510(k) Data Aggregation

    K Number
    K151890

    Validate with FDA (Live)

    Date Cleared
    2015-12-23

    (166 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The "GBOX-15A/15B" are indicated for use in surgical applications requiring the hemostasis, ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: aesthetic (dermatology and plastic surgery), gastroenterology, general surgery, genitourinary surgery (urology), qynecology (GYN), neurosurgery, otolaryngology (ENT), ophthalmology, arthroscopy, podiatry, pulmonology, and thoracic surgery; and Laser Assisted Lipolysis (980 nm only).

    The "VELASII-30A/30B" are indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology (BPH), Genitourinary (UJrology), Thoracic Surgery, Plastic Surgery and Dermatology, Aesthetics including vascular lesions and hair removal, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery (PLDD), Gastroenterology, Head/neck/ENT and Radiology Endovascular coagulation, Oral Surgery and Dental procedures, laser assisted lipolysis.

    The "VELASII -60A/60B/60C" are indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue. Such as: Dermatology, Aesthetics, Plastic Surgery, Vascular Surgery, Podiatry, Endovenous Occlusion of the greater saphenous vein of the thigh in patients with superficial vein reflux.

    Device Description

    Medical diode laser systems are a kind of laser with semiconductor as working material. It consists of working material, cavity resonator and power source.

    The medical diode laser systems for this unit is GaAlAs diode bar, and the wavelength are 810mm/980mm of GBOX-15A/15B, 810/980mm of VELAS II -30A/30B, 810/980/940nm of VELAS II -60A/60B/60C. It features impact structure, high efficiency and long lifetime. The beam divergence angle is too large lead to poor beam quality. In order to better meet the needs of Laser Diode practical applications, Laser Diode carried beam shaping is very necessary. With the GIGAA'S unique fiber-coupling technology, the laser beam can be coupled efficiently into the fiber.

    To start the laser unit, turn the main switch ON into " | " position. Immediately the power indicator will be green with the system fans working. At the same time, the LCD screen lights up as the picture shows below.

    After finishing setting the parameters, press the "Ready" button and then the system will remind you to wear the safety goggles (protect wavelength from 800nm to 1100nm). At this time when you press down the footswitch, the laser will emit.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from Wuhan Gigaa Optronics Technology Company, Ltd. for their Medical Diode Laser Systems, including models GBOX-15A/15B, VELASII-30A/30B, and VELASII-60A/60B/60C. The purpose of a 510(k) submission is to demonstrate that the new device is "substantially equivalent" to a legally marketed predicate device, meaning it is as safe and effective.

    The submission focuses on comparing the proposed devices to existing predicate devices rather than establishing novel acceptance criteria based on a specific clinical study of the proposed device's performance against a ground truth.

    Here's an analysis based on the provided text, addressing your questions:

    1. Table of Acceptance Criteria and Reported Device Performance

    This document does not present "acceptance criteria" in the typical sense of quantitative performance metrics (e.g., sensitivity, specificity, accuracy) that a device must meet in a direct study against a ground truth. Instead, it demonstrates "substantial equivalence" to predicate devices by comparing their specifications and intended uses. The "reported device performance" is largely framed within the context of these comparative specifications and adherence to recognized standards.

    Here's a table summarizing the comparison, highlighting where the proposed devices align with or are similar to the predicate devices, which implicitly serves as the "acceptance criteria" for substantial equivalence.

