Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K153546

    Validate with FDA (Live)

    Date Cleared
    2016-09-28

    (292 days)

    Product Code
    Regulation Number
    872.3760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fabrication of dentures.

    Device Description

    PINK CAD/CAM disc are blanks used in different milling machines (dental CAD-CAM systems) by professional dental technicians. They are composed of hot cured polymethyl methacrylate (PMMA) and pigments. Device is available in different pink shades. Discs of all shades are available in different dimensions (diameter, thickness and profile margin).

    AI/ML Overview

    The provided document describes the 510(k) premarket notification for the "PINK CAD/CAM Disc" for fabrication of dentures and compares it to a predicate device ("M-PM-Disc (Pink)") and a reference device ("PMMA CAD/CAM disc").

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Standard Requirement)Reported PINK CAD/CAM Disc PerformancePredicate Device (M-PM-Disc Pink) PerformanceReference Device (PMMA CAD/CAM disc) Performance
    E-modulus (MPa) (Minimum 2000 MPa per EN ISO 20795-1)3040 MPa2678 MPa2771 MPa
    Flexural Strength (MPa) (Minimum 65 MPa per EN ISO 20795-1)115.8 MPa90 MPa114 MPa
    Residual Methyl Methacrylate (%) (Maximum 2.2 wt-% per EN ISO 20795-1)1.02 wt-%1.14 wt-% (from literature)1.02 wt-%
    Biocompatibility: Cytotoxicity (EN ISO 10993-5)Met requirementsMet requirementsMet requirements
    Biocompatibility: Irritation (EN ISO 10993-10)Met requirementsTest not performedMet requirements
    Biocompatibility: Sensitization (EN ISO 10993-10)Met requirementsMet requirementsMet requirements

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: The document does not explicitly state the sample size for each test. For the E-modulus, flexural strength, and residual methyl methacrylate, it refers to "The test results... have been obtained with PINK CAD/CAM disc" and "The performance characteristics have been determinat_ed_ with PINK CAD/CAM disc," implying a test set was used, but the number of samples is not specified. For biocompatibility tests, it mentions that "testing was conducted on reference device PMMA CAD/CAM disc composed of Polident PMMA CAD/CAM disc material," again without specifying the sample size.
    • Data Provenance: The data appears to be prospective as the tests were "conducted" for the purpose of this 510(k) submission. The country of origin for the data generation is not explicitly stated, but the manufacturer (Polident d.o.o.) is based in Slovenia, suggesting the studies were conducted or commissioned there.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • This information is not applicable to this type of device submission. The "ground truth" for the performance characteristics of dental materials like E-modulus, flexural strength, and residual monomer content is established by standardized laboratory testing methods (e.g., EN ISO 20795-1) and objective measurements, not by expert consensus or interpretation of subjective data. Similarly, biocompatibility studies (EN ISO 10993 series) involve standardized biological assays and observation, not expert ground-truthing in the sense of clinical image analysis.

    4. Adjudication Method for the Test Set:

    • This information is not applicable for the reasons stated above. Adjudication methods like "2+1" or "3+1" are typically used in clinical or diagnostic studies where subjective interpretation is involved and consensus among experts is required to establish a ground truth. The tests performed for this device are quantitative and objective.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is not applicable. The device "PINK CAD/CAM disc" is a material for fabricating dentures, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study related to AI assistance is irrelevant to this submission.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • This is not applicable. The device is a physical material, not an algorithm.

    7. The Type of Ground Truth Used:

    • The "ground truth" for the device's performance is established through standardized laboratory testing using recognized international standards (EN ISO 20795-1 for mechanical and chemical properties, EN ISO 10993 series for biocompatibility). These standards define the test methodologies and acceptable limits, providing an objective "ground truth" for material properties and biological safety.

    8. The Sample Size for the Training Set:

    • This is not applicable. The device is a physical material, not a machine learning model or algorithm, so there is no concept of a "training set" for its development or evaluation in this context. The manufacturing process is controlled, and the performance is verified through testing of representative samples.

    9. How the Ground Truth for the Training Set Was Established:

    • This is not applicable as there is no training set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1