    Device Name
    ClearRead CT
    Date Cleared
    2016-09-09

    (134 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ClearRead CT™ is comprised of computer assisted reading tools designed to aid the radiologist in the detection of pulmonary nodules during review of CT examinations of the chest on an asymptomatic population. The ClearRead CT requires both lungs be in the field of view. ClearRead CT provides adjunctive information and is not intended to be used without the original CT series.

    Device Description

    ClearRead CT is a dedicated post-processing application that generates a secondary vessel suppressed Lung CT series with CADe marks and associated region descriptors intended to aid the radiologist in the detection of pulmonary nodules.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the ClearRead CT device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Localization Receiver Operating Characteristic (LROC) AUCClearRead CT found to significantly increase the AUC compared to the unaided read, indicating superior performance for detecting nodules.
    Radiologists' Interpretation TimeClearRead CT found to decrease read times with and without outliers.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the exact sample size for the test set (number of cases). It refers to a "multi-reader multi-case (MRMC) study."

    The data provenance (country of origin, retrospective/prospective) is not specified in the provided text.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number of experts used to establish ground truth or their qualifications. It mentions a "multi-reader multi-case (MRMC) study," implying multiple readers were involved in the evaluation, but not necessarily in establishing the initial ground truth.

    4. Adjudication Method for the Test Set

    The document does not explicitly describe the adjudication method used for the test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    Yes, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done.

    • Effect Size of Human Readers Improvement with AI vs. Without AI Assistance: The study found that using ClearRead CT "significantly increase[d] the AUC" of the LROC response. It also found that ClearRead CT "decrease[d] read times with and without outliers." While a specific numerical effect size (e.g., a percentage increase in AUC or specific time reduction) is not provided in this summary, the terms "significantly increase" and "decrease" indicate a positive and measurable improvement.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The document focuses on the multi-reader multi-case study, explicitly stating that ClearRead CT is "designed to aid the radiologist" and "provides adjunctive information and is not intended to be used without the original CT series." This implies the primary evaluation was human-in-the-loop. It also mentions that the device "generates a secondary vessel suppressed Lung CT series with CADe marks," which could be considered a standalone function, but the performance metrics provided are for the combined human-AI workflow.

    7. The Type of Ground Truth Used

    The document does not explicitly state the type of ground truth used (e.g., expert consensus, pathology, outcomes data).

    8. The Sample Size for the Training Set

    The sample size for the training set is not mentioned in the provided text.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide information on how the ground truth for the training set was established.