Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K160602

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-06-23

    (113 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The JustRight Surgical® Vessel Sealing System is indicated for use in open and laparoscopic general surgical procedures to seal blood vessels and vascular bundles up to and including 5 mm in diameter for use in adult and pediatric populations, wherever vessel ligation is required.

    The device is contraindicated for use in ENT procedures.

    Device Description

    The JustRight Surgical® Vessel Sealing System consists of the JustRight™ Generator and a JustRight™ Sealer. The JustRight™ Generator is designed to provide low power bipolar RF energy for vessel-sealing applications. The JustRight™ Generator is for use only with JustRight™ Sealer instruments. The JustRight™ Sealer is a hand held ring handle design with either a 10 cm or a 20 cm shaft length that is compatible with a 3mm (ID) cannula. The JustRight™ Sealer attaches to the generator via a 10' cord.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the JustRight Surgical® Vessel Sealing System. It focuses on demonstrating substantial equivalence to predicate devices, particularly for an expanded indication to include pediatric use.

    Here's an analysis of the acceptance criteria and the study information based on the provided document.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with specific numerical targets. Instead, it relies on demonstrating equivalence through various tests and a literature review. The performance claims are comparative to predicate devices.

    Acceptance Criterion (Implied)Reported Device Performance
    Bench Testing - Device Function and DurabilityConfirmed device performance and characteristics.
    Electrical Safety (in accordance with IEC 60601-1-1)Conducted and met standards.
    Electromagnetic Compatibility (EMC) (in accordance with IEC 60601-1-2)Conducted and met standards.
    Electrosurgical Generator Safety (in accordance with IEC 60601-2-2)Conducted and met standards.
    Biocompatibility (in accordance with ISO 10993)Conducted and met standards.
    Vessel Sealing Performance (in vivo) (equivalent to predicate device in pediatric and adult populations)Animal testing confirmed device performance to be equivalent to the predicate device. Literature review and post-market feedback support safety and efficacy for pediatric indication.

    2. Sample Size for the Test Set and Data Provenance

    • Animal Testing: The document states "Animal testing... conducted," but does not specify the sample size (number of animals or number of seals) used for the in vivo evaluation.
    • Data Provenance:
      • Bench Testing: Internal JustRight Surgical LLC.
      • Animal Testing: Not explicitly stated, but typically conducted in a laboratory setting.
      • Literature Review: Publicly available medical literature, including meta-analyses, focusing on complication rates in pediatric vs. adult populations using vessel sealers.
      • Post-Market Feedback: Data collected on the existing JustRight Surgical® Vessel Sealing System.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This type of information is not applicable or not provided in the context of this 510(k) submission. The document describes engineering and animal studies, not human clinical trials or diagnostic AI studies requiring expert ground truth establishment for a test set.

    4. Adjudication Method for the Test Set

    This is not applicable as the studies are technical and animal tests, not clinical or diagnostic studies requiring expert adjudication of results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The document focuses on device performance and equivalence to predicate devices, not on human reader performance with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable as the device is a physical surgical instrument (vessel sealing system), not an AI algorithm.

    7. The Type of Ground Truth Used

    • Bench Testing: Engineering specifications and standardized test methods (e.g., electrical safety standards, force/pressure measurements for seals, durability cycles).
    • Animal Testing: Direct observation of vessel sealing efficacy in an in vivo model, likely based on burst pressure and histological assessment of sealed vessels, comparing outcomes to the predicate device.
    • Literature Review: Published clinical outcomes and complication rates for vessel sealing devices.
    • Post-Market Feedback: Real-world data reported by users of the device.

    8. The Sample Size for the Training Set

    This is not applicable. The device is a hardware system, not an AI algorithm that requires a "training set." The development of the device would involve engineering design and testing, not machine learning model training.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1