510(k) Data Aggregation

    K Number
    K160271

    Validate with FDA (Live)

    Date Cleared
    2016-07-15

    (164 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The N95 Particulate Respirator and Surgical Mask, Models TN01-12 are single-use, disposable devices, provided non-sterile, and are intended to be worn by operating room personnel or other healthcare workers to protect both the patient and operating room personnel from the transfer of microorganisms, body fluids, and particulate material.

    Device Description

    The N95 Particulate Respirator and Surgical Mask, Models #TN01-11 and #TN01-12 are NIOSH certified (TC 84A-3348), pouchedshaped respirators when worn. The flat-folded masks are composed of four layers of materials consisting of polypropylene and polyethylene (inner layer), polypropylene meltblown (two filter layers), and polypropylene spunbond (outer layer). The masks contain a conformable nose clamp enclosed in a binding tape welding the top edge to conform to the contours of the face. In addition, the masks contain an ultrasonically welded, polyurethane elastic headband not made with natural rubber latex, to secure the masks in place on the wearer. The masks are offered in orange or white.

    AI/ML Overview

    The provided document describes the acceptance criteria and performance of the N95 Particulate Respirator and Surgical Mask, Models TN01-11 and TN01-12, in comparison to a predicate device.

    Here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    CriteriaAcceptance Criteria (Implied by Predicate & Standards)Reported Device Performance (Models TN01-11 and TN01-12)
    Fluid ResistancePass at 160 mmHg (ASTM F1862)Pass at 160 mmHg (ASTM F1862). Models TN01-11 and TN01-12 meet the requirements of ASTM F1862.
    Particulate Filtration Efficiency (PFE)NIOSH Certification (e.g., 84A-5216)NIOSH Certification Number: TC 84A-3348 (includes TN01-11 & TN01-12)
    Bacterial Filtration Efficiency (BFE)NIOSH Certification (e.g., 84A-5216)NIOSH Certification Number: TC 84A-3348 (includes TN01-11 & TN01-12)
    Differential Pressure (△ P)NIOSH Certification (e.g., 84A-5216)NIOSH Certification Number: TC 84A-3348 (includes TN01-11 & TN01-12)
    FlammabilityClass 1 (16 CFR 1610)Class 1 (16 CFR 1610). Meets 16 CFR 1610 Standard for class 1 flammability.
    BiocompatibilityNon-cytotoxic, non-sensitizing, non-irritating (ISO 10993)Under the conditions of the study, the device was non-cytotoxic, non-sensitizing, and non-irritating (ISO 10993).
    - Cytotoxicity(ISO 10993-5)Under the conditions of the study, the subject device was non-cytotoxic (ISO 10993-5).
    - Irritation(ISO 10993-10)Under the conditions of the study, the subject device was non-irritating (ISO 10993-10).
    - Sensitization(ISO 10993-10)Under the conditions of the study, the subject device was non-sensitizing (ISO 10993-10).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the specific sample sizes used for each performance test (e.g., number of masks tested for fluid resistance or flammability). It simply states that the models meet the requirements of the standards. The provenance of the data is not specified beyond indicating it was a submission to the FDA. The tests are described as performance testing and non-clinical tests, which are typically prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not applicable to the type of device and testing described. The "ground truth" for N95 masks is established through standardized laboratory performance tests according to recognized standards (e.g., ASTM, NIOSH, ISO) and not through expert consensus or interpretation of medical images.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. Adjudication methods like 2+1 are typically used in studies where human readers interpret medical data (e.g., radiology studies) and their findings need to be reconciled. For performance testing of a physical device like an N95 mask, objective laboratory measurements are taken against predefined criteria.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The device is a physical N95 mask, not an AI-powered diagnostic tool. Therefore, MRMC studies and "human readers improving with AI" are irrelevant to this document.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable. The device is a physical N95 mask; there is no algorithm or AI component involved. The testing described is for the physical performance of the mask.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the performance of the N95 mask is based on objective measurements against established performance standards set by regulatory bodies and testing organizations. These include:

    • ASTM F1862 for Fluid Resistance
    • NIOSH Certification (TC 84A-3348) for Particulate Filtration Efficiency, Bacterial Filtration Efficiency, and Differential Pressure
    • 16 CFR 1610 for Flammability
    • ISO 10993 series for Biocompatibility (cytotoxicity, irritation, sensitization)

    8. The sample size for the training set

    This information is not applicable. There is no "training set" in the context of performance testing for a physical medical device like an N95 mask. Training sets are relevant for machine learning or AI models.

    9. How the ground truth for the training set was established

    This information is not applicable, as there is no training set for this type of device and study.