510(k) Data Aggregation

    K Number
    K160114

    Validate with FDA (Live)

    Date Cleared
    2016-03-15

    (56 days)

    Product Code
    Regulation Number
    888.3050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Xspan™ Laminoplasty Fixation System is intended for use in the lower cervical and upper thoracic spine (C3-T3) after a laminoplasty has been performed. The Xspan Laminoplasty Fixation System holds or buttresses the allograft in place in order to prevent expulsion of the allograft or impingement of the spinal cord.

    Device Description

    The Xspan™ Laminoplasty Fixation System is a posterior implant system comprised of laminoplasty plates, hinge plates, and mounting screws. Various forms and sizes of these implants are available so that adaptations can be made to take into account the pathology and anatomy of an individual patient. The system components are manufactured from Titanium based alloy in accordance with ASTM F136. The single use only implants are provided sterile [gamma radiation] and should not be reused under any circumstances.

    The system also includes instrumentation to aid with implantation. The instruments are supplied in a tray which is used for instrument sterilization and storage.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Xspan™ Laminoplasty Fixation System, a medical device for spinal surgery. This document primarily focuses on establishing substantial equivalence to predicate devices based on technological characteristics and mechanical performance. It does not contain information typically found in a study demonstrating the device meets specific acceptance criteria related to a diagnostic or AI algorithm's performance.

    Therefore, many of the requested categories cannot be filled as they pertain to studies evaluating the performance of AI/diagnostic devices, which is not the subject of this 510(k) summary. I will answer based on the information available in the document, and indicate "Not Applicable" or "Not Provided" where the information is not present.

    Acceptance Criteria and Study Summary for Xspan™ Laminoplasty Fixation System

    This device is a surgical implant, not a diagnostic or AI-driven device. The "acceptance criteria" here refer to the performance benchmarks established through mechanical testing to demonstrate equivalence to predicate devices, rather than criteria for diagnostic accuracy (e.g., sensitivity, specificity).

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    Mechanical PerformanceEquivalent performance to predicate devices through static and dynamic 4-point bend testing (referenced ASTM F2193 and ASTM F382).The Xspan™ Laminoplasty Fixation System demonstrated equivalent performance to the predicate systems through static and dynamic 4-point bend testing.
    BiocompatibilityConformance to ASTM F136 for Wrought Titanium-6Aluminum-4Vanadium ELI Alloy for Surgical Implant Applications. Material historically proven biocompatible, corrosion-resistant, and non-toxic.Implants are made of titanium alloy conforming to ASTM F136. The material has a long history of use and proven biocompatibility, corrosion-resistance, and non-toxicity.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated as "sample size" in the context of a clinical test set. For mechanical testing, the information on the number of devices or test specimens is not provided.
    • Data Provenance: Not applicable in the context of clinical data. For mechanical testing, the data origin is from internal testing performed by X-spine Systems, Inc.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not applicable. This device is a surgical implant, and the performance evaluations primarily involve mechanical testing and material conformance, not expert-adjudicated ground truth for diagnostic accuracy.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods (e.g., 2+1, 3+1) are for establishing ground truth in diagnostic studies, not for mechanical performance testing of a spinal implant.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study and Effect Size

    Not applicable. This is a medical device (spinal implant), not an AI-driven diagnostic tool. Therefore, an MRMC study comparing human reader performance with and without AI assistance was not conducted.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. This is a medical device, not an AI algorithm.

    7. Type of Ground Truth Used

    • For Mechanical Testing: Engineering specifications and standardized test methods (ASTM F2193 and ASTM F382) served as the "ground truth" or benchmarks for performance. Comparison was made against established performance profiles of predicate devices.
    • For Biocompatibility: Material standards (ASTM F136) and historical evidence of biocompatibility for the specified titanium alloy served as the "ground truth."

    8. Sample Size for the Training Set

    Not applicable. This device does not involve an AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This device does not involve an AI algorithm or a training set.