Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K152225

    Validate with FDA (Live)

    Date Cleared
    2015-11-24

    (109 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Edwards eSheath Introducer Set is indicated for the introduction and removal of devices used with the Edwards SAPIEN 3 Transcatheter Heart Valve.

    Device Description

    The Edwards eSheath Introducer Set consists of a sheath and 2 introducers. It is available with inner sheath diameters of 14 French (model 914ES) and 16 French (model 916ES). The 14 French introducer set is used to facilitate introduction of the 23mm and 26mm SAPIEN 3 THV and Commander Delivery System into the vasculature, and the 16 French introducer set is used to facilitate introduction of the 29mm SAPIEN 3 THV and Commander Delivery System into the vasculature.

    The sheath shaft is comprised of 2 layers of material (HDPE/TecoFlex coextruded outer layer and PTFE liner). The outer and inner layer are folded, creating a seam which allows the distal region of the sheath to temporarily expand in diameter when a device is inserted. A tapered strain relief at the proximal end of the sheath shaft helps provide hemostasis when the sheath is inserted.

    A radiopaque marker on the distal end indicates the location of the sheath tip in the body and a hydrophilic coating on the sheath tubing exterior facilitates introduction into the vessel. The sheath tubing mates with a housing, which holds three seals (valves) to provide hemostasis: a duckbill seal, a disc seal, and a cross slit seal. The housing also includes flushport tubing and a stopcock to allow for flushing of the sheath.

    Two introducers are provided to aid in the introduction of the target vessel, and can also be used for dilation of the vessel prior to sheath insertion. The introducers are radiopaque and feature a tapered tip and guidewire lumen.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Edwards eSheath Introducer Set, focusing on acceptance criteria and the study that proves the device meets them. Please note that the document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than a comprehensive "study" in the academic sense of proving efficacy through clinical trials. Therefore, some of the requested information may not be present as it pertains more to clinical studies of diagnostic AI devices.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of numerical acceptance criteria with corresponding performance metrics for the Edwards eSheath Introducer Set. Instead, it lists the types of non-clinical tests performed to demonstrate that "performance characteristics... are equivalent to the predicates, and to verify that design requirements are satisfied." The stated outcome for all tests is "successfully completed."

    Acceptance Criteria Category (Implicit)Specific Test PerformedReported Device Performance
    Physical Integrity / Visual QualityVisual Surface InspectionSuccessfully completed (meets design requirements)
    Dimensional AccuracyDimensional InspectionSuccessfully completed (meets design requirements)
    Imaging VisibilityRadiopacity/VisualizationSuccessfully completed (meets design requirements)
    Functional CompatibilityGuidewire CompatibilitySuccessfully completed (meets design requirements)
    Hemostasis ControlHemostasisSuccessfully completed (meets design requirements)
    Ease of Use / DurabilityLubricity and Durability of the SheathSuccessfully completed (meets design requirements)
    Structural ResilienceKink ResistanceSuccessfully completed (meets design requirements)
    Expandability / Return CapabilitySeam Return After ExpansionSuccessfully completed (meets design requirements)
    Component StrengthBond StrengthSuccessfully completed (meets design requirements)
    InteroperabilityDevice InteractionSuccessfully completed (meets design requirements)
    Surface PropertiesHydrophilic Coating CharacterizationSuccessfully completed (meets design requirements)
    CleanlinessUSP Particulate TestSuccessfully completed (meets design requirements)
    Sterility AssuranceSterilization ValidationSuccessfully completed (meets design requirements)
    BiocompatibilityCytotoxicitySuccessfully completed (meets design requirements)
    HemocompatibilitySuccessfully completed (meets design requirements)
    Systemic ToxicitySuccessfully completed (meets design requirements)
    Material Mediated PyrogenicitySuccessfully completed (meets design requirements)
    Irritation/Intracutaneous ReactivitySuccessfully completed (meets design requirements)
    SensitizationSuccessfully completed (meets design requirements)
    Chemical AcceptabilitySuccessfully completed (meets design requirements)
    Blood CompatibilityThrombogenicitySuccessfully completed (meets design requirements)
    Product ProtectionPackaging IntegritySuccessfully completed (meets design requirements)
    Product LifespanShelf Life VerificationSuccessfully completed (meets design requirements)

    2. Sample Size Used for the Test Set and Data Provenance

    This document describes non-clinical (bench) testing, not a clinical study involving patient data. Therefore, the concept of a "test set" in the context of human data or AI algorithms is not applicable here. The data provenance is internal to Edwards Lifesciences, derived from various laboratory and engineering tests; it is not data from human subjects or from a specific country of origin in that sense. These were likely prospective tests performed on manufactured samples of the device.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. Ground truth for non-clinical engineering tests (e.g., bond strength, dimensional accuracy, sterilization) is established through standardized methodologies, engineering specifications, and validated measurement equipment, not through expert human interpretation in the way a diagnostic imaging study would use radiologists.

    4. Adjudication Method for the Test Set

    Not applicable. Since the tests are objective, performance is measured against established specifications or standards. There is no human interpretation or adjudication required in the sense of resolving conflicting expert opinions.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. This device is a physical medical introducer set, not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies and the concept of human readers improving with AI assistance are irrelevant to this submission.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is a physical medical introducer set, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical tests is based on:

    • Engineering Specifications: Predetermined design requirements and tolerances.
    • International/National Standards: Adherence to standards like USP for particulates, ISO standards for sterilization, and specific test methods for biocompatibility (e.g., ISO 10993 series).
    • Predicate Device Performance: The primary goal is to demonstrate equivalence in performance characteristics to the legally marketed predicate devices.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device requiring a training set. The "training" for such a device would be its design and manufacturing process, which is refined through iterative engineering development and testing.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As this is not an AI/ML device, the concept of a "training set" and associated ground truth establishment does not apply. The analogous process for a physical medical device involves:

    • Defining design inputs and requirements based on intended use, user needs, and regulatory standards.
    • Developing and manufacturing prototypes.
    • Conducting iterative non-clinical testing (similar to the "Summary of Non-Clinical Testing" listed) to ensure the device meets those requirements.
    • Refining the design based on test results.

    Summary Takeaway:

    The document is a 510(k) summary for a physical medical device (catheter introducer set). Its purpose is to demonstrate substantial equivalence to existing predicate devices through comprehensive non-clinical (bench) testing. The questions regarding AI acceptance criteria, training/test sets, expert ground truth, adjudication, and MRMC studies are not applicable to the nature of this particular device and its regulatory submission pathway. The acceptance criteria are implicitly met by "successfully completing" the listed non-clinical tests, which are designed to show that the device performs equivalently to its predicates and meets all relevant design requirements and safety standards.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1