivalence to the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the exact number of plates or screws tested for each non-clinical test.
    • Data Provenance: The tests are "Bench tests," which implies they were conducted in a laboratory setting. The country of origin of the data is not explicitly stated, but the submitter (Jeil Medical Corporation) is based in Korea. The studies are retrospective in the sense that they are laboratory tests on manufactured devices, not prospective clinical trials.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This section is not applicable as the study is a non-clinical bench testing study, not a clinical study involving diagnosis or interpretation by experts. Ground truth is established by the physical testing results against standard specifications.

    4. Adjudication Method for the Test Set

    This section is not applicable for a non-clinical bench testing study. Adjudication typically refers to expert review in clinical trials or diagnostic studies.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A MRMC comparative effectiveness study was not done. The document explicitly states: "No clinical studies were considered necessary and performed."

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This section is not applicable. The device is a physical medical implant (bone plate and screw), not an algorithm or AI system.

    7. Type of Ground Truth Used

    The ground truth used for these non-clinical mechanical tests is based on engineering standards and specifications (ASTM F382-99 for plates and ASTM F543-07 for screws). The performance of the subject device is compared against these standards and against the predicate device's performance.

    8. Sample Size for the Training Set

    This section is not applicable. There is no "training set" as this is a physical medical device undergoing mechanical testing, not a machine learning model.

    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable for the same reason as above.