510(k) Data Aggregation

    K Number
    K160916

    Validate with FDA (Live)

    Date Cleared
    2016-06-28

    (88 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NuVasive CoRoent Ti-C System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion.

    The CoRoent Ti-C System is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar devices are to be used in patients who have had at least six months of non-operative treatment. The System is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the lumbar spine.

    Device Description

    The NuVasive® CoRoent Ti-C System is manufactured from PEEK-OPTIMA® (Polyetherether-ketone) conforming to ASTM F2026, commercially pure titanium conforming to ASTM F1580. Ti-6Al-4V ELI conforming to ASTM F136/1472 or Tantalum (Ta) conforming to ASTM F560 or ISO 13782. The implants are available in a variety of different shapes and sizes to suit the individual pathology and anatomical conditions of the patient.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for a medical device called the CoRoent® Ti-C System, an intervertebral body fusion device. This document focuses on demonstrating substantial equivalence to previously cleared predicate devices rather than proving performance against specific acceptance criteria through a clinical study. Therefore, much of the requested information (like specific performance criteria, expert adjudication, MRMC studies, and detailed ground truth establishment) is not present.

    However, I can extract the available information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Since this is a 510(k) submission based on substantial equivalence, there are no specific "acceptance criteria" in the traditional sense that a clinical study would aim to meet for a new, novel device. Instead, the device's performance was compared to established standards and predicate devices.

    Acceptance Criterion (Established Standard/Predicate Comparison)Reported Device Performance (Assessment)
    Static and dynamic axial compression and compression shear per ASTM F2077Meets requirements; presents no new worst-case for performance testing compared to predicates.
    Wear debris testing per ASTM F2077, ASTM F1714 and ASTM F1877Meets requirements; presents no new worst-case for performance testing compared to predicates.
    Bacterial endotoxin testing (BET) per ANSI/AAMI ST-72:2011Meets requirements; demonstrates sterility information for devices labeled as sterile.
    Substantial equivalence in design, intended use, and material composition to predicate devicesFound to be substantially equivalent.

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: This document does not describe a "test set" in the context of clinical data. The performance evaluations were nonclinical (bench testing).
    • Data Provenance: The data provenance for the nonclinical testing is internal to NuVasive, as they performed the tests (e.g., "Nonclinical testing was performed to demonstrate that the subject CoRoent Ti-C System..."). It's not clinical data, so country of origin or retrospective/prospective does not apply.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable. The performance data is based on nonclinical, laboratory testing against engineering standards, not on expert-adjudicated clinical cases.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable. There was no clinical test set requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is an intervertebral body fusion system, not an AI-powered diagnostic or imaging device. No MRMC study was performed.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is a physical implant, not an algorithm or software.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this device's performance is adherence to established ASTM (American Society for Testing and Materials) standards for mechanical and material properties, and ANSI/AAMI standards for sterility. It's based on engineering and material science principles, not clinical outcomes or expert consensus.

    8. The sample size for the training set:

    Not applicable. This document refers to a physical medical device, not a machine learning algorithm that requires a "training set."

    9. How the ground truth for the training set was established:

    Not applicable for the same reason as point 8.