Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K152531

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-01-15

    (134 days)

    Product Code
    Regulation Number
    874.4760
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended to clear the end of a rigid rod endoscope in order to maintain clear visualization of endoscopic procedures without having to remove the scope from the surgical site.

    The device is indicated for use during routine diagnostic procedures and during endoscopic sinus surgery.

    Device Description

    The InstaClear Lens Cleaner system is equivalent to the predicate device cleared under K982594. The system includes a reusable AC powered console with pump, a reusable footpedal, and sterile disposable single-use accessories.

    The system is used to clear the end of a rigid endoscope in order to maintain clear visualization of endoscopic procedures without having to remove the scope from the surgical site. The device is indicated for use during routine diagnostic procedures and during endoscopic sinus surgery (ESS).

    The system accessories include sterile sheaths to fit various sizes of 4mm Olympus and Storz endoscopes, a sterile irrigation tubeset, and a non-sterile footpedal. The console includes software to control irrigation delivery and suction.

    AI/ML Overview

    This document describes the InstaClear Lens Cleaner device and its performance testing to support its substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and the Reported Device Performance

    Acceptance Criteria CategorySpecific Test/CriterionReported Device Performance
    BiocompatibilityCytotoxicity, Sensitization, Irritation (for patient-contacting materials)Full GLP biocompatibility testing on similar devices containing all patient-contacting materials on file for other Gyrus ACMI marketed devices.
    Electrical SafetyCompliance with IEC 60601-1Device complies.
    Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1-2Device complies.
    SterilizationSterility Assurance Level (SAL) of 10^-6 for disposable items (sheaths, tubesets)Sterilized with Ethylene Oxide, cycle validated in accordance with ISO 11135 to provide SAL of 10^-6.
    Shelf LifeInitial one-year shelf life for disposable itemsSupported by analysis of shelf-life stability of materials and packaging, and accelerated aging studies. Real-time aging testing will confirm.
    Cleaning PerformanceCleaning compatibility (e.g., with blood/mucous analog)Testing demonstrated cleaning compatibility.
    Noise LevelsAcceptable noise levels during operationTesting performed. (Specific acceptable level not provided, but implies met.)
    Heat GenerationAcceptable heat generation during operationTesting performed. (Specific acceptable level not provided, but implies met.)
    Tabletop StabilityDevice stability on tabletopTesting performed. (Implies met.)
    Burst PressureIntegrity of components under pressureTesting performed. (Implies met.)
    Suction StrengthEffective suction capabilityTesting performed. (Implies met.)
    Hub StrengthMechanical strength of hubsTesting performed. (Implies met.)
    Torque StrengthMechanical strength under torqueTesting performed. (Implies met.)
    Leak TestingAbsence of leaks in the systemTesting performed. (Implies met.)
    Ship TestingDevice integrity after shipping conditionsRepresentative samples subjected to environmental conditioning and ship testing. (Implies met.)
    Baseline Performance TestingGeneral functional performanceTesting performed. (Implies met.)
    Cleaning Cycle Time (compared to predicate)Comparable or improved cleaning cycle time post-debrisDevice performs as well as or better than the predicate device.
    Clearing Performance (compared to predicate)Comparable or improved clearing performance (removal of debris)Device performs as well as or better than the predicate device.
    Leak/Drip Occurrence Rate (compared to predicate)Comparable or improved absence of leaks/drips post-cleaningDevice performs as well as or better than the predicate device.
    Risk AnalysisCompliance with ISO 14971Risk analysis carried out in accordance with established in-house acceptance criteria based on ISO 14971.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the specific sample sizes used for each performance test. It refers to "representative samples" for stability and environmental testing.

    • Data Provenance: The studies were non-clinical (electrical, mechanical, functional) performance testing and preclinical (bench) evaluations and testing. These are laboratory-based tests, not human trials.
    • Country of Origin: Not specified, but conducted by Gyrus ACMI, Inc. a US-based company with manufacturing in Bartlett, TN, USA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. This device's performance was evaluated through non-clinical and preclinical bench testing against objective criteria and comparison to a predicate device, not through expert-reviewed clinical data or human studies.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, the testing involved objective performance metrics and comparison to a predicate device, not expert adjudication of human subject results.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a lens cleaner, not an AI-powered diagnostic or interpretive tool that would involve human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical medical device, not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the non-clinical and preclinical performance tests was based on:

    • Objective Engineering Specifications: For electrical safety, EMC, burst pressure, torque strength, etc.
    • Established Sterilization Standards: For sterility assurance level.
    • Simulated Use Conditions with Analogs: For cleaning performance (using a blood analog).
    • Performance Characteristics of the Predicate Device: For comparative evaluation of features like cleaning cycle time, clearing performance, and leak/drip rate.

    8. The Sample Size for the Training Set

    Not applicable. This document describes a physical medical device, not a machine learning or AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1