Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K153498

    Validate with FDA (Live)

    Device Name
    Operio
    Manufacturer
    Date Cleared
    2016-07-20

    (226 days)

    Product Code
    Regulation Number
    878.5070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Operio is a portable device for use in a surgical operating room that produces a directed, non-turbulent flow of air to the surgical site during ophthalmic surgery and to the sterile instruments used during surgery. The air flow from the device is HEPA-filtered to reduce the presence of particulate matter to reduce the microorganisms to a level of < 5 CFU per m3 at the surgical site and instruments.

    The air flow is intended to be directed parallel to the surgical site and/or instruments within: 20" (50 cm) in width, 47" (120 cm) in length and 15" (40 cm) in height. Device effectiveness may not be reliably detectable at a distance of 47 inches (120 cm) from the air flow outlet, and effectiveness depreciates beyond this specified area.

    Device Description

    The air zone unit Operio is used in a surgical operating room for cleaning the air in a given area by re-circulating the ambient air and cleaning it from airborne particles. The air is filtered through a HEPA filter and the air is generated over the areas where the demands of clean air are especially high. The air zone unit Operio is equipped with a control panel for adjustments and for positioning of the unit.

    The air zone unit Operio is a mobile unit that with the help of castors easily can be moved around the patient to an optimal position or, when needed, transported inside the hospital between wards. It is also equipped with an optional instrument tray.

    By using a unique sterile shield as a protective barrier the air zone unit can be placed close to the OR table and deliver HEPA filtered air to the surgical site and instruments to reduce the presence of airborne particulate and microorganisms.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study details for the Operio device, based on the provided document:

    Acceptance Criteria and Device Performance for Operio

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaDevice Performance (Operio)
    Microorganisms at Surgical Site & Instruments: < 5 CFU per m³0.4 CFU/m³ (clinical environment, mean value)
    Particulate Matter (0.5 µm/f³): Reduction in comparison to predicate device1.1 particles of 0.5 µm/f³ (compared to 68,122 particles of 0.5 µm/f³ for predicate)
    HEPA Filtration Efficiency: At least 99.97% against 0.3 µm particlesAt least 99.995% against 0.3 µm particles
    Airflow Coverage (Width): 20" (50 cm)Not explicitly stated if tested independently, but the indication mentions this as the intended directed flow area.
    Airflow Coverage (Length): 47" (120 cm)Not explicitly stated if tested independently, but the indication mentions this as the intended directed flow area, with effectiveness depreciating beyond.
    Airflow Coverage (Height): 15" (40 cm)Not explicitly stated if tested independently, but the indication mentions this as the intended directed flow area.

    2. Sample Sizes and Data Provenance

    • Test Set (Clinical): 31 samples from 23 patients.
    • Data Provenance: Clinical tests were carried out in a Swedish University Hospital. The data is prospective as it's stated "measurements were made near the surgical site and over the instruments."

    3. Number of Experts and their Qualifications for Ground Truth

    The document does not explicitly state the number of experts used to establish ground truth or their qualifications. The clinical tests were performed in a "Swedish University Hospital," suggesting medical professionals were involved in the procedures and data collection, but their specific roles in ground truth establishment for the microbial and particulate counts are not detailed.

    4. Adjudication Method for the Test Set

    The document does not specify an adjudication method for the clinical test set (e.g., 2+1, 3+1, none). The CFU and particulate counting results are presented as measured values.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The study focuses on the device's performance in reducing airborne contamination, not on evaluating human reader improvement with or without AI assistance. This device is an air-handling apparatus, not an AI diagnostic tool.

    6. Standalone Performance Study (Algorithm Only)

    Yes, a standalone performance study was done for the device itself. The clinical and non-clinical tests evaluate the device's capability to clean the air and reduce microorganisms and particulates without human intervention in the purification process itself. The "algorithm" here refers to the device's mechanical and filtration processes.

    7. Type of Ground Truth Used

    The ground truth used was based on direct measurements of:

    • Colony Forming Units (CFU): Indicative of microbial load.
    • Particulate counts: Indicative of particulate matter.
      These are direct, empirical measurements taken from the environment.

    8. Sample Size for the Training Set

    The document does not mention a training set as this is a physical device, not a machine learning algorithm that requires a "training set" in the conventional sense. The "training" of the device is inherent in its engineering and design, validated by the tests described.

    9. How Ground Truth for the Training Set was Established

    Not applicable, as this device does not utilize a training set in the context of machine learning. The "ground truth" for the device's design and operational parameters would be based on engineering specifications, regulatory standards for air quality, and previous research on surgical site infection prevention.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1