510(k) Data Aggregation

    K Number
    K153148

    Validate with FDA (Live)

    Date Cleared
    2015-12-01

    (32 days)

    Product Code
    Regulation Number
    866.3980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Influenza A/H5 Subtyping Kit contains reagents and controls of the CDC Human Influenza Virus RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an Applied Biosystems (ABI) 7500 Fast Dx Real-Time PCR instrument in conjunction with clinical and epidemiological information:

    • For the presumptive identification of virus in patients who may be infected with influenza A subtype A/HS (Asian lineage) from viral RNA in human respiratory specimens and viral culture in conjunction with clinical and epidemiological risk factors;

    · To provide epidemiologic information for surveillance of circulating influenza viruses.

    Performance characteristics for influenza were established during a seasonal influenza viruses A/H1 and A/ H3 were the predominant influenza A viruses in circulation and during a season when the A/H1pdm09 influenza virus was the predominant influenza A virus in circulation. Performance characteristics may vary with other emerging influenza A viruses.

    Testing with the influenza H5a and H5b primer and probe sets should not be performed unless the most current U.S. Department of Health and Human Services (DHHS) clinical and epidemiologic criteria for testing suspect A/ H5 specimens. The definitive identification of influenza A/H5 (Asian lineage) either directly from patient or from virus cultures requires additional laboratory testing, along with clinical and epidemiological assessment in consultation with national influenza surveillance experts.

    Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out bacterial infection with other viruses. The agent detected may not be the definite cause of disease.

    If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to a state or local health department for testing. Viral culture should not be attempted unless a BSL 3+ facility is available to receive and culture specimens.

    All users, and any person reporting results from use of this device should be trained to perform and interpret the results from this procedure by a competent instructor prior to use. CDC Influenza Division will limit the distribution of this device to only those users who have successfully completed a training course provided by CDC instructors or designees.

    Device Description

    The CDC Human Influenza Real-Time RT-PCR Diagnostic Panel Influenza A/H5 Subtyping Kit is a real-time RT-PCR) assay that utilizes the Applied Biosystems® (ABI) 7500 Fast Dx Realtime PCR system. The Influenza A/H5 Subtyping Kit consists of oligonucleotide primers, fluorescently labeled hydrolysis probes, and controls which are used in rRT-PCR assays for the in vitro qualitative detection and characterization of influenza virus RNA in respiratory specimens from patients presenting with influenza-like illness (ILI). The oligonucleotide primers and probes for detection of influenza A viruses (InfA) were selected from highly conserved regions of the matrix (M) protein. Oligonucleotide primers and probes for characterization of avian influenza A/H5 (Asian lineage) viruses (H5a and H5b) were selected from highly conserved regions of their HA genes. The Influenza A/H5 Subtyping kit also contains primers and probes to detect the human RNase P gene (RP) in control samples and clinical specimens.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and the study that proves the device meets the acceptance criteria:

    Device Name: CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel, Influenza A/H5 Subtyping Kit (VER 3)
    K Number: K153148


    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes the analytical and clinical performance studies, but directly stated "acceptance criteria" for each metric are not explicitly listed in a formatted table with pass/fail thresholds. Instead, the results are presented as the device's performance, implying these values met internal criteria for submission and approval. Inferring from the nature of diagnostic device validation, these would typically be high percentages for sensitivity and specificity.

    Based on the provided data, here's a table summarizing the observed performance:

