Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K161723

    Validate with FDA (Live)

    Date Cleared
    2016-12-21

    (182 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ulrich Transfer Set is indicated for the transfer of OmnipaqueTM (Iohexol) contrast media as supplied in an Imaging Bulk Package to non pre-filled single-use only sterile syringes on syringe-based contrast injection systems indicated for the controlled, automatic administration on the venous side of contrast media for CT procedures. The Transfer Set is to be discarded after the contrast media container has been depleted or 8 hours have elapsed since the container was penetrated. whichever occurs first.

    Device Description

    The ulrich Transfer Set (Transfer Set) is a sterile tubing set connection between an imaging bulk package and empty sterile syringes on single-use-only, syringe based contrast injection systems. The Transfer Set consists of a spike with protective cap and integrated air filter on one side, 20" (50.8 cm) tubing segment and one female swabable (also referenced as swabbable) valve (Swabsite Swabbable Valve - K002689) with a protective cap on the other side. The Transfer Set is to be discarded after the contrast media container has been depleted or 8 hours have elapsed since the container was penetrated, whichever occurs first.

    AI/ML Overview

    The provided text describes the ulrich Transfer Set and its successful 510(k) submission for substantial equivalence to a predicate device. This is a medical device, and the acceptance criteria and supporting studies are focused on its safety and performance standards rather than AI/algorithm performance.

    Here's a breakdown of the requested information based on the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/TestAcceptance Criteria (Standard Compliance)Reported Device Performance
    SterilizationISO 11135-1:2014 (Sterilization of health care products - Ethylene oxide) to a SAL of 10⁻⁶Complies with the standard and achieves SAL of 10⁻⁶
    Shelf-LifeISO 11607-1:2006 (Packaging for terminally sterilized medical devices - Part 1)Complies with the standard
    ISO 11607-2:2006 (Packaging for terminally sterilized medical devices - Part 2)Complies with the standard
    BiocompatibilityISO 10993-4:2009 (Interactions with blood)Materials comply with the standard
    ISO 10993-5:2009 (In vitro cytotoxicity)Materials comply with the standard
    ISO 10993-10:2010 (Irritation and Skin Sensitization)Materials comply with the standard
    ISO 10993-11:2006 (Systemic toxicity)Materials comply with the standard
    ASTM F756-13 (Haemolytic Properties Materials)Materials comply with the standard
    ASTM F619-14 (Extraction of Medical Plastics)Materials comply with the standard
    USP 38 <151> (Pyrogen Test)Materials comply with the standard
    USP 38 <85> (Bacterial Endotoxins Test)Materials comply with the standard
    USP 38 <161> (Transfusion and infusion assemblies)Materials comply with the standard
    Performance - BenchISO 8536-4:2010 (Infusion equipment for medical use - Part 4)Complies with its predetermined specifications and applicable standards
    ISO 22413:2010 (Transfer sets for pharmaceutical preparations)Complies with its predetermined specifications and applicable standards
    ISO 594-2 (Conical fittings with 6% (Luer) taper)Complies with its predetermined specifications and applicable standards
    Microbial Ingress Testing(Implicit: no microbial ingress)Performed, results indicate compliance with specifications
    Chemical Compatibility Testing(Implicit: conformance to approved release specification of Omnipaque)Performed, results indicate compliance with specifications
    Extractables and Leachables Testing(Implicit: within safe limits)Performed, results indicate compliance with specifications
    Intended Use & Indications for UseSubstantially equivalent to predicate deviceMet through design comparison and testing

    2. Sample size used for the test set and the data provenance

    The document does not specify exact sample sizes for each test mentioned (e.g., number of transfer sets tested for sterility, biocompatibility, or performance). The studies are "verification results" indicating compliance with standards, implying a sufficient sample size was used as per the standard's requirements.

    • Data Provenance: The studies are laboratory-based bench tests and materials testing, conducted by ulrich GmbH & Co. KG or their contracted testing facilities. The country of origin of the data would likely be Germany (where ulrich GmbH & Co. KG is located) or the location of the testing laboratories. The studies are prospective in the sense that they were conducted specifically for this 510(k) submission to demonstrate compliance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This section is not applicable to this device submission. The ulrich Transfer Set is a physical medical device (an intravascular administration set), not an AI/software device that requires expert-established ground truth for diagnostic or interpretative tasks. The "ground truth" here is defined by meeting established engineering, chemical, and biological standards (e.g., a device is sterile if it meets SAL 10⁻⁶; materials are biocompatible if they pass ISO 10993 tests).

    4. Adjudication method for the test set

    This is not applicable as it pertains to expert consensus on diagnostic or interpretative AI performance. The tests described are objective, quantitative measurements against predefined standards.

    5. (If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance)

    This is not applicable. The device is an intravascular administration set, not an AI or imaging diagnostic tool. Therefore, no MRMC study, human reader improvement, or AI assistance is relevant.

    6. (If a standalone (i.e. algorithm only without human-in-the-loop performance) was done)

    This is not applicable. The device is a physical medical device, not an algorithm or AI.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    As explained in point 3, the concept of "ground truth" in the AI/diagnostic sense is not directly applicable. The "ground truth" for the ulrich Transfer Set's performance is established by:

    • International Standards: Compliance with ISO 11135-1, ISO 11607-1/2, ISO 10993 series, ISO 8536-4, ISO 22413, ISO 594-2.
    • National Standards: Compliance with ASTM F756-13, ASTM F619-14, USP 38.
    • Predetermined Specifications: The device's own internal design and performance specifications confirmed by testing.

    8. The sample size for the training set

    This is not applicable. The device is a physical medical device, not an AI model that requires a training set.

    9. How the ground truth for the training set was established

    This is not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1