Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K161303

    Validate with FDA (Live)

    Date Cleared
    2017-01-26

    (261 days)

    Product Code
    Regulation Number
    888.3030
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The dynaMXTM Compression Plate is indicated for fixation of small fragments, osteolomics, atthrodeses, replantations, and reconstructions of small bones and small fragments.

    Device Description

    The dynaMX™ Compression Plate provides a means of fixation for small bones and small fragments. The dynaMÄ™ Compression Plate is made of biocompatible Nitinol. The bridge of the implant is designed to exhibit superelastic properties at room temperature.

    AI/ML Overview

    I apologize, but the provided text is a 510(k) summary for a medical device (dynaMX™ Compression Plate) and primarily focuses on establishing substantial equivalence to predicate devices through laboratory bench testing.

    Therefore, the input document does not contain any information regarding clinical studies, acceptance criteria related to clinical performance, or the involvement of human readers or experts for ground truth establishment.

    The document describes bench tests conducted to verify suitability and substantial equivalence, such as:

    • Elastic Static Bending Testing
    • Bending Fatigue Testing
    • Cyclic Potentiodynamic Polarization Corrosion Testing
    • Galvanic Corrosion Testing
    • Compressive Force Performance
    • Transformation Temperature Determination
    • Package Seal Strength Verification

    It also mentions testing for locking screws, including:

    • Insertion and Removal Torque
    • Torsional Strength
    • Pullout Strength

    Without information about clinical performance acceptance criteria and studies linked to human or algorithm performance, and ground truth establishment, I cannot fulfill your request for the specific details you've asked for.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1