    K Number
    K161672

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-09-29

    (112 days)

    Product Code
    Regulation Number
    876.5130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The catheter is indicated for use by patients with urine retention and patients with a post void residual volume (PVR) due to neurogenic and non-neurogenic voiding dysfunction. The catheter is inserted into the urethra to reach the bladder allowing urine to drain. The product is for male patients only.

    Device Description

    The SpeediCath Flex Coudé catheter is a sterile single use hydrophilic coated polyurethane catheter for men. The catheter is to be used for intermittent drainage of the bladder through the urethra by males with missing or reduced bladder control. The catheter has a flexible bendable tip that facilitates passage through the urethra to the bladder. The catheter is shielded by a sleeve, which serves as protection from the user's touch during insertion.

    AI/ML Overview

    Here is an analysis of the provided FDA document regarding the SpeediCath Flex Coudé device, focusing on acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

      Acceptance Criteria / Performance AspectReported Device Performance
      Bench Testing
      Flow ratePassed (according to EN1616/EN1618 & ASTM F623-99:2013)
      Coefficient of frictionPassed (according to ASTM D1894:2014)
      Tensile strengthPassed (according to EN1616/EN1618)
      Connector securityPassed (according to EN1616)
      BiocompatibilityPassed (according to ISO 10993-1 and applicable parts of FDA Blue Book Memorandum #G95-1)
      Clinical Performance (Cadaver Study)
      Insertion & navigation in difficult male urethral anatomiesComparable to a standard coudé catheter in terms of insertion and navigation through difficult male anatomies (enlarged prostate and strictures). Implies meeting performance equivalent to established devices for these challenging cases.
      Clinical Performance (Clinical Evaluation of Published Data)Safety and efficacy profile equivalent to the predicate and reference devices.

      Note: The document states "All tests passed" for the bench performance data, implying that the acceptance criteria for each of these tests were met.

    2. Sample Size and Data Provenance

      • Test set (cadaver study): The document mentions "male cadavers with difficult urethral anatomies." It does not specify the exact number of cadavers used.
      • Data Provenance (cadaver study): Not explicitly stated, but implies a laboratory setting for the "pre-clinical study." It is a prospective observational study in a pre-clinical setting.
      • Clinical Evaluation: Based on "available published data," which would be retrospective analysis of existing clinical literature. No specific sample size mentioned for this review beyond the studies themselves.
    3. Number of Experts and Qualifications for Ground Truth

      • Cadaver Study: The document does not specify the number of experts or their qualifications involved in establishing "ground truth" or evaluating insertion/navigation in the cadaver study. It only states the device "was comparable to a standard coudé catheter." This comparison would likely have been made by experienced individuals, but their number and specific qualifications are not detailed.
      • Clinical Evaluation: For the "clinical evaluation of available published data," the "experts" would be the authors of the published studies being reviewed. The document does not state how many individuals conducted this literature review or their qualifications.
    4. Adjudication Method for the Test Set

      • The document does not describe any formal adjudication method (like 2+1 or 3+1) for the cadaver study or the clinical evaluation. The comparison in the cadaver study ("comparable to a standard coudé catheter") implies a direct observation and assessment, but multi-reader adjudication is not mentioned.
    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

      • No MRMC comparative effectiveness study involving human readers with and without AI assistance was performed. This device is a medical device (catheter) for physical use, not an AI diagnostic/interpretive tool.
    6. Standalone Performance (Algorithm Only)

      • Not applicable. This is a physical medical device, not a software algorithm.
    7. Type of Ground Truth Used

      • Bench Testing: Engineering measurements and adherence to specified standards (e.g., ASTM, EN, ISO).
      • Cadaver Study: Direct observation and assessment of the device's physical performance (insertion and navigation) in anatomical models (cadavers) with known difficult anatomies (enlarged prostate and strictures). The "ground truth" is the physical reality of successful navigation and the comparative ease with a predicate.
      • Clinical Evaluation: Clinical outcomes and safety profiles reported in existing published literature for predicate and reference devices.
    8. Sample Size for the Training Set

      • Not applicable as this is a physical medical device, not a machine learning algorithm. The "development" or "training" would refer to design iterations and engineering evaluations, not a data-driven training set in the AI sense.
    9. How the Ground Truth for the Training Set was Established

      • Not applicable for the same reason as point 8. The "ground truth" for the device's design would stem from engineering principles, user needs, and clinical requirements for urinary catheterization.