Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K161090

    Validate with FDA (Live)

    Device Name
    Cranial PSI
    Manufacturer
    Date Cleared
    2017-01-19

    (276 days)

    Product Code
    Regulation Number
    882.5330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OSSDSIGN Cranial PSI is intended for the reconstruction of cranial defects. It is indicated for non-load bearing applications for patients in whom cranial growth is complete and for use with or with or with or without duraplasty.

    Device Description

    OSSDSIGN Cranial PSI (Patient Specific Implant) is a device that replaces native bone in the cranial skeleton. Each Cranial PSI is a patient-specific device specifically created for a patient's unique anatomical requirements. Cranial PSI consists of a rigid titanium mesh that is largely covered by biocompatible ceramic tiles. The ceramic tiles are in a mosaic pattern that provides space between tiles to allow free circulation of fluids.

    AI/ML Overview

    The provided document is a 510(k) summary for the OSSDSIGN Cranial PSI device. It includes information about acceptance criteria and mechanical/biocompatibility testing comparing it to a predicate device, but it does not describe studies involving human performance or assessment by experts. Therefore, many of the requested fields cannot be answered from this document.

    Here's the information that can be extracted:

    1. A table of acceptance criteria and the reported device performance

    Test Method and Relevance SummaryAcceptance CriteriaOSSDSIGN Cranial PSI PerformanceKLS Martin - PCI Titanium Mesh (Predicate) Performance
    Dynamical Load Test: To verify that the Cranial PSI supports the forces exerted onto the implant from sleeping during the device lifetime. Method: orbital shaker, 125 rpm, 60 hours. Relevance: test simulates changed head position every 20 minutes during 8 hours sleep for 50 years.No deformationNo deformationNo deformation
    Max Force [N]: To establish the maximum force that can be applied to the device before failure. Method: Universal testing machine at 1 mm/min. Relevance: protection from falling objects and blunt trauma.>100 N461 N383 N
    Energy absorption [mJ]: To establish the maximum energy the device can absorb before failure. Method: Universal testing machine at 1 mm/min. Relevance: protection from falling objects and blunt trauma.>1000 mJ2260 mJ3760 mJ
    Resistance to Deformation [mm]: To establish device resistance to deformation before device failure. Method: Universal testing machine at 1 mm/min to 100 N applied force. Relevance: Both the subject device and predicate passed the test.<6 mm0.5 mm1.1 mm

    Biocompatibility Acceptance Criteria and Results:

    TestTest Method SummaryAcceptance Criteria (Implied by results)Results for OSSDSIGN Cranial PSI
    CytotoxicityISO 10993-5 Method: L-929 mouse fibroblast cellsNot cytotoxic; No evidence of causing cell lysis or toxicityNot cytotoxic; No evidence of causing cell lysis or toxicity
    SensitizationISO 10993-10 Method: guinea pig maximization testNon-sensitizer; No evidence of causing delayed dermal contact sensitizationNon-sensitizer; All test article extracts showed no evidence of causing delayed dermal contact sensitization in the guinea pig.
    Sensitivity, IrritationISO 10993-6 Method: polar and non-polar test article extracts intracutaneously injected into five separate sites on the right side of the back of three rabbits. Observations for erythema and edema at 24, 48, and 72 hours.Non-irritant; Difference between test and control mean scores <1.0Non-irritant; Difference between the test and the control mean scores <1.0, confirming the device is a non-irritant
    Systemic (acute) toxicityISO 10993-11 Method: 20 mice observed at 4, 24, 48 and 72 hours.Non-toxic; No mortality or evidence of systemic toxicityNon-toxic; No mortality or evidence of systemic toxicity
    GenotoxicityISO 10993-3 Method: mouse lymphoma forward gene mutation assay.Not mutagenic; <2-fold increase in mean mutant frequencyNot mutagenic; <2-fold increase in mean mutant frequency of the L5178Y/TK cell line.
    MutagenicityISO-10993-3 Method: bacterial reverse mutation studyNon-mutagenic; <2-fold increase in mutagenic frequencyNon-mutagenic; <2-fold increase in mutagenic frequency.
    ImplantationISO 10993-6 Test for local effects after implantation, 2- and 6-week subcutaneous implantations in rabbits.Non-irritant compared to controlNon-irritant; Non-irritant compared to control.
    Analytical Extractable Chemical AnalysisISO 10993-12 Analysis by GC-MS, ICP-MS, LC-MS and HPLC-ELSDNon-hazardous; No hazardous materials detected in various extractionsNon-hazardous in all results, including extractions in polar, non-polar and mid-polar solvents. No hazardous materials detected.
    Indirect HemolysisISO 10993-4 Biological evaluation of medical devices - Part 4: Selection of tests for interactions with bloodHemolytic index within acceptable criteria (implied by "All samples passed the acceptance criteria")The hemolytic index for the test article extract was 3.8% and the test article extract was slightly hemolytic. All samples passed the acceptance criteria.
    Material Mediated PyrogenicityISO 10993-11 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity, Pyrogen TestTotal rise of temperatures during the 3 hour observation period within acceptable limits; Nonpyrogenic (implied by results)The total rise of temperatures during the 3 hour observation period was within acceptable limits. The test article was judged as nonpyrogenic. All samples passed the acceptance criteria.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: The document does not specify a distinct "test set" sample size in the context of human data or image analysis. The tests mentioned are mechanical and biocompatibility tests performed on the device itself or in animal models/in vitro assays.
      • For the mechanical tests, the "size" of the device tested was 197 cm² for both the subject device and the predicate. The number of individual samples tested is not stated.
      • For biocompatibility:
        • Cytotoxicity: L-929 mouse fibroblast cells (number of cells/replicates not specified).
        • Sensitization: Guinea pigs (number not specified).
        • Irritation: 3 rabbits.
        • Systemic toxicity: 20 mice.
        • Genotoxicity: Mouse lymphoma cell line (L5178Y/TK).
        • Mutagenicity: Bacterial strains (not specified).
        • Implantation: Rabbits (number not specified) for 2- and 6-week implantation.
    • Data Provenance: The document does not provide information on data provenance (e.g., country of origin, retrospective/prospective) as it relates to clinical or imaging data, which are not part of this 510(k) summary. The company is based in Uppsala, Sweden.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable as the document describes mechanical and biocompatibility testing of a physical device, not a diagnostic algorithm requiring ground truth established by human experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable as the document describes mechanical and biocompatibility testing of a physical device, not a diagnostic algorithm requiring adjudication of human expert interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. No MRMC comparative effectiveness study involving human readers or AI assistance is mentioned in this 510(k) summary. The study is a non-clinical comparison of a device's physical properties with a predicate device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. This document evaluates a physical medical device (a cranial implant), not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the mechanical tests, the "ground truth" is defined by the objective performance measurements against pre-defined engineering requirements (e.g., no deformation, >100 N max force, <6 mm deformation).
    For biocompatibility tests, the "ground truth" is established by the results of standardized biological assays (e.g., looking for cell lysis, sensitization, irritation, toxicity, mutagenicity, pyrogenicity) as per ISO standards. These are objective measures based on biological reactions.

    8. The sample size for the training set

    This information is not applicable. The device is a physical implant, not an AI algorithm that requires a training set. The term "training set" is not relevant to the described testing.

    9. How the ground truth for the training set was established

    This information is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1