    Performance MetricAcceptance Criteria (Inferred)Reported Device Performance
    Analytical Sensitivity (LOD)Specific LOD values for tested H5N1 and H5N8 strains. For example, the lowest concentration where ≥95% of replicates test positive.- A/Vietnam/1203/2004 x A/Puerto Rico/8/34 reassortant (A/H5N1): - Invitrogen SuperScript™: 10^3.8 EID50/mL - Quanta qScript™: 10^2.4 EID50/mL - A/duck/Vietnam/NCVD-1544/2012 (A/H5N1): - Invitrogen SuperScript™: 10^3.1 EID50/mL - Quanta qScript™: 10^3.1 EID50/mL - A/gyrfalcon/Washington/41088-6/2014 (A/H5N8): - Invitrogen SuperScript™: 10^3.35 EID50/mL - Quanta qScript™: 10^3.35 EID50/mL Reported as the lowest concentration where InfA and both H5a/H5b showed uniform detection.
    Analytical InclusivityDetection of all listed A/H5 (Asian lineage) isolates representative of different geographic locations and phylogenetic clades, at or near the established LOD.- Reactive with all 15 tested H5 (Asian lineage) isolates (all 3/3 positive for InfA, H5a, and H5b) with both enzyme systems. - Predicted to be reactive with A/Bangladesh/3222/2011 (in silico analysis).
    Analytical Specificity (Cross-Reactivity)No detection of common human influenza A (non-H5), human influenza B, swine influenza, avian influenza (non-H5), canine influenza, or equine influenza viruses when tested at high titers.- InfA assay detected all tested human influenza A and various animal influenza A strains (as expected for an InfA primer/probe set). - H5a and H5b assays showed no cross-reactivity with human influenza A (non-H5), human influenza B, most animal influenza A (non-H5), and specifically did not detect A/duck/Singapore-Q/F119-3/1997 (H5N3) when tested with Invitrogen SuperScript™. (Note: The H5a/H5b assays did detect A/duck/Singapore-Q/F119-3/1997 H5N3, which is an H5 subtype, albeit not specified as Asian lineage. This is an expected positive result for H5 detection.)
    Analytical Specificity (Exclusivity)No detection (no false positives) of common non-influenza human respiratory viruses, respiratory bacteria, and commensal organisms of the human respiratory tract when tested at high concentrations.- No detection (all negative) for all 18 bacteria, 1 yeast, and 16 non-influenza human respiratory viruses tested with Invitrogen SuperScript™ for InfA, H5a, and H5b assays.
    Clinical Performance (Positive Agreement)High Positive Agreement (e.g., >90% with a reasonable lower bound for 95% CI).- Quanta qScript™: 100% (95% CI: 92.9 – 100.00) (50/50) - Invitrogen SuperScript™: 88.0% (95% CI: 76.2 - 94.4) (44/50)
    Clinical Performance (Negative Agreement)High Negative Agreement (e.g., >90% with a reasonable lower bound for 95% CI).- Quanta qScript™: 100% (95% CI: 94.4 – 100.00) (65/65) - Invitrogen SuperScript™: 100% (95% CI: 94.4 - 100.00) (65/65)

    2. Sample Size Used for the Test Set and Data Provenance

    • Analytical Sensitivity (LOD):

      • Test Set Sample Size: 20 replicates for each enzyme kit (Invitrogen SuperScript™ and Quanta qScript™) for the highest virus dilution for each of the three A/H5 strains to confirm LOD. Initial testing used 3 replicates.
      • Data Provenance: Laboratory-generated (characterized viruses).
    • Analytical Inclusivity:

      • Test Set Sample Size: 3 replicates per virus isolate for 15 A/H5 (Asian lineage) viruses for both enzyme systems. One virus was tested in silico. Total real wet-lab tests: 15 viruses * 3 replicates * 3 assays (InfA, H5a, H5b) * 2 enzyme systems = 270 individual reactions.
      • Data Provenance: Laboratory-generated (characterized viruses representing different geographic locations and phylogenetic clades).
    • Analytical Specificity (Cross-Reactivity):

      • Test Set Sample Size: 3 replicates per virus for 14 human and animal influenza A and influenza B viruses tested. Total: 14 viruses * 3 replicates * 3 assays (InfA, H5a, H5b) = 126 individual reactions.
      • Data Provenance: Laboratory-generated (characterized viruses).
    • Analytical Specificity (Exclusivity):

      • Test Set Sample Size: A single test for each of 35 non-influenza organisms (16 viruses, 18 bacteria, 1 yeast). Total: 35 organisms * 3 assays (InfA, H5a, H5b) = 105 individual reactions.
      • Data Provenance: Laboratory-generated (propagated, titered, and characterized organisms).
    • Clinical Performance Evaluation:

      • Test Set Sample Size:
        • Positive Contrived Samples: 50 samples.
        • Negative Specimens: 65 specimens.
      • Data Provenance:
        • Positive Contrived Samples: Laboratory-generated (grown virus added to A549 cell suspension).
        • Negative Specimens: Obtained from a clinical study conducted during the 2011-2012 influenza season. This suggests the negative specimens were retrospective clinical samples (human respiratory specimens), originating likely from the US where CDC studies are typically conducted.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not explicitly state the number or qualifications of experts used to establish the ground truth for the test set.

