510(k) Data Aggregation

    K Number
    K160164

    Validate with FDA (Live)

    Date Cleared
    2016-02-25

    (29 days)

    Product Code
    Regulation Number
    866.3328
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BD Veritor™ System for Rapid Detection of Flu A+B is a rapid chromatographic immunoassay for the direct and qualitative detection of influenza A and B viral nucleoprotein antigens from nasopharyngeal wash, aspirate, and swab in transport media samples from symptomatic patients. The BD Veritor System for Rapid Detection of Flu A+B is a differentiated test, such that influenza A viral antigens can be distinguished from influenza B viral a single processed sample using a single device. The test is to be used as an aid in the diagnosis of influenza A and B viral infections. A negative test is presumptive and it is recommended that these results be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay. Outside the U.S. a negative test is presumptive and it is recommended that these results be confirmed by viral culture or a molecular assay cleared for diagnostic use in the country of use. FDA has not cleared this device outside the U.S. Negative test results do not preclude influenza viral infection and should not be used as the sole basis for treatment or other management decisions. The test is not intended to detect influenza C antigens.

    Device Description

    The BD Veritor™ Flu A+B test is an immunochromatographic assay for the qualitative detection of influenza A and B viral antigens in respiratory specimens. The patient specimen is added to a reaction tube prefilled with RV Reagent C, gently mixed, and then added to the test device. RV Reagent C contains mucolytic agents that function to break down mucus in a patient specimen thereby exposing viral antigens and enhancing detection on the assay device. Processed specimens are expressed through a filter tip into a single sample well on the BD Veritor™ Flu A+B test device.

    After addition to the test device, any influenza A or influenza B viral antigens present in the specimen bind to anti-influenza antibodies coniugated to detector particles on the Veritor'™ Flu A+B test strip. The antigen-conjugate complexes migrate across the test strip to the reaction area and are captured by a line of antibody striped on the membrane. The Veritor'™ Flu A+B test devices are designed with five spatially-distinct zones including positive and negative control line positions, separate test line positions for the target analytes, and a background zone. The test lines for the target analytes are labeled on the test device as 'A' for Flu A position, and 'B' for Flu B position. The onboard positive control ensures the sample has flowed correctly and is indicated on the test device as 'C'. Two of the five distinct zones on the test device are not labeled. These two zones are an onboard negative control line and an assay background zone. The onboard negative control zone addresses non-specific signal generation. The remaining zone is used to measure the assay background.

    The BD Veritor™ Flu A+B assay incorporates an active negative control feature in each test to identify and compensate for sample-related, nonspecific signal generation. The BD Veritor™ System Reader uses a proprietary algorithm that subtracts nonspecific signal at the negative control line from the signal present at both the Flu A and Flu B test lines. If the resultant test line signal is above a pre-selected assay cutoff, the specimen scores as positive. If the resultant test line signal is below the cutoff, the specimen scores as negative. Use of the active negative control feature allows the BD Veritor™ System reader to correctly interpret test results that cannot be scored visually because the human eve is unable to accurately perform the subtraction of the nonspecific signal.

    AI/ML Overview

    The provided document describes the BD Veritor™ System for Rapid Detection of Flu A + B Laboratory Kit (K160164). However, it is a 510(k) summary for a modification to an already marketed device (K120049, K121797, K132256, K132693). The modification solely involves adding strain reactivity data to the labeling for two specific avian influenza strains (A/Northern Pintail/Washington/40964/2014 H5N2 and A/Gyrfalcon/Washington/41088-6/2014 H5N8).

    The document does not contain acceptance criteria for the original device's performance, nor does it present a study demonstrating that the device meets such criteria. Instead, it states that the addition of strain reactivity data for the new strains did not alter the intended use or fundamental scientific technology of the device, and a risk analysis identified no new safety and effectiveness issues. Therefore, the device is considered substantially equivalent to the previously cleared predicate devices.

    As such, I cannot provide the requested information regarding acceptance criteria, device performance, sample sizes, expert qualifications, adjudication methods, MRMC studies, or standalone performance based on this document, as this information pertains to the original device clearance, not this specific modification.

    The document only discusses the change to the labeling and its impact.