510(k) Data Aggregation

    K Number
    K153563

    Validate with FDA (Live)

    Device Name
    AirFit F20
    Manufacturer
    Date Cleared
    2016-05-13

    (151 days)

    Product Code
    Regulation Number
    868.5905
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AirFit F20 is a non-invasive accessory used for channeling airflow (with or without supplemental oxygen) to a patient from a positive airway pressure (PAP) device such as a continuous pressure (CPAP) or bilevel system.

    The AirFit F20 is:
    • to be used by patients weighing more than 66 lb (30 kg) for whom positive airway pressure therapy has been prescribed
    • intended for single-patient reuse in the home environment and multi-patient reuse in the hospital/institutional environment.

    Device Description

    The AirFit F20 is an externally placed vented mask covering the mouth and the nose of the patient. It provides a seal such that pressure from a positive pressure source is directed to the patient's nose and/or mouth. The mask connects via a standard conical connector to a conventional air delivery hose, which in turn connects to the positive pressure source. The mask is held in place with an adjustable headgear that straps the mask to the face. The AirFit F20 mask system comprises four subassemblies: cushion, frame, elbow and headgear. The elbow incorporates the vent array and anti-asphyxia valve safety features. The cushion and headgear are available in various sizes to fit a wide patient population.

    AirFit F20 is a prescription device supplied non-sterile.

    AI/ML Overview

    The provided document is a 510(k) summary for the ResMed AirFit F20, a non-invasive accessory for CPAP/bilevel systems. It focuses on demonstrating substantial equivalence to a predicate device (Quattro Air K123979) rather than presenting a study to prove acceptance criteria for a new, unique device performance. Therefore, many of the requested categories (e.g., sample size for test set, number of experts for ground truth, MRMC study, training set information) are not applicable or not explicitly detailed as one might find in a study specifically designed to establish novel performance claims.

    Here's an analysis based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't provide a direct table of specific acceptance criteria values alongside numerical performance results for the AirFit F20. Instead, it states that the device was designed and tested in accordance with ISO 17510-2: Sleep apnoea breathing therapy - Part 2: Masks and application accessories, and that testing confirmed the new device met "predetermined acceptance criteria" and its performance is "substantially equivalent" to the predicate.

    Acceptance Criteria Category (Derived from ISO 17510-2 and testing)Reported Device Performance
    CO2 RebreathingMet predetermined acceptance criteria; demonstrated performance substantially equivalent to Quattro Air (K123979).
    AAV performanceMet predetermined acceptance criteria for activation/deactivation, inspiratory/expiratory resistance, inadvertent activation/deactivation, and response to extreme humidification; demonstrated performance substantially equivalent to Quattro Air (K123979).
    Total Mask FlowMet predetermined acceptance criteria; demonstrated performance substantially equivalent to Quattro Air (K123979).
    Flow ResistanceMet predetermined acceptance criteria; demonstrated performance substantially equivalent to Quattro Air (K123979).
    Through ImpedanceMet predetermined acceptance criteria; demonstrated performance substantially equivalent to Quattro Air (K123979).
    Mechanical Integrity & PerformanceVerified to simulated normal and reasonable abuse scenarios (home cleaning, multi-patient reuse, transportation, storage); met predetermined acceptance criteria.
    BiocompatibilityMet predetermined acceptance criteria in accordance with ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-12, and ISO 10993-17 for new materials.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated. The testing appears to be primarily bench/non-clinical and biocompatibility. For bench testing of physical characteristics like flow, rebreathing, and resistance, samples would typically consist of a certain number of manufactured units of the AirFit F20. The specific number is not provided.
    • Data Provenance: The testing was non-clinical (bench testing) and focused on material and functional performance. It does not involve human patient data in the context of clinical outcomes. The manufacturer, ResMed Ltd., is based in Australia, and the submission was made to the US FDA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • This question is not applicable in the context of this 510(k) submission. The "ground truth" for the non-clinical tests (e.g., CO2 rebreathing, flow resistance) is established by measurement against predefined engineering specifications and international standards (like ISO 17510-2). These are objective physical measurements, not subjective expert interpretations that require "ground truth" establishment by multiple experts with specific qualifications like radiologists.

    4. Adjudication Method for the Test Set

    • Not applicable. Since the tests are objective physical measurements against standards and predicate performance, there is no need for expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No. The document explicitly states: "Clinical data was not relied upon to demonstrate substantial equivalence to the predicate device. Bench testing demonstrates that the new AirFit F20 device performs in an equivalent manner and is as safe and as effective as the predicate device." MRMC studies apply to imaging or diagnostic devices where human readers interpret results, which is not the case for this device.

    6. Standalone (i.e., algorithm only without human-in-the-loop performance) Performance Study

    • Yes, indirectly. The non-clinical data presented (CO2 rebreathing, AAV performance, total mask flow, flow resistance, through impedance, mechanical integrity, biocompatibility) represents testing of the device's inherent physical and functional performance, essentially "standalone" in a bench setting without human interaction influencing the core functional measurements. However, it's not an "algorithm" in the typical sense for AI, but rather the performance of the physical mask.

    7. The Type of Ground Truth Used

    • The "ground truth" for the non-clinical tests is based on:
      • International Standards: Specifically, ISO 17510-2: Sleep apnoea breathing therapy - Part 2: Masks and application accessories.
      • Engineering Specifications: Predetermined acceptance criteria derived from the standard and design requirements.
      • Predicate Device Performance: Demonstrating substantial equivalence to the performance characteristics of the legally marketed predicate device (Quattro Air (K123979)).

    8. The Sample Size for the Training Set

    • Not applicable. This device is a physical medical device (mask), not an AI/algorithm-based diagnostic system that requires a "training set" in the machine learning sense.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As above, there is no AI training set for this device.