    Manufacturer
    Date Cleared
    2016-01-19

    (165 days)

    Product Code
    Regulation Number
    872.3770
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DENTALOS PLUS discs are milling blanks consisting of Polymethylmethacrylate (PMMA) and designed for the fabrication of long-term temporary crown and bridgework. They are machined by use of the CAD/CAM technique.

    DENTALOS PLUS discs are recommended for manufacturing substructures of single tooth crowns and bridgeworks with up to two pontics.

    Device Description

    DENTALOS PLUS discs are milling blanks composed of hot cured polymethylmethacrylate (PMMA). They are intended to be used by dental professionals e.g. dental technicians for the fabrication of long-term temporary crowns and bridgeworks as custom-made restorations for the sole use of a particular patient. These restorations are designed virtually by dental technicians using the CAD technology on the basis of intraoral scans or scans from impressions and/or models. The designed restorations can thereafter be machined in all appropriate CAM Milling Centers out of DENTALOS PLUS Discs. In a further step the milled workpiece can be individually characterized with veneering materials and polished for to improve the aesthetic appearance of the finished restoration. DENTALOS PLUS discs are offered as monochrome discs (DENTALOS PLUS MONO) in nine different shades and as multicolored discs (DENTALOS PLUS MULTICOLOR) in five shades, all discs in different thicknesses.

    AI/ML Overview

    The document provided is a 510(k) Summary for a medical device called "DENTALOS PLUS MONO" and "DENTALOS PLUS MULTICOLOR," which are milling blanks made of polymethylmethacrylate (PMMA) for fabricating temporary crowns and bridges. This submission is for establishing substantial equivalence to a predicate device, not for a novel AI/software device.

    Therefore, the requested information about acceptance criteria and study details related to AI/software performance (e.g., sample sizes for training/test sets, expert adjudication, MRMC studies, standalone performance, ground truth establishment) is not applicable to this document.

    The document focuses on demonstrating physical and material property comparability of the new device to a predicate device (BRIGHTGLASS and BRIGHTGLASS M).

    However, I can extract the relevant information from the document regarding the performance testing that was conducted:

    1. A table of acceptance criteria and the reported device performance

    The document reports performance parameters for the "DENTALOS PLUS MONO" and "DENTALOS PLUS MULTICOLOR" devices, and implicitly expects them to be comparable to the predicate devices. The table below presents the reported physical properties. The acceptance criteria are implied to be "comparable to the predicate devices," as stated in the "Performance Testing" section.

    Performance ParameterReported Device Performance (DENTALOS PLUS MONO/MULTICOLOR)Acceptance Criteria (Implicit: Comparable to Predicate)
    e-module3370 MPA3370 MPA (Predicate: BRIGHTGLASS/BRIGHTGLASS M)
    Tensile Strength76.3 MPA76.3 MPA (Predicate: BRIGHTGLASS/BRIGHTGLASS M)
    Elastic Limit70 MPA70 MPA (Predicate: BRIGHTGLASS/BRIGHTGLASS M)
    Bending Strength106 MPA106 MPA (Predicate: BRIGHTGLASS/BRIGHTGLASS M)
    Water Absorption19.36 µg/m³19.36 µg/m³ (Predicate: BRIGHTGLASS/BRIGHTGLASS M)
    Water Solubility< 1 µg/m³< 1 µg/m³ (Predicate: BRIGHTGLASS/BRIGHTGLASS M)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample size for the tests conducted (tensile strength, elastic limit, bending strength, e-module, water solubility, and water absorption). It only states that "a series of testing was performed." The provenance of the data (country of origin, retrospective/prospective) is not mentioned.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is a material properties comparison for a dental device, not an AI/software evaluation requiring expert-established ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is a material properties comparison, not an AI/software evaluation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a material properties comparison, not an AI/software evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a material properties comparison, not an AI/software evaluation.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The "ground truth" in this context is based on physical and chemical testing specified by ISO standards (ISO 10477:2004, ISO 527-1:2012, and ISO 572-2:2012). These standards define the methods and parameters for evaluating the properties of dental polymer-based materials. The performance of the new device is compared directly to the predicate device based on these objective measurements.

    8. The sample size for the training set

    Not applicable. This is a material properties comparison, not an AI/software evaluation that involves a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is a material properties comparison, not an AI/software evaluation that involves a training set.