Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K160166

    Validate with FDA (Live)

    Device Name
    X-MIND trium
    Manufacturer
    Date Cleared
    2016-11-15

    (293 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    0 - 21
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    X-MIND trium is a digital panoramic, cephalometric and tomographic extra-oral X-ray system, indicated for use in:

    • producing panoramic X-ray images for diagnostic examination of dentition (teeth), jaws and oral structures;
    • producing radiographs of maxillofacial region and parts of the skull for cephalometric examination, if equipped with CEPH arm;
    • producing radiographs of hands and wrists for carpus examination, if equipped with CEPH arm;
    • producing tomographic images of the oral and maxillofacial region, for diagnostic examination of dentition (teeth), jaws, oral structures and some cranial bones, if equipped with CBCT option.

    From a clinical point of view, X-MIND trium can be applied for the following medical indications :

    • Generic dentistry
    • Dental implantology
    • Dental surgery
    • Maxillo-facial surgery
    • Cephalometric analysis
    • Carpus radiology
    Device Description

    X-MIND trium is a digital panoramic, cephalometric and tomographic extra-oral Xray system, which consists of:

    1. Operator's workstation
    2. X-MIND trium remote control and X-MIND trium light
    3. X-MIND Trium medical device (Control panel, X-ray generator, Sliding body, U-arm, Detector sliding group, CBCT detector, PAN detector, Patient support, F group, Column)
      Cephalometric extension (CEPH arm extension, CEPH control panel, CEPH patient support, CEPH detector sliding group, CEPH secondary collimator, CEPH detector)
      Accessories (Temple rest, Chin rest, Sliding bite block and chin rest assembly, TMJ nose support, Calibration tray and geometric calibration phantom, Quality phantom support for Germany)
    AI/ML Overview

    The provided text is a 510(k) summary for the X-MIND trium device, which is a digital panoramic, cephalometric, and tomographic extra-oral X-ray system. The document compares the proposed device to a predicate device, the Hyperion X9, to establish substantial equivalence.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance:

    The document implicitly uses the performance of the predicate device, Hyperion X9, as the acceptance criteria for substantial equivalence, particularly for image quality and radiation dose, while also assessing compliance with relevant safety standards. The table below summarizes the key performance comparisons presented as part of the substantial equivalence claim. Since specific numerical acceptance criteria (e.g., "must meet X value") are not explicitly stated as distinct criteria but rather as comparison outcomes, this table presents the comparative performance as reported.

