510(k) Data Aggregation

    K Number
    K160326

    Validate with FDA (Live)

    Date Cleared
    2016-02-23

    (18 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    5MP Monochrome LCD Monitor MS55i2 (ML21055, MD211G5) and 5MP Color LCD Monitor CCL550i2 (CL21550) are intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners. They are to be used in digital mammography PACS, modalities including FFDM, and breast tomosynthensis

    Device Description

    MS55i2 (ML21055, MD211G5) is a 21.3 inch (54 cm) monochrome LCD monitor and CCL550i2 (CL21550) is a 21.3 inch (54 cm) color LCD monitor. Their displays' resolution is 2048 x 2560 (landscape), 2560 x 2048 (portrait) supporting DVI (digital visual interface) and DisplayPort.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the JVC Kenwood medical monitors, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA 510(k) summary for the JVC KENWOOD monitors (MS55i2 and CCL550i2) primarily focuses on demonstrating substantial equivalence to a predicate device (EIZO RadiForce GX540) rather than presenting specific, quantitative acceptance criteria and results. The "acceptance criteria" can be inferred as matching or being equivalent to the key performance specifications of the predicate device, especially those related to image quality and functionality for diagnostic medical imaging.

    Feature / Acceptance Criteria (Inferred from Predicate Equivalence)Predicate Device (EIZO RadiForce GX540) SpecificationJVC KENWOOD MS55i2 (ML21055, MD211G5) PerformanceJVC KENWOOD CCL550i2 (CL21550) Performance
    Response Time (typical)25ms (On/Off)25ms (On/Off)25ms (On/Off)
    Resolution or Matrix Size5MP (2048 x 2560)5MP (2048 x 2560)5MP (2048 x 2560)
    Screen TechnologyTFT Monochrome LCD Panel (IPS)TFT Monochrome LCD Panel (IPS)TFT Color LCD Panel (IPS)
    BacklightingLEDLEDLED
    Maximum LuminanceMin. 850cd/m², Typ. 1200cd/m²Min. 850cd/m², Typ. 1200cd/m²Min. 700cd/m², Typ. 1000cd/m²
    DICOM Calibrated Luminance500cd/m²500cd/m²500cd/m²
    Viewing Angle (CR>50)Horizontal: Typ. 176°, Vertical: Typ. 176°Horizontal: Typ. 176°, Vertical: Typ. 176°Horizontal: Typ. 176°, Vertical: Typ. 176°
    Display AreaHorizontal: 337.92mm, Vertical: 422.4mmHorizontal: 337.92mm, Vertical: 422.4mmHorizontal: 337.92mm, Vertical: 422.4mm
    Aspect Ratio4:54:54:5
    Pixel PitchHorizontal: 0.165mm, Vertical: 0.165mmHorizontal: 0.165mm, Vertical: 0.165mmHorizontal: 0.165mm, Vertical: 0.165mm
    Contrast RatioMin. 900:1, Typ. 1200:1Min. 900:1, Typ. 1200:1Min. 1000:1, Typ. 1300:1
    Grayscale Tones (10-bit DisplayPort)1,024 from a palette of 16,369 tones1,024 from a palette of 12,277 tones1,024 from a palette of 65,473 tones
    Grayscale Tones (8-bit)256 from a palette of 16,369 tones256 from a palette of 12,277 tones256 from a palette of 65,473 tones
    Non-Uniformity CompensationDigital Uniformity Equalizer (DUE)Digital Uniformity correction SystemDigital Uniformity correction System
    Input Video SignalDVI-D x1, DisplayPort x1DVI-D x1, DisplayPort x1DVI-D x1, DisplayPort x1
    QA SoftwareRadiCSQA Medivisor / Medivisor NX, F-CALQA Medivisor / Medivisor NX, F-CAL
    Built-in Front SensorIntegrated Front Sensor (IFS)Built-in Front SensorBuilt-in Front Sensor
    Ambient Light SensorAmbient Light Sensor (ALS)Built-in ambient Light SensorBuilt-in ambient Light Sensor

    Explanation of Differences and Performance Justification:

    • Grayscale Tones: While the palette size for grayscale tones differs, the document states, "But our monitors pass the exams in AAPM-TG18 4.3 'Luminance Response'. Our monitors are equivalent to the predicated device." This indicates that despite the numerical difference in palette size, the devices meet the relevant image quality standards for grayscale reproduction.
    • Non-Uniformity Compensation: The JVC KENWOOD system is considered "equivalent to the predicated device's one" in making visual performance "much better in the point of luminance (monochrome and color version) and chromaticity."
    • Maximum Luminance (CCL550i2): The color monitor has a lower maximum luminance, but the document clarifies, "However DICOM Calibrated luminance is same as the predicated device. Then actual performance for diagnosis is the same." This is a crucial point, as the calibrated luminance is what's critical for diagnostic viewing.
    • Contrast Ratio (CCL550i2): The color model's contrast ratio is higher, which is generally a positive difference.

    2. Sample Size Used for the Test Set and Data Provenance

    This document does not contain information about a "test set" in the context of clinical images or patient data. The evaluation is focused on the technical performance specifications of the monitors themselves. Therefore, sample sizes for test sets in this sense are not applicable.

    • Data Provenance: The data presented is from the manufacturer's (JVC KENWOOD Corporation) testing and comparison against the specifications of the predicate device. The company is based in Japan. This is more akin to bench testing of hardware specifications rather than a clinical study with patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This is a technical comparison of monitor specifications, not an evaluation of diagnostic accuracy based on expert interpretation of images. The "ground truth" here is the adherence to technical standards and performance specifications, likely established through engineering measurements and adherence to industry standards like AAPM-TG18.

    4. Adjudication Method for the Test Set

    Not applicable, as no human reader studies or clinical test sets requiring adjudication are described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This 510(k) summary is for medical display monitors and does not involve AI or reader studies.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This 510(k) summary is for medical display monitors and does not involve algorithms or AI performance.

    7. The Type of Ground Truth Used

    The "ground truth" used in this submission is primarily technical specifications and performance measurements against established industry standards (e.g., DICOM, AAPM-TG18) and direct comparison with a legally marketed predicate device. The goal is to demonstrate that the new devices perform equivalently or better in terms of visual characteristics critical for medical image diagnosis.

    8. The Sample Size for the Training Set

    Not applicable. There is no AI algorithm being developed or trained described in this document.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set mentioned in this document.