510(k) Data Aggregation

    K Number
    K152128

    Validate with FDA (Live)

    Date Cleared
    2015-12-23

    (145 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 2-in-1 Abdominal Muscle Training & Back Pain Relief System has two modes of stimulation; a transcutaneous electrical nerve stimulation (TENS) BP-313 mode and a powered muscle stimulation (PMS) TS-212 mode.

    The TENS (BP-313 Mode) is indicated for temporary relief of pain associated with sore and aching muscles in the waist. back, arm, leg due to strain from exercise or normal household and work activities.

    The PMS (TS-212 Mode) is indicated for the improvement of abdominal muscle tone, for strengthening of abdominal muscles and for the development of a firmer abdomen. This system is intended to stimulate healthy muscles in order to improve and facilitate muscle performance.

    Device Description

    2-in-1 Abdominal Muscle Training & Back Pain Relief System is a portable multifunctional device, offering both Transcutaneous Electronic Nerve Stimulator (TENS) and Powered Muscle Stimulator (PMS) qualities.

    2-in-1 Abdominal Muscle Training & Back Pain Relief System has 12 operation modes, which can give certain electrical pulse through 4 of electrode pads placed on the skin to help users to enjoy body stimulation.

    The support belt which fastens around your waist provides appropriate stimulation position for a quick and easy exercise for users of all different waist sizes. The adhesive pads adhere to the belt to conduct the signal from the unit to your abdominal muscles. Snap cables are for connecting the stimulator and the support belt. Each unit comes with one battery (6F22/1604) so you can start exercising right away.

    AI/ML Overview

    The provided document is a 510(k) Summary for a medical device called "2-in-1 Abdominal Muscle Training & Back Pain Relief System." This type of document is submitted to the FDA to demonstrate that a new device is substantially equivalent to a legally marketed predicate device. As such, it focuses on comparing the new device to existing ones rather than detailing a specific clinical study for the new device's performance against defined acceptance criteria in the format requested.

    Therefore, the specific information about detailed acceptance criteria, a dedicated study proving performance, sample sizes for test/training sets, experts, adjudication methods, MRMC studies, standalone performance, and ground truth establishment for this specific device's clinical performance is not present in the provided text.

    The document primarily outlines design validation tests performed to ensure electrical safety, electromagnetic compatibility, biocompatibility, and software functionality, as well as a waveform test to verify output specifications. These are essential for establishing substantial equivalence but are not clinical performance studies with acceptance criteria in the sense of demonstrating diagnostic accuracy or treatment efficacy in human subjects.

    However, I can extract information related to the tests performed and the standards they comply with:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide specific quantitative acceptance criteria or reported device performance in the format of a clinical trial. Instead, it states that the device was evaluated against various standards. The "performance" in this context refers to compliance with these standards and the device's technical specifications.

    Test CategoryAcceptance Criteria (Stated as Compliance Standard)Reported Device Performance
    Electrical SafetyIEC 60601-1 and IEC 60601-2-10 standardsComplies with IEC 60601-1 and IEC 60601-2-10 requirements.
    Electromagnetic CompatibilityIEC 60601-1-2 standardComplies with IEC 60601-1-2 requirements.
    BiocompatibilityISO 10993-5 and ISO 10993-10 standardsComplies with ISO 10993-5 and ISO 10993-10 requirements for user-contacting materials.
    Software Verification/ValidationFDA "Guidance for Pre Market Submissions and for Software Contained in Medical Devices"Complies with FDA guidance requirements.
    Waveform Test (Output Specifications)FDA "Guidance for Transcutaneous Electrical Nerve Stimulator for Pain Relief Intended for Over the Counter Use" and "Guidance for Powered Muscle Stimulator for Muscle Conditioning"Verified output specifications according to FDA guidance. (Specific parameters like Voltage, Current, Pulse Duration, etc., are compared to predicate devices and deemed compliant with standards/guidance).
    Patient Leakage CurrentComply with IEC 60601-1 and IEC 60601-2-10AC: 54.5µA, DC: 0.5µA (Subject Device) vs. Comply with IEC 60601-1 and IEC 60601-2-10 (Predicate)
    Average DC current through electrodes (no pulses)Comply with IEC 60601-1 and IEC 60601-2-100A (Subject Device) vs. < 0.01μA (Predicate)

    Notes on the "Reported Device Performance" for Output Specifications:
    The document provides detailed technical specifications for the subject device and predicate devices (e.g., Maximum Output Voltage, Maximum Output Current, Pulse Duration, Pulse Frequency, Net Charge, etc.). It states that "Although the... are a little different from the predicate devices, they all comply with IEC 60601-1, IEC 60601-2-10 requirement for Transcutaneous Electrical Nerve Stimulator for Pain Relief and FDA guidance requirement for Powered Muscle Stimulator for Muscle Conditioning." This implies that the device's performance on these specific electrical parameters meets the established regulatory and guidance thresholds, even if they differ numerically from the predicate devices.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    Not applicable. The document describes engineering and bench testing, not clinical studies involving human subjects with test sets in the traditional sense.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    Not applicable. Ground truth for clinical performance is not established as this is not a clinical study. The "truth" in this context refers to compliance with international standards and FDA guidance through lab testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    Not applicable. This is not a clinical study requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This device is an electrical stimulator, not an AI-assisted diagnostic or treatment planning tool that would typically involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
    Not applicable. This is a physical medical device, not an algorithm. Its performance is evaluated through bench testing and compliance with standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    The "ground truth" or reference standard for the device's design and technical specifications is compliance with established international medical device standards (IEC 60601-1, IEC 60601-2-10, ISO 10993-5, ISO 10993-10, IEC 60601-1-2) and relevant FDA guidance documents for Transcutaneous Electrical Nerve Stimulators and Powered Muscle Stimulators.

    8. The sample size for the training set
    Not applicable. There is no mention of a training set for an algorithm in this document, as it describes a physical electrical stimulator.

    9. How the ground truth for the training set was established
    Not applicable, as there is no training set mentioned.