Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K153742

    Validate with FDA (Live)

    Date Cleared
    2016-08-08

    (224 days)

    Product Code
    Regulation Number
    870.5100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The balloon dilatation catheter is indicated for:

    • balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion
    • balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction
    • balloon dilatation of a stent after implantation (balloon models 2.00 mm - 5.00 mm only)
    Device Description

    The Apollo Balloon Dilatation Catheter is designed to allow easy exchange of the catheter using a standard length guidewire. Balloon diameters range from 2.0mm to 5.0mm. The balloon material is made of a minimally compliant material with a rated burst pressure of 22 atmospheres for Ø2.0- 4.0mm and 20 atmospheres for Ø4.5-5.0mm balloon respectively. The minimally compliant balloon material will allow high pressure dilatation while maintaining precise control of the balloon diameter and length. The proximal shaft of the catheter is composed of a female luer connector bonded to a PTFE coated stainless steel tube. The proximal shaft allows superior proximal pushability with a smooth transition to a distal shaft composed of an outer tube of nylon and a tri-extrusion inner tube with a balloon laser welded to both tubes at the distal tip. Two radiopaque platinum/iridium marker bands are positioned within the balloon shoulders. The inner tube accepts a standard 0.014 inch PTCA guidewire. The guidewire enters the catheter tip and advances coaxially out the distal Rx port, thereby allowing both coaxial guidance and rapid exchange of catheter with a single standard length guidewire. Two marked sections, 5mm in length located on the proximal shaft, indicate catheter position relative to the tip of either a brachial or femoral guiding catheter. The design of this dilatation catheter does not incorporate a lumen for distal dye injections or distal pressure measurements.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Apollo Balloon Dilatation Catheter, a medical device. The information details the device's technical characteristics and performance data, but it does not describe a study that proves the device meets acceptance criteria in the context of an AI/human-in-the-loop performance study.

    Instead, the performance data section summarizes in vitro bench tests for the physical characteristics of the catheter. This is a very different type of study than what would be described for an AI-powered diagnostic device.

    Therefore, many of the requested details, such as the sample size for the test set, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone AI performance, and training set details, are not applicable to this document as it concerns an entirely different type of medical device evaluation.

    However, I can extract the acceptance criteria and the reported device performance for the physical properties of the catheter as described.

    Here's the information based on the provided text, while acknowledging that it does not correspond to an AI performance study:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategoryReported Device Performance
    General Performance (in vitro tests)The test results met all acceptance criteria, and were similar to predicate devices for:
    Dimensional VerificationMet acceptance criteria.
    Balloon Preparation, Deployment, RetractionMet acceptance criteria.
    Balloon Rated Burst PressureMet acceptance criteria. (Stated burst pressure: 22 atmospheres for Ø2.0-4.0mm and 20 atmospheres for Ø4.5-5.0mm balloons)
    Balloon FatigueMet acceptance criteria.
    Balloon ComplianceMet acceptance criteria.
    Balloon Inflation and Deflation TimeMet acceptance criteria.
    Catheter Bond StrengthMet acceptance criteria.
    Tip Pull StrengthMet acceptance criteria.
    Flexibility and KinkingMet acceptance criteria.
    Torque StrengthMet acceptance criteria.
    Coating IntegrityMet acceptance criteria.
    Particulate EvaluationMet acceptance criteria.
    RadiopacityLeveraged from predicate devices based on the use of same materials and processes as the predicates (K133852; cleared September 19, 2014).
    BiocompatibilityLeveraged from predicate devices based on the use of same materials and processes as the predicates (K133852; cleared September 19, 2014).

    Regarding the remaining points, they are not applicable because this document describes the evaluation of a physical medical device (catheter) through bench testing, not an AI/software-based diagnostic system.

    1. Sample size used for the test set and the data provenance: Not applicable. The "test set" refers to physical samples of the catheter for in vitro bench testing, not a dataset of patient information. The document does not specify the number of catheters tested for each performance characteristic.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth in this context would be engineering specifications and measurements, verified by qualified engineers or lab personnel.
    3. Adjudication method: Not applicable. Bench testing relies on objective measurements against engineering specifications.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is for AI-assisted diagnostic devices.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is for AI algorithms.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): For the physical characteristics, the ground truth would be established engineering specifications, material properties, and physical measurements (e.g., precise dimensions, burst pressures, bond strengths).
    7. The sample size for the training set: Not applicable. This device is evaluated via physical bench testing, not machine learning model training.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1