    K Number
    K161691

    Validate with FDA (Live)

    Date Cleared
    2017-03-20

    (273 days)

    Product Code
    Regulation Number
    862.1475
    Age Range
    2 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For the quantitative in vitro determination of LDL-cholesterol concentration in human plasma and serum. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders mellitus), atherosclerosis and various liver and renal diseases.

    This in vitro diagnostic device is intended for prescription use only.

    Device Description

    The LDL Cholesterol kit assay consists of ready to use reagent solutions. CATALOGUE NUMBER: CH8312

    R1. Enzyme Reagent 1 4 x 20 mL R2. Enzyme Reagent 2 4 x 9 mL

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    Device Name: Direct LDL Cholesterol (LDL)

    Acceptance Criteria CategorySpecific MetricAcceptance Criteria (Implied/Direct)Reported Device Performance
    PrecisionTotal CV % (within run, among run, among day)Generally expected to be low for clinical assays, with specific targets often dependent on concentration levels.QC 1 (92.0 mg/dl): 5.9% total CV QC 2 (135.9 mg/dl): 4.6% total CV QC 3 (186.7 mg/dl): 4.4% total CV Serum pool 1 (65.0 mg/dl): 5.9% total CV Serum pool 2 (154.0 mg/dl): 5.0% total CV Serum pool 3 (200.1 mg/dl): 5.0% total CV Serum pool 4 (343.7 mg/dl): 5.3% total CV
    Linearity/Reportable RangeLinear Regression Correlation Coefficient (r)Close to 1.0 (indicating a strong linear relationship)r = 0.997
    Reportable RangeDefined range where results are linear.21 - 740 mg/dl
    Detection LimitLimit of Blank (LoB)Very low, ideally close to zero, to ensure no signal from blank.1.94 mg/dl
    Limit of Detection (LoD)Low enough to reliably detect the analyte.3.19 mg/dl
    Limit of Quantitation (LoQ)Low enough for precise and accurate quantification at low concentrations (typically ≤20% imprecision).16.1 mg/dl (with ≤20% imprecision)
    Analytical SpecificityInterference (% of Control)% of Control ± 10% for potential interferents at specified concentrations.Hemoglobin: No significant interference up to 1000mg/dl.Total Bilirubin: No significant interference up to 60mg/dl.Conjugate Bilirubin: No significant interference up to 60mg/dl.Triglycerides: No significant interference up to 500mg/dl.Intralipid®: No significant interference up to 500mg/dl.Ascorbic Acid: No significant interference up to 6mg/dl.
    Method ComparisonLinear Regression Correlation Coefficient (r)Close to 1.0 when compared to a predicate device, indicating substantial equivalence.r = 0.998 (compared to predicate device)
    Matrix ComparisonLinear Regression Correlation Coefficient (r)Close to 1.0 when comparing serum and lithium heparin plasma, indicating equivalent performance across matrices.r = 0.998 (serum vs. lithium heparin plasma)
    TraceabilityConformance to reference materials/standardsTraceable to an internal master reference material. Not certified by CRMLN (stated as a disclaimer in labeling).Traceable to an internal master reference. Labeling states "device has not been certified by the CRMLN."

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Precision (Analytical Performance): 80 determinations for each of 7 pools/QC levels (total of 560 determinations). The samples included control material and "unaltered human serum samples that were spiked with LDL cholesterol concentrations or diluted to achieve concentrations based on established ranges" (e.g., <100mg/dl, 130-159mg/dL, etc.). Data provenance is unspecified, but implied to be from a laboratory setting (prospective testing) rather than patient cohorts, using human serum.
      • Linearity/Assay Reportable Range: 11 levels, run in replicates of five on two lots of reagent. This means 11 * 5 * 2 = 110 individual measurements. The samples were prepared using "low and high serum pools." Data provenance is unspecified, but implied to be laboratory-prepared (prospective testing) using serum.
      • Detection Limit: 240 determinations (across 4 low-level samples) for LoD. Data provenance is unspecified, but implied to be laboratory-based (prospective testing).
      • Analytical Specificity: Two LDL Cholesterol concentrations (96.75 mg/dl and 193.5 mg/dl) were tested against various interferent levels. Each interferent was tested. The specific number of replicates or total determinations per interferent is not specified, but inferred to be multiple to allow for mean value calculation. Data provenance is unspecified, but implied to be laboratory-based (prospective testing).
      • Method Comparison with Predicate Device: 139 serum patient samples. Data provenance is unspecified, but these are patient samples, making this a retrospective or prospectively collected patient sample study. The country of origin for these patients is not specified.
      • Matrix Comparison: 70 matched patient sample pairs (serum and lithium heparin plasma). Data provenance is unspecified, but these are patient samples, making this a retrospective or prospectively collected patient sample study. The country of origin for these patients is not specified.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • For this type of in vitro diagnostic device (quantitative measurement of a biomarker), "ground truth" is typically established by reference methods or highly characterized calibrators/control materials, rather than expert human interpretation.
      • Precision, Linearity, Detection Limit, Analytical Specificity: Ground truth is based on the known concentrations of analytes in control materials, spiked samples, or reference preparations.
      • Method Comparison and Matrix Comparison: The ground truth for comparative studies is the measurement obtained from the "predicate device" or the paired sample (e.g., serum for matrix comparison). This is an analytical comparison rather than a clinical ground truth requiring expert consensus on a diagnosis.
      • Therefore, the concept of "number of experts" and "qualifications of experts" used for establishing ground truth as it would apply to imaging or diagnostic interpretation is not applicable here.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Given that this is an in vitro diagnostic device for quantitative measurement, adjudication methods typically used for subjective interpretations (like medical imaging diagnosis) are not applicable. The device provides a numerical output.
      • Results are compared to established reference values or predicate device measurements directly. Any discrepancies would be investigated through re-testing or troubleshooting, not through a formal adjudication process between experts primarily for the test set results.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. This type of study applies to devices that assist human readers (e.g., radiologists interpreting images). The "Direct LDL Cholesterol (LDL)" device is an in vitro diagnostic assay that generates a quantitative result directly from a patient sample; it does not involve human "readers" in the context of interpretation that an MRMC study would assess.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, this entire submission describes the standalone performance of the Direct LDL Cholesterol (LDL) assay system. It is an automated in vitro diagnostic test that provides a measurement result without human intervention for interpretation beyond operating the instrument and reviewing the numerical output.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The primary ground truth used is measurements from reference methods, predicate devices, or traceable calibrator/control values.
      • For method comparison, the predicate device's measurements serve as the comparator.
      • For precision, linearity, and detection limit, the "ground truth" refers to the known concentrations of the analyte in the prepared samples and controls.
      • The device mentions traceability to an "internal master reference material" for calibrators, which is a form of analytical ground truth.
    7. The sample size for the training set:

      • The document does not provide details on a "training set" in the context of machine learning, because this device is a chemical reagent-based assay, not an AI/ML algorithm.
      • The study describes the validation of the analytical performance of the reagent system and its associated instrument for quantitative measurement. The concept of training data, as used in AI, does not apply here.
    8. How the ground truth for the training set was established:

      • As there is no "training set" in the AI/ML sense, this question is not applicable. The analytical performance of the assay is based on chemical reactions and spectrophotometric measurements, calibrated against materials with known LDL Cholesterol concentrations.