    • For analytical studies (LOD, Inclusivity, Specificity), the ground truth is established by the known characteristics of the laboratory-controlled and characterized viral and bacterial stocks. This does not typically involve human expert consensus in the same way clinical diagnostics do.
    • For the clinical performance evaluation:
      • The "positive contrived samples" have their ground truth by definition (known virus added).
      • The "negative specimens" were previously tested as negative for influenza A by the CDC Human Influenza rRT-PCR Diagnostic Panel. The ground truth for these would have been established by the predicate device's results, which were likely interpreted by trained laboratory personnel, rather than experts in a concensus panel for the purpose of this study.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for the test set.

    • For analytical studies, the results are typically quantitative (e.g., Cq values) or qualitative (positive/negative call) based on predetermined assay thresholds. There's no indication of a need for expert adjudication of these raw results.
    • For clinical performance, the comparison is made against the known status of contrived samples or the prior result from a predicate device. An adjudication process (like 2+1, 3+1, etc.) is typically used when the ground truth itself is uncertain or to resolve discrepancies between readers/tests, which is not stated as part of this study design.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done.

    • MRMC studies are typically used to compare the performance of human readers (e.g., radiologists, pathologists) with and without AI assistance or against another diagnostic method, often looking at reader agreement and diagnostic accuracy.
    • This submission describes a RT-PCR diagnostic panel, which is an in vitro diagnostic (IVD) device. The performance is determined by the assay's ability to detect viral RNA, not by human reader interpretation of complex images or clinical data in the same way an MRMC study would apply. Therefore, the concept of "how much human readers improve with AI vs without AI assistance" is not applicable here.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, the studies described are essentially standalone performance evaluations of the device (assay) itself.

    • The analytical studies (LOD, Inclusivity, Specificity, Exclusivity) inherently assess the algorithm/reagent's ability to detect targets or avoid off-target detection.
    • The 'clinical performance evaluation' also tests the device's ability to correctly identify positive and negative contrived/clinical samples.
    • While a human laboratory technologist performs the PCR steps and interprets the output (e.g., reads Cq values and applies the interpretation algorithm to call positive/negative), the study evaluates the device's performance characteristics given those outputs, not the technologist's diagnostic ability with or without the device. The data presented is the direct output of the device's reaction, not a human interpretation layer being separately assessed for its impact.

    7. The Type of Ground Truth Used

    • Analytical Studies (LOD, Inclusivity, Cross-Reactivity, Exclusivity):
      • Known Characteristics of Propagated and Characterized Viruses/Organisms: For these studies, the ground truth is established by the known identity, titer, and genetic makeup of the laboratory-prepared viral and bacterial stocks. For example, an A/H5N1 virus is known to be A/H5N1 because it was characterized as such.
    • Clinical Performance Evaluation:
      • Contrived Samples: The ground truth for positive samples was established by the known addition of virus to a cell suspension.
      • Negative Specimens: The ground truth for negative specimens was established by their prior categorization as "negative for influenza A" using the CDC Human Influenza rRT-PCR Diagnostic Panel (likely the predicate device or a similar established method) in a previous clinical study. This is a form of comparator method ground truth.

    8. The Sample Size for the Training Set

    The document does not explicitly mention a training set or its sample size.

    • For IVD assays like RT-PCR, the development process (which would involve "training" or optimization) typically focuses on primer/probe design, reaction conditions, and calibration using initial lab-generated samples. The 'training' in this context is implicit in the assay development and refinement, rather than a distinct, formal training set for a machine learning algorithm.
    • The studies described are for validation (testing), not for development or training.

    9. How the Ground Truth for the Training Set Was Established

    As no explicit training set is mentioned (see point 8), the document does not describe how ground truth for a training set was established.

    • If an implicit 'training' phase (assay development) occurred, the ground truth would have been established through well-characterized laboratory reference materials and known viral isolates, similar to the ground truth for the analytical validation studies.