Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K160985

    Validate with FDA (Live)

    Device Name
    Pantera Pro
    Manufacturer
    Date Cleared
    2016-10-07

    (182 days)

    Product Code
    Regulation Number
    870.5100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pantera Pro is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The Pantera Pro (balloon diameter 2.0 – 4.0 mm) is also indicated for post-delivery expansion of balloon expandable stents.

    Device Description

    BIOTRONIK's Pantera Pro is a sterile, single-use, intravascular balloon catheter for the dilatation of stenotic segments in coronary arteries or bypass grafts. The dilatation balloon is designed to inflate to a known diameter and length at a specific inflation pressure. Pantera Pro has balloon diameters ranging from ∅ 1.25 mm – 4.0 mm and balloon lengths ranging from ∅ 6 mm – 30 mm. All product sizes of Pantera Pro have a usable catheter length of 140 cm. The Pantera Pro has a soft tapered tip at the distal end of the catheter to facilitate advancement of the catheter. The balloon is folded to achieve a low crossing profile and is connected to the tip. The Pantera Pro has two coatings; a hydrophobic coating on the outer surface of the hypotube and a hydrophilic polymeric based coating on the outer surface of the distal outer shaft. In addition, a hydrophobic coating is applied on the balloons of sizes ∅ 2.5 – 4.0 mm whereas a hydrophilic coating is applied on the balloons of sizes ∅ 1.25 – 2.0 mm. The coating is activated by immersing the catheter in sterile saline prior to use. The Pantera Pro is compatible with guide wires of 0.014" (0.36 mm) diameter and guiding catheters with an inner diameter of ≥0.056" (1.42 mm; 5F). For all Pantera Pro product sizes, the nominal pressure (NP) required to achieve the nominal diameter is 7 atm and the rated burst pressure (RBP) is 14 atm.

    AI/ML Overview

    The provided text describes a medical device, the Pantera Pro Coronary Dilatation Catheter, and its acceptance for market. However, it does not detail a study involving AI or human readers, so I will answer based on the information available regarding the device's performance through non-clinical testing.

    The acceptance criteria for the Pantera Pro Coronary Dilatation Catheter are based on various in-vitro bench tests and biocompatibility testing, ensuring it meets performance specifications and is biologically safe for its intended use. The study proving the device meets these criteria is a non-clinical design verification in-vitro testing and biocompatibility assessment.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific TestAcceptance Criteria (Implied/Derived from context)Reported Device Performance (Summary from text)
    Design VerificationDimensional Verification (including visual inspection)Meets design specificationsCompleted with satisfactory results.
    Balloon Preparation, Deployment and RetractionFunctionality within specificationCompleted with satisfactory results.
    Balloon Rated Burst Pressure≥ 14 atmNominal pressure (NP) is 7 atm, Rated Burst Pressure (RBP) is 14 atm, consistent across sizes. Testing confirmed performance.
    Balloon FatigueWithstands repeated inflation cyclesCompleted with satisfactory results.
    Balloon ComplianceMeets specified diameter at pressureCompleted with satisfactory results.
    Balloon Inflation and Deflation TimeWithin clinically acceptable timesCompleted with satisfactory results.
    Catheter Bond StrengthSufficient to maintain integrityCompleted with satisfactory results.
    Tip Pull TestSufficient to prevent detachmentCompleted with satisfactory results.
    Flexibility and Kink TestMaintains patency and navigabilityCompleted with satisfactory results.
    Torque StrengthSufficient for precise navigationCompleted with satisfactory results.
    RadiopacitySufficient for visualization during procedureCompleted with satisfactory results.
    Coating IntegrityCoatings remain intact and functionalCompleted with satisfactory results.
    Particulate EvaluationMeets cleanliness requirementsCompleted with satisfactory results.
    Shelf Life testingSupports claimed shelf lifeProvided to support a 3-year shelf life.
    For Stent Expansion - Balloon Rated Burst Pressure in Stent≥ 14 atm (same as stand-alone but within stent)Completed with satisfactory results.
    For Stent Expansion - Balloon Fatigue / Repeat Balloon Inflation in StentWithstands repeated inflation cycles within stentCompleted with satisfactory results.
    BiocompatibilityCytotoxicity, Sensitization, Irritation/Intracutaneous Reactivity, Acute Systemic Toxicity, Pyrogenicity, Hemolysis, Complement activation, In Vivo Thrombogenicity, Hemocompatibility - Chandler Test, Bacterial Reverse Mutation Assay - AMES test, Mouse Lymphoma Assay, Mammalian Micronucleus TestMeets ISO 10993-1:2009 standards and
    FDA recommendations for limited exposureAll tests completed and the device "met all specified design and performance requirements. No new safety or effectiveness issues were raised during the testing." Biologically safe for intended use.
    SterilizationSterility Assurance Level (SAL)1 x 10-6Sterilized with EO gas to achieve a SAL of 1 x 10-6.
    Endotoxin Limit< 20 EU / deviceMeets the endotoxin limit of 20 EU / device.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a distinct "test set" in terms of patient data or clinical samples for the non-clinical studies. Instead, the tests were performed on device samples (Pantera Pro Coronary Dilatation Catheter units). The sample size for each specific bench test (e.g., number of balloons tested for burst pressure, number of catheters for bond strength) is not explicitly stated in this summary but would be detailed in the full design verification report.

    Data Provenance: The data provenance for these non-clinical tests is from in-vitro bench testing conducted by the manufacturer (BIOTRONIK, Inc.). There is no mention of country of origin for the data beyond the manufacturer's location in Lake Oswego, Oregon, USA. These are retrospective data in the sense that they are derived from controlled laboratory experiments and quality assessments during the device's development and validation, rather than prospective clinical trials.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This question is not applicable to the provided document. The "ground truth" for non-clinical performance and biocompatibility testing is established by engineering specifications, regulatory standards (e.g., FDA guidance documents, ISO standards), and pre-defined pass/fail criteria for each test. These are objective measurements and do not rely on expert consensus or interpretation of complex medical images or patient outcomes. The performance is compared against these objective, quantitative targets.

    4. Adjudication Method for the Test Set

    This is not applicable. Adjudication methods (like 2+1, 3+1) are typically used in studies involving human interpretation or subjective assessments to resolve discrepancies, often in imaging studies. Since the studies described are non-clinical, objective bench tests and biocompatibility assays, no adjudication method was employed. The results are quantitative and directly measured against established criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document explicitly states: "No clinical investigation has been performed for this device." Therefore, there is no information on the effect size of how much human readers improve with AI vs. without AI assistance, as AI is not a component of this device, and no human-reader studies were conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    No, a standalone (algorithm only) performance study was not done. This device is a physical medical device (a catheter), not an artificial intelligence algorithm or software. Therefore, the concept of algorithm-only performance is not applicable.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical studies relies on:

    • Engineering Specifications: Defined parameters for device dimensions, material properties, mechanical performance (e.g., burst pressure, flexibility, bond strength).
    • Regulatory Standards: Compliance with industry standards (e.g., ISO 10993 for biocompatibility) and FDA guidance documents setting performance expectations.
    • Established Test Methodologies: Standardized methods to objectively measure performance against the above specifications and standards.

    8. The Sample Size for the Training Set

    This question is not applicable. The device described, the Pantera Pro Coronary Dilatation Catheter, is a physical medical device; it is not an AI/ML algorithm that requires a "training set" of data. Its design and performance are validated through engineering and biocompatibility testing as opposed to machine learning techniques.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable, as there is no "training set" for this physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1