510(k) Data Aggregation

    K Number
    K161498

    Validate with FDA (Live)

    Date Cleared
    2016-08-22

    (82 days)

    Product Code
    Regulation Number
    888.3075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Streamline OCT Occipito-Cervico-Thoracic System is intended to provide immobilization of spinal segments as an adjunct to fusion when using autograft for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine (T1 to T3): traumatic spinal fractures and/ or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability.

    The system is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

    In order to achieve additional levels of fixation, the Streamline OCT System may be connected to the Quantum Spinal Fixation System, Streamline MIS Spinal Fixation System or Streamline TL Spinal System using connectors and/or transition rods.

    Device Description

    The Streamline OCT System consists of a variety of rods, hooks, polyaxial screws, high-angle screws, locking caps, occipital plates, occipital screws, and connecting components used to build an occipito-cervico-thoracic spinal construct. System components are manufactured from ASTM F136 medical grade titanium alloy and ASTM F1537 medical grade cobalt chromium molybdenum alloy. Medical grade titanium alloy and medical grade cobalt chromium molybdenum alloy may be used together. The system should be implanted using only the surgical instruments designed for the system.

    The purpose of this submission is to: expand the use of the system to include the use of screws in the posterior cervical spine, add line extension components and modify an existing component.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the "Streamline OCT Occipito-Cervico-Thoracic System." It describes the device, its intended use, and argues for its substantial equivalence to previously cleared predicate devices. The information provided is about the device itself and its regulatory clearance process, rather than a study proving performance against specific acceptance criteria for AI or diagnostic performance metrics.

    Therefore, many of the requested sections about acceptance criteria and study details for an AI-powered diagnostic device cannot be extracted from this document.

    However, I can extract information related to the non-clinical testing performed for this physical medical device.

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table of precise acceptance criteria with numerical thresholds for performance (e.g., tensile strength > X, fatigue life > Y), nor does it report specific measured values for device performance against such criteria. Instead, it generally states that engineering analysis and re-creation testing demonstrated improvements and confirmed that the physical components met regulatory expectations.

    Acceptance Criteria CategoryReported Device Performance (General Statement from Document)
    Mechanical Performance (General)Engineering analysis confirmed that subject components do not introduce new worst-case components or impact previously completed construct testing.
    Mechanical Performance (Modifications)Re-creation testing was completed to demonstrate the modifications improve performance.
    Compliance with StandardsPreviously completed mechanical testing (ASTM F1717 and ASTM F2706) demonstrate that the subject and predicate systems are substantially equivalent.
    Safety & EffectivenessNo new risks to safety or effectiveness were raised by the published literature search or non-clinical testing.

    2. Sample size used for the test set and the data provenance

    For the non-clinical mechanical testing, the document does not specify the sample size used for each test (e.g., number of constructs tested for fatigue or static loading). It simply refers to "engineering analysis" and "re-creation testing."

    The term "data provenance" (e.g., country of origin, retrospective/prospective) is not applicable here as the testing is physical mechanical testing on device components, not clinical data from patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable to the non-clinical mechanical testing of this physical medical device. "Ground truth" in this context would typically refer to clinical diagnoses or outcomes, which are not part of the described engineering or mechanical testing.

    4. Adjudication method for the test set

    This question is not applicable to the non-clinical mechanical testing of this physical medical device. Adjudication methods like "2+1" or "3+1" are used in clinical studies to resolve discrepancies in expert interpretations of medical data, which is not what was performed for this submission.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation of medical images or data, often with AI assistance, to assess improvements in reader performance. This document is about a spinal implant system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a standalone (algorithm only) performance study was not done. This is also not applicable to a physical spinal implant system.

    7. The type of ground truth used

    For the non-clinical mechanical testing described, the "ground truth" would be established by the physical properties and behavior of the materials and constructs under specific, controlled loading conditions, as defined by engineering standards (e.g., ASTM F1717, ASTM F2706). It is not "expert consensus," "pathology," or "outcomes data" in the clinical sense.

    8. The sample size for the training set

    This concept is not applicable as this document is not describing an AI/machine learning device that requires a training set. The "training set" refers to data used to train an algorithm.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.