    FeatureAcceptance Criteria (Predicate Device K082721 and K122567/K100558)Reported Device Performance (Wuhan Gigaa Optronics Technology Co., Ltd. Medical Diode Laser Systems)Comparison Conclusion (as stated in document)
    General
    Product CodeGEXGEXSame
    Device TypeDiode LaserDiode LaserSame
    ClassIIIISame
    Safety ClassificationClass I Type BClass I Type BSame
    Laser Class44Same
    Operation InterfaceColor LCD touch screenColor LCD touch screenSame
    Microprocessor ControlYesYesSame
    Indications for UseSurgical applications requiring ablation, hemostasis, vaporization, excision, incision, and coagulation of soft tissue in specified medical specialties (aesthetic, gastroenterology, general surgery, genitourinary surgery, gynecology, neurosurgery, otolaryngology, ophthalmology, arthroscopy, podiatry, pulmonology, thoracic surgery, and Laser Assisted Lipolysis for 980 nm). For some models, also includes use with endoscopic equipment and specific conditions like BPH, vascular lesions, hair removal, endovenous occlusion.Covers the same range of surgical applications and medical specialties: ablation, hemostasis, vaporization, excision, incision, and coagulation of soft tissue, including specific uses like Laser Assisted Lipolysis and endoscopic applications. Indications for GBOX-15A/B are identical to LaserPro; for VELAS II-30A/B and VELAS II-60A/B/C, they are "same" or "similar" to JOULE and QUANTA systems.Same or Similar. "The indications for use of 'VELAS II -30A/0B' are same to that of Quanta System QUANTA Diode Laser Family. The indications for use of 'VELAS II -60A/60B/60C' are same to that of JOULE 810/940/980 Multi-Platform System." Minor structural differences are deemed not to affect safety and effectiveness.
    Technical Specifications (Examples)
    Wavelength (GBOX-15A/B)LaserPro 810: 810nm±10%; LaserPro 980: 980nm±10%GBOX-15A: 810nm±10nm; GBOX-15B: 980nm±10nmSimilar. "Center wavelength is consistent, and GBOX wavelength accuracy is better than that of the predicate device."
    Wavelength (VELAS II)JOULE 810/940/980 (810nm, 940nm, 980nm); QUANTA 808/980 (808nm, 980nm)VELAS II-30A, VELAS II-60A: 810nm $\pm$ 10nm; VELAS II-30B, VELAS II-60B: 980nm $\pm$ 10nm; VELAS II-60C: 940nm $\pm$ 10nmSimilar. "Center wavelength is consistent. Reasonable accuracy range of VELAS II does not affect safety and effectiveness."
    Output Power (GBOX)Predicate: 1-15W, 1-20W, 1-25WGBOX: 1-15WSimilar. "Same as 1-15W. GBOX design output power has claimed to meet intended use."
    Output Power (VELAS II)JOULE: $\leq$ 100W; QUANTA: 30WVELAS II-30A/B: 1-30W; VELAS II-60A/B/C: 1-60WSimilar. "Output power of 'VELAS II -30A/30B' is Same to that for 'QUANTA 808, QUANTA 980'. Output power range of 'VELAS II -60A/60B/60C' is contained in that of 'JOULE 810, JOULE 940, JOULE 980', and it has been designed to meet the intended use requirement claimed."
    Pulse width (GBOX)Predicate: 0.05s-10sGBOX: $\le$ 2.5sSimilar. "Pulse width range of the device is contained in that of the predicate device, and it has been designed to meet the intended use requirement claimed."
    Pulse width (VELAS II)JOULE: $\le$ 2500ms (2.5s); QUANTA: 0.25-10000ms (0.25s-10s)VELAS II: 10ms-2.5sSimilar. "Pulse width range of 'VELAS II -30A/30B, VELAS II -60A/60B/60C' is contained in that of the two types predicate device, and it has been designed to meet the intended use requirement claimed."
    ComponentsLaser system, Color touch screen, Fiber and the handpiece, Foot switch, Power detector/calibratorLaser system, Color touch screen, Fiber, Foot switch, Power detector/calibrator. (GBOX lacks handpiece, VELAS II does not explicitly list handpiece as a separate component distinct from fiber)Similar. "Handle as the laser transmission system, should be used with optical fiber. So GBOX only uses fiber as the laser transmission system does not affect the safety and effectiveness." (Similar statement for VELAS II).
    Power SupplyPredicate K082721: 100-240 VAC, 540VA max, 50/60Hz; Predicate K122567: 230VAC, 50/60Hz; Predicate K100558: 100-240VAC, 50-60Hz, 350VAGBOX: 100-240VAC, 200VA, 50-60Hz; VELAS II-30A/B: 100-240VAC, 50/60Hz, 350VA; VELAS II-60A/B/C: 100-240VAC, 50/60Hz, 650VASimilar. "Only rated power is different, not affect safety and effectiveness."
    SterilityNon-sterile (main engine)Non-sterile (main engine), but fiber is sterile for single use.Similar. "Fiber steriled for single use of GBOX (and VELAS II) is more effective to ensure the safety of patients."
    Standards ComplianceIEC60601-1, IEC60601-1-2, IEC60601-2-22, IEC60825-1, ISO10993-1, ISO10993-4, ISO10993-5, ISO10993-10, ISO10993-11IEC60601-1, IEC60601-1-2, IEC60601-2-22, IEC60825-1, ISO10993-1, ISO10993-5, ISO10993-10 (For VELAS II, also ISO10993-4 and ISO10993-11 are listed in Section 9 Non-Clinical Tests)Same (referring to adherence to these specific standards for safety and performance testing).

    2. Sample size used for the test set and the data provenance

    The document describes non-clinical tests performed on the "GBOX" and "VELAS II" Medical Diode Laser Systems for safety and performance in accordance with various IEC and ISO standards. It also mentions "software verification has been carried out according to the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

    • Test Set: No explicit "test set" in the context of clinical data (e.g., patient cases, retrospective or prospective) is described for evaluating device performance against primary clinical endpoints. The testing described is primarily engineering and bench testing to demonstrate compliance with general safety and performance standards and comparison to predicate device specifications.
    • Sample Size: Not applicable in the context of a clinical test set. The non-clinical tests would have involved specific units of the device.
    • Data Provenance: The tests were performed by the manufacturer, Wuhan Gigaa Optronics Technology Co., Ltd. The document does not specify a country of origin for the non-clinical test data other than the manufacturer's location in China. The "tests" here refer to regulatory compliance and engineering evaluations, not clinical trials with human subjects.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided as the submission is not based on a clinical study requiring ground truth establishment by experts. The "ground truth" for this submission is effectively the established safety and performance profile of the predicate devices and compliance with recognized international standards for medical electrical equipment and laser products.

    4. Adjudication method for the test set

    Not applicable. There is no clinical test set requiring expert adjudication described in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a medical diode laser system, not an AI-powered diagnostic or interpretive tool that would involve "human readers" or "AI assistance" in the sense of image analysis or diagnostic support. Therefore, an MRMC study is irrelevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is not an algorithm or AI system for diagnostic or interpretive purposes. It's a surgical laser system where the performance is physical (e.g., power output, wavelength, tissue interaction) and its operation is directly controlled by a human operator.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this 510(k) submission is primarily:

    • Predicate Device Performance: The established safety and effectiveness of the legally marketed predicate devices to which the proposed devices are compared.
    • Compliance with Recognized Standards: Adherence to international standards for medical electrical equipment, laser safety, and biological evaluation (e.g., IEC 60601 series, IEC 60825-1, ISO 10993 series).
    • Engineering and Performance Specifications: The technical specifications (e.g., wavelength, output power, pulse width) are tested against internal design requirements and predicate device specifications.

    There is no mention of ground truth established via expert consensus, pathology, or outcomes data from a clinical trial for the proposed device itself. The claim of "Substantial Equivalence" relies on demonstrating that the proposed device is sufficiently similar to devices already found safe and effective.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device that requires a "training set."

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML device.