Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K152318

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-11-17

    (92 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MD-2300S Ultrasonic A/B Scanner for Ophthalmology is intended for ophthalmological ultrasonic diagnosis and AL biometric measurement.

    The 10MHz B-Probe applies to normal ophthalmological ultrasonic diagnosis; and the 20MHz B-Probe is suitable for observing the details of intraocular and retina structure.

    The device should be operated by trained medical staff.

    Patients with eyelid trauma and severe eye infection are prohibited from using B-Scan; and patients and cornea trauma are prohibited from using A-Biometric scan.

    The A-biometric scan should be used cautiously on patients without independent behavior abilities or who are highly sensitive to contacting measurements, whose ineffective cooperation may result in inaccurate measurements,

    Device Description

    MD-2300S Ultrasonic A/B Scanner for Ophthalmology is an ultrasonic imaging instrument specialized in ophthalmological diagnosis. It consists of a main unit, a power adaptor, a 10MHz B-Scan probe, a 20MHz B-Probe, a 10MHz A-Probe, a footswitch, a keyboard and a mouse.

    MD-2300S includes two most popular operation modes for ophthalmological clinical diagnosis: B-mode ultrasonic section imaging and A-mode axial biometric parameter measurement.

    The 10MHz B-Probe applies to normal ophthalmological ultrasonic diagnosis; and the 20MHz B-Probe is suitable for observing the details of intraocular and retina structure.

    The part of B-mode uses 10MHz and 20MHz mechanical sector-scan probes for scanning. It acquires echo signal through ultrasonic pulse transmitting and receiving circuits; and converts it into digital information to save in FIFO; the data information of each echo line is transmitted to the embedded board through USB port. The application software running on Windows XPE platform realizes real time display, saving and case report generation for images of eye tissue.

    The part of A-mode uses 10MHz A-biometric probe for A-mode scanning. It acquires echo signal through ultrasonic pulse transmitting and receiving circuits; and converts it into digital information to save in FIFO; the data information of each echo line is transmitted to the embedded board through USB port. Based on the interface reflection of ultrasound wave in three different tissues of anterior chamber, lens and vitreous body, the application software running on Windows XPE platform measures the transmitting time of ultrasound wave in different tissues and then calculates distances of each segment according to the acoustic velocity of different tissues to get the axial length and generates case report of A-biometry automatically, thus provides parameters for intraocular lens implantation operation.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the MD-2300S Ultrasonic A/B Scanner for Ophthalmology. The information provided primarily focuses on demonstrating substantial equivalence to predicate devices rather than a standalone clinical study with explicit acceptance criteria and statistical performance metrics for novel AI algorithms.

    Here's a breakdown of the requested information based on the provided text, recognizing that some of the questions may not be fully applicable to this type of submission which is focused on substantial equivalence for a medical device rather than an AI/ML algorithm:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly state "acceptance criteria" in the typical sense of a target performance threshold for a new AI algorithm. Instead, it demonstrates the device's technical specifications and clinical utility are substantially equivalent to predicate devices. The "performance" is described in terms of technical characteristics and observational clinical findings.

    FeatureAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance (MD-2300S)
    B-mode
    10MHz Probe Axial ResolutionNo more than 0.2mm (Predicate: Aviso)No more than 0.1mm (Better than predicate)
    10MHz Probe Lateral ResolutionNo more than 0.6mm (Predicate: Aviso)No more than 0.2mm (Better than predicate)
    20MHz Probe Axial ResolutionNo more than 0.1mm (Predicate: Aviso)No more than 0.08mm (Better than predicate)
    20MHz Probe Lateral ResolutionNo more than 0.25mm (Predicate: Aviso)No more than 0.15mm (Better than predicate)
    B-Scan Image Quality (Clinical)Comparable to existing devices, provides valuable reference for diagnosis."can display the intraocular structure clearly and provide valuable reference base for clinical diagnosis. The image resolution as well as the scanning scope, frame rate, gain adjustment, saving, image processing, cineloop, case management and other functions are able to satisfy the clinical demands."
    A-mode
    AL Biometric Measuring AccuracyNo more than ±0.05mm (Explicitly stated as a characteristic, likely an internal requirement or derived from similar predicate devices, though not directly compared to a specific predicate's accuracy in the given table).No more than ±0.05mm (Reported as a characteristic).
    AL Measuring Data ConsistencyNo obvious difference with predicate (ODM-2100) using Bland-Altman method."the result of statistical processing for the AL measuring data showed that the measuring results of both products have good consistency."
    IOL CalculationFunctionality substantially equivalent to MD-1000A (for non-refractive surgery) and identical to Aviso (for refractive surgery).For non-refractive surgery cases, IOL calculation is "substantially equivalent" to MD-1000A. For refractive surgery cases, IOL calculation is "identical" with Aviso. Supports SRK-T, SRK-II, BINK-II, HOLLADAY, HOFFER-Q, HAIGIS. Supports IOL calculation after refractive surgery with comparative display. Automatic measurement and 10 groups averaging and display standard deviation.
    Safety & EMCCompliance with IEC 60601-1, IEC 60601-2-37, IEC 60601-1-2.Complies with IEC 60601-1: 2005 + AM1: 2012, IEC 60601-2-37: 2007, IEC 60601-1-2: 2007. Acoustic output parameters comply with IEC 60601-2-37 and FDA Guidance.
    BiocompatibilityConforms to ISO 10993 requirements.Tests for in vitro cytotoxicity, skin sensitization and irritation conducted on patient-contacting materials and conform to biocompatibility requirements.
    Software & Essential PerformancePassed verification and validation.Results comply with requirements.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • B-Mode Test Set:
      • Sample Size: "randomly outpatients and inpatients that need to make B-mode ultrasonic examination" (no specific number provided). The selection aimed to cover "typical cases of ultrasonic diagnosis for ophthalmology."
      • Data Provenance: Prospective (implied by "make ultrasonic inspection") clinical validation study conducted at Tianjin Eye Hospital, China.
    • A-Mode Test Set:
      • Sample Size: "50 eyes randomly from outpatients and inpatients that need to measure axial length (AL)."
      • Data Provenance: Prospective clinical validation study conducted at Tianjin Eye Hospital, China. Patients' sex and age were not restricted.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • B-Mode: "Clinical doctors validate the intended use according to the image quality of B-Scan on orbit and intraocular tissues." The number and specific qualifications of these "clinical doctors" are not stated.
    • A-Mode: The ground truth for A-mode measurements was established by comparing measurements from the MD-2300S against a predicate device, the ODM-2100 Ultrasonic A/B Scanner for Ophthalmology. This implies the predicate device's measurements served as the reference. The operators for both devices are described as "expected user representatives," implying trained medical staff, but specific qualifications are not detailed.

    4. Adjudication Method for the Test Set

    • B-Mode: Not explicitly stated as a formal adjudication process. The text indicates "Clinical doctors validate the intended use according to the image quality." This suggests a qualitative assessment by medical professionals.
    • A-Mode: "The statistical processing results for AL measuring data with Bland-Altman statistical method should have no obvious difference." This is a statistical method for assessing agreement between two different methods of measurement, rather than a formal expert adjudication of ground truth.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size of how much human readers improve with AI vs without AI assistance

    This document describes the premarket notification for a diagnostic ultrasound system, not an AI/ML device assisting human readers. Therefore, an MRMC comparative effectiveness study involving AI assistance for human readers was not conducted or reported in this submission. The "clinical effectiveness validation" was about the device's inherent diagnostic capabilities.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is not an AI/ML algorithm. The device, an ultrasonic scanner, performs measurements (A-mode) and generates images (B-mode) in a standalone capacity. Operators (human-in-the-loop) are required to use the device. The A-mode's automated measurement of AL and IOL calculation is a function of the device, validated against another device's measurements. The B-mode image quality is validated by clinical doctors' assessment.

    7. The Type of Ground Truth Used

    • B-Mode: Expert consensus/clinical assessment based on qualitative evaluation of image quality by "Clinical doctors."
    • A-Mode: Comparison against a legally marketed predicate device (ODM-2100 Ultrasonic A/B Scanner for Ophthalmology) using the Bland-Altman statistical method to assess agreement. This acts as a reference standard for the AL measurement. Pathology or outcomes data are not mentioned as ground truth sources.

    8. The Sample Size for the Training Set

    This submission does not discuss an AI algorithm or a "training set." The device is a traditional medical imaging device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no mention of a training set or AI algorithm in this submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1