Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K161376

    Validate with FDA (Live)

    Date Cleared
    2016-08-25

    (99 days)

    Product Code
    Regulation Number
    874.4250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DragonFly™ Surgical Drill System has been designed for the light drilling of bones as part of surgical operations such as stapedotomy or ossiculoplasty.

    Device Description

    The DragonFly™ Surgical Drill System consists of a small surgical power tool (handpiece); a battery powered footswitch for control of speed and direction; various cutting instruments (burs); angled adaptors to adjust the angle between the handpiece and the bur; and a sterilization tray. Accessories include a battery charger and a battery replacement kit.

    AI/ML Overview

    This document describes the DragonFly Surgical Drill System, a surgical power tool for light drilling of bones in procedures like stapedotomy or ossiculoplasty.

    Here's an analysis of the acceptance criteria and the studies conducted:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" with quantitative targets for each performance metric, as might be seen for AI/Software-as-a-Medical-Device (SaMD) assessments. Instead, it describes performance verification and validation studies by comparing the device against predicate devices and demonstrating compliance with relevant standards. The "acceptance criteria" are implied by successful completion of these tests and comparable or superior performance to predicates.

    Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance
    Functionality After Repeated Autoclave CyclesMaintain normal specifications (not more than 15% decline in RPM and/or magnet strength, no discernable increase in vibration) for a minimum of 100 autoclave cycles.13 handpieces subjected to 270 autoclave cycles. First handpiece failed at 130 cycles, last failed after 270 cycles.
    Drill Speed (RPM)Consistent speed, within normal specifications.Average maximum RPM at t=0 was 9,535; at t=100 was 9,563. Highest RPM achieved was 9,840. (Note: Predicate devices have max 12,000 RPM, but DragonFly's lower RPM was deemed acceptable by surgeons in bench testing).
    Simulated Use (After Autoclave)No issues related to intended performance.No issues reported after 100 autoclave cycles.
    User's Understanding of IFUUsers successfully determine how to use the device.Users successfully read IFU and determined how to turn on/off, charge, assemble, perform startup, and evaluate functional characteristics.
    User's Ability to Determine Performance FailureUsers successfully recognize failures in essential performance.Users successfully recognized failures in essential performance.
    Software Detection of Essential PerformanceSoftware successfully detects failure, displays alarm, and shuts down system.Foot control software successfully detected failure, displayed alarm, and shut down.
    BiocompatibilityPass ISO 10993-1 tests (Intracutaneous Irritation, Acute Systemic Injection, Mem Elution L-929 Mouse Fibroblast Cells, Guinea Pig Sensitization).Representative burs passed all listed ISO 10993-1 tests.
    Software Safety LevelClassified as moderate level of concern (FDA guidance) and Risk Class A (IEC 62366), with verification and validation complete.Software classified as moderate level of concern and Risk Class A. Verification and validation conducted.
    Electrical SafetyCertified according to IEC 60601-1 and related standards.Certified according to IEC 60601-1-2 and ANSI/AAMI ES60601-1:2005/(R) 2012, CSA CAN/CSAC22.2 NO. 60601-1:14, IEC 60601-1 Edition 3.1 (2012) / EN 60601-1 (2006) + A11 + A1 + A12.
    Cleaning and ReprocessingValidation of cleaning and steam autoclavability.Cleaning validation (radioactive and protein markers) and steam autoclavability validation performed by external labs.
    Shelf-life (Burs)Validated for 3 years.Gamma irradiated burs validated for a 3-year shelf-life.
    Handling and Bur Runout (Comparison to Predicate)At least as good as or better than predicate devices.Tested in parallel with Xomed Skeeter Drill and OSSEOSTAP; confirmed handling and bur runout were at least as good or better.
    Surgical Evaluation (Comparison to Predicate)Comparable to or better than predicate devices in operation, handling, cutting performance, and precision suitable for intended use.Evaluated by 15 otologic surgeons in temporal bone labs, deemed at least as good as predicate Xomed Skeeter drill. Board-certified otolaryngologist compared to OSSEOSTAP and Skeeter, scored 20 characteristics, deemed comparable or better, and suitable for intended use.
    Magnetic Latching MechanismAverage axial force to remove bur: adequate. Lateral force to stall bur: adequate. Minimum lateral force to disarticulate bur: adequate.Average axial force of 1.1 lb. required to remove a bur. Lateral force of 0.24 lbs required to stall a 2.3mm cutting bur. Minimum lateral force of 1.2 lb. required to disarticulate a bur.
    Magnetic Drive MechanismCapable of transferring adequate torque.Capable of transferring a minimum of 0.45 in-oz of torque.
    Movable Outer Bur GuardNo new risks created, desirable, safe, and adequate for intended use.No new risks created. Surgeon feedback indicated it was desirable, safe, and adequate.

