    K Number
    K161012

    Validate with FDA (Live)

    Date Cleared
    2016-07-06

    (86 days)

    Product Code
    Regulation Number
    886.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Easy-Load Lens Delivery System is indicated for the folding and injection of Bausch + Lomb intraocular lenses identifying the Easy-Load Lens Delivery System in their approved labeling.

    Device Description

    The Easy-Load Lens Delivery System is used for folding and delivering a Bausch + Lomb three-piece silicone IOL into the eye. An IOL is placed into the loading area and the drawer is closed. This compresses the IOL. The plunger is advanced until it stops at a detent position. The distal end is filled with viscoelastic or balanced salt solution and placed through an incision into the eye. The haptic puller is used to place the leading haptic in the correct loading position. Once the tip is in the eye, the plunger is advanced until the lens is fully expressed into the capsular bag.

    AI/ML Overview

    This document describes the premarket notification for the "Easy-Load Lens Delivery System" (Model EZ-24). It's an intraocular lens guide used for folding and injecting intraocular lenses.

    Here's the information regarding the acceptance criteria and study, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    ItemTESTAcceptance CriteriaRESULTS
    Function Performance(Before and after insertion) IOL surface & bulk homogeneityFree from surface and bulk defects and all edges should appear smooth when viewed at 10X magnificationPass
    (Before and after insertion) IOL lens powerThe diopter power of the lenses shall be within the allowed tolerancePass
    (Before and after insertion) IOL image qualityThe image quality of the lenses shall meet or exceed the minimum resolution group element of the Air Force target specified in the protocol.Pass
    (Before and after insertion) IOL dimensionsThe lenses must meet the dimensional requirements specified in the protocol.Pass
    IOL delivery outcomeThe IOLs do not flip over upon delivery; The IOL must exit inserter upon completion of delivery; No cosmetic defects to the IOL's haptic from folding and/or deliveryPass
    Damage to insertion deviceNo damage to the insertion device due to lens deliveryPass
    Lubricant transfer studyNo detectable transferPass
    Particle studyComparable to or lower than controlPass
    Sterilization verificationComparative Resistance Performance DeterminationFor a given exposure time the average recovery of viable test organisms from biological indicators placed within the barrel of either the EZ-24 or the EZ-28 shall differ by no more than one logarithm of the average of the values determined for both inserters. Alternatively, the average recovery of viable test organisms from biological indicators placed within the barrel of the EZ-24 may be less than the average recovery of viable test organisms from biological indicators placed within the barrel of the EZ-28 insertersPass
    EO transfer test$\le$ 0.5μg EO per IOLPass
    Endotoxin test$<$ 0.2 EU/devicePass
    Packaging verificationPlunger push force and cartridge detachment testingCartridge detachment force is greater than the force to disengage the plunger from the detents in the injector body.Pass

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    The document does not explicitly state the sample size for each specific test or the data provenance. It mentions "nonclinical tests" and "accepted ISO standards," implying laboratory testing rather than human clinical trials. The data is internal to the manufacturer, Bausch + Lomb.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This information is not provided. The tests described are primarily objective physical and chemical measurements ("Free from surface and bulk defects," "diopter power," "resolution group element," "dimensional requirements," "IOL delivery outcome," "no damage," "no detectable transfer," "comparable to or lower than control," "recovery of viable test organisms," "$\le$ 0.5μg EO," "$<$ 0.2 EU/device," "force greater than"). Expert judgment for these specific criteria would likely be embedded in the test protocol design and execution, but the document doesn't detail the number or qualifications of individuals involved in setting up or interpreting these nonclinical tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving interpretation of medical images or subjective evaluations by multiple human readers. For the nonclinical tests described in this document, such adjudication methods are not applicable. The results are based on direct measurements and adherence to specified physical/chemical parameters.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No MRMC study was mentioned. This device is an intraocular lens delivery system, not an AI-powered diagnostic tool, so MRMC studies and "human readers improve with AI" are not relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This is not an algorithm or AI product, so the concept of standalone performance for an algorithm is not applicable. The tests performed are on the physical device itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the nonclinical tests is based on established engineering specifications, accepted ISO standards (ISO 11979-2, ISO 11979-3, and ISO 11135 where applicable), and internal protocols. For example:

    • IOL surface & bulk homogeneity: "Free from surface and bulk defects and all edges should appear smooth when viewed at 10X magnification." (Direct observation against a visual quality standard)
    • IOL lens power: "within the allowed tolerance." (Measurement against a quantitative specification)
    • Sterilization verification: "recovery of viable test organisms... differ by no more than one logarithm..." (Microbiological standard)
    • EO transfer test: "$\le$ 0.5μg EO per IOL." (Chemical quantitative standard)

    8. The sample size for the training set:

    There is no mention of a "training set" as this is not an AI/machine learning device. The tests are on the physical product and its components.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for an AI model. For the nonclinical tests conducted, the acceptance criteria (which serve as the "ground truth" for pass/fail) were established based on ISO standards and internal protocols.