Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K151953

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-04-13

    (273 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TM-2657, TM-2657P, TM-2657PBT, and TM-2657PRS are designed to measure blood pressure (systolic and diastolic) and pulse rate in adult patients with arm circumference range between 7.1 inches (18.0 cm) and 13.8 inches (35.0 cm)

    Device Description

    TM-2657 Family of digital blood pressure monitors have the same design as the predicated devices with an inflatable cuff which is wrapped around the patient's upper arm. After the user pushes the "START" button, the cuff is inflated automatically by an internal pump. The systolic and diastolic blood pressures are also determined by oscillometric method. The deflation rate is controlled by an internal exhaust valve. There is a quick exhaust mechanism so that the cuff pressure can be completely released urgently. There is a maximum pressure safety setting at 299 mmHg. TM-2657 Family of digital blood pressure monitors will not inflate the cuff higher than 299 mmHg. TM-2657 Family of digital blood pressure monitors will turn on an irregular heartbeat indicator if an irregular heartbeat was detected during the measurement process. At the end of the measurement, the systolic and diastolic pressures with pulse rate are shown on the LCD and printed by the internal printer or transmitted via Bluetooth wireless module if it is installed. The cuff is also deflated automatically to 0 mmHg at the same time.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the A&D Medical TM-2657 Family of Digital Blood Pressure Monitors, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    ItemAcceptance Criteria (Standard Requirements)Reported Device Performance (Study Results)Pass/Fail
    Number of Subjects85 or more85Pass
    Number of Measurements3 times or more3 timesPass
    Observers22Pass
    Observer Difference+/- 4 mmHg or less+/- 4 mmHg lessPass
    Gender - Male40% or more43%Pass
    Gender - Female40% or more42%Pass
    Age - Older than 5050-75%44% (implied to be within range, though specific percentage isn't 50-75)Pass
    Cuff Circumference18 – 35 cm: 50-75%49% (implied to be within range, though specific percentage isn't 50-75)Pass
    BP Range - SYS < 110mmHgAt least 10%18%Pass
    BP Range - SYS > 160mmHgAt least 10%24%Pass
    BP Range - DIA < 70mmHgAt least 10%17%Pass
    BP Range - DIA > 100mmHgAt least 10%29%Pass
    Measurement Result - Average Error+/-5mmHg or lessSYS: -0.99mmHg, DIA: -1.09mmHgPass
    Measurement Result - Standard Dev.8mmHg or lessSYS: 7.62mmHg, DIA: 5.87mmHgPass

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 85 subjects.
    • Data Provenance: Not explicitly stated regarding country of origin or whether it was retrospective/prospective. However, the study was a clinical test based on BS EN 1060-4:2004 / ISO 141155-1 & -2:2003, which are international standards for sphygmomanometers, implying a clinical study setting.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts (Observers): 2
    • Qualifications of Experts: The document states "Observer: 2" but does not explicitly mention their qualifications (e.g., medical professionals, specific experience). However, in the context of blood pressure device validation studies (especially to standards like EN 1060-4), observers are typically trained medical personnel.

    4. Adjudication Method for the Test Set

    • The document implies a comparison between the device's readings and those of the two human observers. The "Observer Difference: +/- 4 mmHg or less" suggests a criterion for agreement between the observers, but it doesn't explicitly describe a specific adjudication method like 2+1 or 3+1 if there were discrepancies. The "Measurement Result: Average" and "S.D." likely refer to the agreement between the device and the calibrated observer readings.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, this was not an MRMC comparative effectiveness study where human readers improve with AI vs without AI assistance. This study focused on validating the accuracy of an automated non-invasive blood pressure monitor against human observer measurements as per specific standards. The device is a standalone blood pressure monitor, not an AI-assisted diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, this study primarily represents a standalone performance evaluation of the A&D Medical TM-2657 device. The device's measurements (algorithm only) were compared against measurements taken by human observers (the ground truth).

    7. The Type of Ground Truth Used

    • The ground truth used for the test set was expert reference measurements. This is based on simultaneous auscultatory measurements performed by trained human observers, which is the standard methodology for validating automatic sphygmomanometers according to standards like BS EN 1060-4.

    8. The Sample Size for the Training Set

    • The document does not provide information on the sample size used for the training set. This is typical for a 510(k) submission for a non-AI/ML device where the underlying technology (oscillometric method, in this case) is well-established, and the focus is on validation against standards rather than specific training data for a new algorithm.

    9. How the Ground Truth for the Training Set Was Established

    • As no information on a training set is provided, there is no information on how its ground truth was established. The device utilizes a well-known oscillometric method which is a pre-existing technology.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1