    2. Sample Size Used for the Test Set and Data Provenance

    • Functionality after repeated autoclave cycles: 13 handpieces.
    • Biocompatibility: Representative burs were tested.
    • Bench Testing and Surgical Evaluation of Performance:
      • Comparison of handling and bur runout: Not explicitly stated beyond "The DragonFly Surgical Drill system was tested in parallel with the predicate device".
      • Temporal bone labs evaluation: 15 Otologic surgeons.
      • Comparison by board-certified otolaryngologist: One board-certified otolaryngologist.
    • Data Provenance: Not explicitly stated as retrospective or prospective or country of origin in the provided text. However, temporal bone labs and surgeon evaluations suggest a prospective, controlled testing environment, likely in the US where the submitter is located.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Autoclave/Drill Speed/Simulated Use/Usability: While the tests involve "users" and "surgeons," the "ground truth" here is objective performance metrics (RPM, presence of issues) and user feedback on usability, rather than diagnostic "ground truth" like in imaging studies.
    • Surgical Evaluation:
      • 15 Otologic surgeons: These surgeons were "familiar with and had used the predicate Xomed Skeeter drill." This implies clinical expertise in oto-laryngology. Their specific years of experience are not stated.
      • 1 board certified otolaryngologist: Performed surgical procedures on human temporal bone and provided comparative scores for 20 performance characteristics. This qualification establishes significant clinical expertise.

    4. Adjudication Method for the Test Set

    • For the surgical evaluations:
      • The 15 otologic surgeons provided individual feedback and generally "deemed" the DragonFly as "at least as good as" the predicate. This suggests consensus-based or individual assessments rather than a formal pre-defined adjudication method (like 2+1 or 3+1).
      • The single board-certified otolaryngologist provided comparative scores for 20 characteristics. This was a direct comparison by a single expert.
    • For other performance tests (autoclave cycles, biocompatibility, electrical safety, etc.), the adjudication involves meeting predefined engineering specifications and regulatory standards, evaluated through testing rather than expert medical adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

    • No, an MRMC comparative effectiveness study involving AI assistance for human readers was not performed. This device is a surgical drill system, a hardware device, not an AI/SaMD diagnostic tool. The "readers" here are surgeons evaluating the physical device's performance, not interpreting medical images or data with AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    • No, this question is not applicable. The device is a surgical drill system; it is not an algorithm or AI. It requires a human operator (surgeon).

    7. The Type of Ground Truth Used

    The concept of "ground truth" as typically used in AI/diagnostic studies (e.g., pathology, clinical outcomes) doesn't directly apply in the same way to a surgical tool. Instead, the "truth" is established through:

    • Objective Performance Metrics: RPM, autoclave cycle durability, force measurements (for magnetic latching/drive), biocompatibility test results, electrical safety certifications, shelf-life validation.
    • Expert Clinical Opinion/Comparative Assessment: Direct subjective evaluation and comparison by qualified otologic surgeons and a board-certified otolaryngologist during bench testing and temporal bone labs, confirming the device's suitability and comparable performance to established predicates.

    8. The Sample Size for the Training Set

    • Not applicable. This is a hardware device; there is no "training set" in the context of machine learning or AI.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. This is a hardware device; there is no "training set" with associated "ground truth."
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1