    CharacteristicAcceptance Criteria (Predicate Device - Hyperion X9)Reported Device Performance (X-MIND trium)Outcome
    Cephalometric Image QualityTest pattern images of certain qualityTest pattern images of equivalent qualitySubstantial equivalence confirmed by metrics
    Cephalometric Radiation DoseUnder DRL for Dental lateral cephalogram adult (32 mGycm²) according to NCRP Report No. 172 (2016)Under DRL for Dental lateral cephalogram adult (32 mGycm²) according to NCRP Report No. 172 (2016)Both Hyperion and X-MIND trium are under the DRL
    Panoramic Image QualityTest pattern images of certain qualityTest pattern images of equivalent qualitySubstantial equivalence confirmed by metrics
    Panoramic Radiation DoseUnder DRL for Dental panoramic (100 mGycm²) according to NCRP Report No. 172 (2016)Under DRL for Dental panoramic (100 mGycm²) according to NCRP Report No. 172 (2016)Both Hyperion and X-MIND trium are under the DRL
    CBCT Image Quality (Standard Quality Protocols)Homogeneity: baseline as Hyperion X9Homogeneity: 2.8-10.9 times lower than Hyperion X9Homogeneity for Hyperion is 2.8-10.9 times higher than X-MIND trium
    CNR: baseline as Hyperion X9CNR: 2.12 ÷ 2.37 times higher than Hyperion X9CNR for Trium is 2.12 ÷ 2.37 times higher than the CNR for Hyperion
    MTF: baseline as Hyperion X9MTF plots show Trium better maintains high frequencies between 1.25 lp/mm and 1.5 lp/mmTrium better maintains high frequencies
    Nyquist frequency: baseline as Hyperion X9EquivalentEquivalent
    Absence of artifacts: baseline as Hyperion X9EquivalentEquivalent
    CBCT Radiation Dose (Standard Quality Protocols)Similar to Hyperion X9, within reasonable variationLower than Hyperion X9, but X-MIND trium dose can be reduced by using other scan protocols (implicitly aiming for equivalence or improvement)Hyperion is lower but the dose for X-MIND trium can be reduced by using other scan protocols (suggesting overall acceptable dose, though initially higher for some protocols if not adjusted)
    CBCT System Indicator ($S_{DAP}$)Substantial equivalence to Hyperion X9Calculated $S_{DAP}$ value (not explicitly reported but claimed to establish equivalence)Substantial equivalence is confirmed by $S_{DAP}$
    CBCT Image Quality (High Quality Protocols)Noise: baseline as Hyperion X9Voxel noise for Hyperion is 43.5% ÷ 53.8% higher than TriumVoxel noise for Hyperion is higher than Trium (implying Trium has lower noise)
    NPS: baseline as Hyperion X9Noise content very similar at low frequencies, Hyperion shows higher noise at medium and high frequencyHyperion shows higher noise at medium and high frequencies (implying Trium has lower noise)
    CBCT Radiation Dose (High Quality Protocols)Substantial equivalence to Hyperion X9Metrics confirmed substantial equivalenceSubstantial equivalence is confirmed by metrics
    Electrical Safety & EMCCompliance with IEC 60601-1 3rd ed., IEC 60601-1-2: 2007, etc. (as per predicate)Compliance with same standards, plus US and Canada deviations (IEC 60601-1 deviations for US and Canada), IEC 60601-1: 2007, ANSI/AAMI ES60601-1: 2005, CAN/CSA C22.2 No. 60601-1:08, IEC 60601-1-3: 2008, IEC 60601-2-63: 2012, IEC 60601-1-6: 2010, IEC 62366: 2007Met and exceeded standards, with formal certifications from NRTLs.
    Software Verification & ValidationAs recommended by FDA GuidanceDocumentation provided as recommended by FDA Guidance for software in medical devicesSoftware considered "moderate" level of concern, V&V testing conducted. General tools and means made available by both software are equivalent.

    2. Sample size used for the test set and the data provenance:

    The document mentions that "bench tests" were performed for image quality and radiation dose. It refers to "test pattern images" and "clinical images obtained with X-MIND trium and predicate device." However, the sample size for the test set (number of images/cases) is not explicitly stated.

    The data provenance is likely retrospective, as the comparison involves existing images from the predicate device and newly generated images from the X-MIND trium under controlled bench test conditions. The country of origin of the data is not specified but given the manufacturer's location in Italy and the use of a Nationally Recognized Testing Laboratory (NRTL) (Intertek Semko, Nemko S.p.A.), it is likely that the testing was performed in Europe.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    The document does not mention the use of human experts to establish ground truth for the test set. The image quality assessments appear to be based on objective metrics derived from phantom studies (e.g., homogeneity, CNR, MTF, Nyquist frequency, artifacts, NPS), rather than subjective evaluations by clinical experts.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable, as human expert evaluation for ground truth determination is not described. The assessment was based on objective metrics from bench tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. This device is an X-ray imaging system, not an AI-assisted diagnostic tool. The focus of the submission is on hardware and software equivalence for image acquisition and display, not on reader performance improvement with AI.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Yes, standalone performance was done, but not in the context of an "algorithm" as typically conceived in AI/ML performance. The "device performance" described (image quality metrics, radiation dose, electrical safety, EMC) is the performance of the X-MIND trium system itself in a standalone capacity, i.e., without human intervention in the image generation part of the test. The "software" component also underwent verification and validation, but this refers to its functionality and reliability in image management and processing features available to the user, not an AI algorithm performing diagnostic tasks.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    The "ground truth" for the performance evaluation mostly relies on objective physical measurements and phantom studies. For image quality, metrics like homogeneity, CNR, MTF, Nyquist frequency, and NPS derived from test patterns or phantoms serve as the "truth" against which the device's imaging capabilities are measured. For radiation dose, established Dosimetry Reference Levels (DRLs) from NCRP REPORT No. 172 (2016) serve as the standard. For electrical safety and EMC, recognized international standards (e.g., IEC 60601 series) are the ground truth for compliance.

    8. The sample size for the training set:

    Not applicable. The device is an X-ray imaging system, not an AI/ML diagnostic algorithm that requires a training set of data.

    9. How the ground truth for the training set was established:

    Not applicable. As stated above, this device does not involve a training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1