510(k) Data Aggregation

    K Number
    K160363

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-10-20

    (254 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Optimus CMF System is implantable bone plates and bone screws for maxillofacial and mandible surgery procedures including:

    1. Fractures
    2. Osteotomies
    3. Reconstructive procedures
    4. Revision procedures where other treatments or devices have failed.
    Device Description

    The system is comprised of plate and screw. The range of subject device plate's sizes is from 0.6 to 1.5mm thick. It is made of unalloyed Titanium (ASTM F67) and anodized in 4 colors (silver, blue, green and gold). The range of screw's diameters is from 1.3 to 2.0mm in lengths of 6.0 to 12mm. The screws are made of Ti-6AI-4V ELI titanium alloy (ASTM F136) and anodized in 3 colors (light blue, silver and gold).

    The device is packed separately in a PE bag.The plate and screw aresingle use only, nonsterile products. The devices must be sterilized before use.

    The screws are anodized in different colors to prevent confusion. The colors are selected for each model by diameter and so on. The devices are used as a pair (O-plate & O-screw, Fplate & F-screw) but sold individually (Not a kit or a set).

    The IMF screw is for maxillomandibular fixation. The head has a relief groove in which wire or elastic bands can be wrapped around the screws which are temporarily implanted in the maxilla and mandible. It is intended for use as a bone screw in temporary fixation of the maxilla and mandible, providing indirect stabilization of fractures of the maxilla and/or the mandible. It is used independently, not with the plates.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the "Optimus CMF System," which consists of implantable bone plates and bone screws for maxillofacial and mandible surgery. As such, it primarily focuses on demonstrating substantial equivalence to previously cleared predicate devices rather than proving a device meets specific clinical performance acceptance criteria in the way an AI/ML device would.

    Therefore, the requested information elements related to AI/ML device studies (such as sample size for test/training sets, experts for ground truth, adjudication methods, MRMC studies, standalone performance, etc.) are not applicable to this submission. This document describes a physical medical implant, not a diagnostic or interpretative AI/ML system.

    However, I can extract information related to the device's performance testing as a substitute for "acceptance criteria and reported performance" in the context of this device type.

    1. A table of acceptance criteria and the reported device performance

    The document describes non-clinical tests performed to evaluate the performance of the Optimus CMF System compared to predicate devices. The "acceptance criteria" here implicitly refer to demonstrating comparable or superior mechanical strength to legally marketed predicate devices, thereby establishing substantial equivalence.

    Acceptance Criteria Category (Implicit)Specific Test TypePerformance Metric EvaluatedReported Device Performance
    Mechanical Strength4 Point Bending Test (ASTM F382-14)Yield strength, bending structural stiffness, bending strength"The test results of the subject device were overall higher than the predicate devices supporting that the subject device is substantially equivalent to the predicate devices."
    Torsional Strength Test (ASTM F543-13)Torsional strength"The test results of the subject device were overall higher than the predicate devices supporting that the subject device is substantially equivalent to the predicate devices."
    Pullout Force Test (ASTM F543-13)Pullout force"The test results of the subject device were overall higher than the predicate devices supporting that the subject device is substantially equivalent to the predicate devices."
    BiocompatibilityBiocompatibility TestingCompatibility with biological systems"Validation of sterilization parameters and biocompatibility of the submission device are supported by sterilization validation and biocompatibility testing as provided in the primary predicate K140037."
    SterilizationSterilization ValidationEfficacy of sterilization process"Validation of sterilization parameters and biocompatibility of the submission device are supported by sterilization validation and biocompatibility testing as provided in the primary predicate K140037."

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size: The document does not explicitly state the number of samples (plates/screws) used for each mechanical test. Standard ASTM testing methods usually specify a minimum number of samples, but this information is not detailed in this summary.
    • Data Provenance: The tests were performed to support a submission by Osteonic Co., Ltd. (located in Seoul, Korea) to the FDA in the USA. The specific location where the testing was physically conducted is not mentioned. These are prospective engineering tests performed on the manufactured devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable, as this is a non-AI physical device. Ground truth, in the context of these mechanical tests, is established by measurement instruments according to standardized ASTM protocols.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable, as this is a non-AI physical device.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable, as this is a non-AI physical device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable, as this is a non-AI physical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the mechanical tests consists of objective measurements (e.g., force, displacement) obtained from standardized engineering tests (ASTM F382-14, ASTM F543-13) using calibrated equipment. For biocompatibility and sterilization, the ground truth is established by adherence to recognized standards and validated reports, often involving laboratory analyses.

    8. The sample size for the training set

    Not applicable, as this is a non-AI physical device and does not involve a training set.

    9. How the ground truth for the training set was established

    Not applicable, as this is a non-AI physical device and does not involve a training set.