Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K151786

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-09-24

    (85 days)

    Product Code
    Regulation Number
    862.1810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Binding assay for the in vitro quantitation of vitamin B12 in human serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of gastrointestinal malabsorption. The electrochemiluminescence immunoassay "ECLIA" is intended for use on Elecsys and cobas e immunoassay analyzers.

    Vitamin B12 II CalSet is used for calibrating the quantitative Elecsys Vitamin B12 II assay on the Elecsys and cobas e immunoassay analyzers.

    Device Description

    The Elecsys Vitamin B12 II assay employs a competitive test principle using intrinsic factor specific for vitamin B12. Vitamin B12 in the sample competes with the added vitamin B12 labeled with biotin for the binding sites on the ruthenium-labeled intrinsic factor complex.

    Results are determined via a calibration curve which is instrument specifically generated by 2-point calibration and a master curve provided via reagent barcode.

    The reagent working solutions include:

    • the rackpack (kit placed on instrument) .
    • Streptavidin coated microparticles, .
    • Reagent 1 (ruthenium labeled intrinsic factor) and .
    • Reagent 2 (vitamin B12 labeled biotin). .
    • Pretreatment 1 (Dithiothreitol) .
    • Pretreatment 2 (sodium hydroxide, sodium cyanide) .

    The Vitamin B12 II CalSet is a lyophilized human serum matrix with added vitamin B12 in two concentration ranges.
    The CalSet includes:

    • Cal 1 (approximately 250 pg/mL vitamin B12 in a Human serum matrix) .
    • Cal 2 (approximately 1500 pg/mL vitamin B12 in a Human serum matrix) .
    AI/ML Overview

    The document describes the Elecsys Vitamin B12 II Assay and Elecsys Vitamin B12 II CalSet, which are used for in vitro quantitative determination of Vitamin B12 in human serum and plasma. The device is intended for use in the diagnosis and treatment of anemias of gastrointestinal malabsorption.

    Here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    Test CategoryAcceptance CriteriaReported Device Performance and Remarks
    PrecisionElecsys Vitamin B12 II Assay (Candidate Device)
    Within-run (Repeatability)≤ 200 pg/mL: SD ≤ 14 pg/mL > 200 pg/mL: CV ≤ 7 %Cobas e 411:HS 1 (176 pg/mL): 8.86 SD (5.0% CV)HS 2 (405 pg/mL): 13 SD (3.2% CV)HS 3 (960 pg/mL): 19.7 SD (2.1% CV)HS 4 (1230 pg/mL): 27.4 SD (2.2% CV)HS 5 (1940 pg/mL): 40.9 SD (2.1% CV)PCV0 (229 pg/mL): 8.96 SD (3.9% CV)PCV1 (447 pg/mL): 12.2 SD (2.7% CV)PCV2 (934 pg/mL): 20.2 SD (2.2% CV)All results met the acceptance criteria.
    Intermediate Precision≤ 200 pg/mL: SD ≤ 24 pg/mL > 200 pg/mL: CV ≤ 12 %Cobas e 411:HS 1 (176 pg/mL): 12.7 SD (7.2% CV)HS 2 (405 pg/mL): 17.5 SD (4.3% CV)HS 3 (960 pg/mL): 31.0 SD (3.2% CV)HS 4 (1230 pg/mL): 46.4 SD (3.8% CV)HS 5 (1940 pg/mL): 72.6 SD (3.7% CV)PCV0 (229 pg/mL): 12.4 SD (5.4% CV)PCV1 (447 pg/mL): 18.6 SD (4.2% CV)PCV2 (934 pg/mL): 38.4 SD (4.1% CV)All results met the acceptance criteria.
    Limit of Blank (LoB)LoB ≤ 50 pg/mLAchieved. (Specific value not reported in the summary, but stated as meeting criterion.)
    Limit of Detection (LoD)LoD ≤ 100 pg/mLAchieved. (Specific value not reported in the summary, but stated as meeting criterion.)
    Limit of Quantitation (LoQ)LoQ (Imprecision) ≤ 20 % at 150 pg/mLAchieved. (Specific value not reported in the summary, but stated as meeting criterion.)
    Linearity≤ 200 pg/mL: ± 20 pg/mL > 200 pg/mL: ± 10 % (for deviation to higher order polynomial regression) Significant level for deviation to higher order polynomial: 5%Reported that "All results met the predefined acceptance criteria for linearity." and "The linearity results support a claimed measuring range."
    Endogenous Interferences< 200 pg/mL: ± 20 pg/mL of unspiked reference value > 200 pg/mL: ± 10% of unspiked reference valueReported that "Predefined acceptance criterion was met." Claims included in the method sheet were set to the concentration without observed interference.
    Exogenous Interferences - Drugs± 10% of the reference value (unspiked sample)Reported that "No interference with the assay was found." for 16 commonly used pharmaceuticals.
    Exogenous Interferences - AnticoagulantsSlope (Passing/Bablok): 0.9 – 1.1 Intercept (Passing/Bablok): < ± 100 pg/mL Pearson's r (linear regression): ≥ 0.95 Bias at 200 pg/mL: ± 10 %Potential effects assessed by regression analysis. (Specific values not reported in summary, but implied to be acceptable based on overall conclusion of substantial equivalence).
    Method Comparison (vs. Predicate)Slope (Passing/Bablok): 1.00 ± 0.05 Correlation (linear regression) Pearson's r: > 0.95 Intercept (Passing/Bablok): < ± 30 pg/mL Bias at 200 pg/mL: ≤ ± 10 %Reported for Candidate Device vs. Predicate:Slope (Passing/Bablok): 0.984Intercept (Passing/Bablok): -26.7Tau/r: 0.996Bias at 200 pg/mL: -15.0The values align with the criteria.
    Reagent Stability (Recovery)80 - 120 % for PC Varia and human serum samples (for real-time stability)Study 1, 2, and 3: Acceptance criterion for recovery was compared to day 0 value. (Implied to be met as no issues stated). Study 4 (real-time): Ongoing, data available for 0, 7, 10, 13, 19, 25 months; criterion is 80-120% recovery.
    Sample Stability (Recovery)≤ 200 pg/mL: ± 28 pg/mL > 200 pg/mL: ± 14 %Study 1, 2, and 3: Recovery was calculated as percent of the reference value. (Implied to be met as no issues stated).
    Calibrator Reconstitution (Signal Recovery)90 to 110 % signal recovery of the reference material valueAchieved. (Specific values not reported in summary, but stated as meeting criterion.)
    Calibrator Stability (-20°C, 2-8°C, Onboard) (Signal Recovery)90 to 110 % signal recovery of the reference material valueAchieved. (Specific values not reported in summary, but stated as meeting criterion.)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Precision: Not explicitly stated as a separate "test set" sample size but evaluated using "human serum samples" (HS 1-5) and "PreciControl Varia" (PCV0-2). The protocol involved testing 2 replicates per run, 2 runs per day for 21 days for precision.
    • Limit of Blank (LoB): n = 60 LoB measurements (5 replicates, 2 runs per day on 2 instruments over 3 days) using a buffered human serum albumin matrix.
    • Limit of Detection (LoD): n = 60 LoD measurements (5 samples, 2 runs per day on 2 instruments over 3 days) using 5 low-level human serum samples.
    • Limit of Quantitation (LoQ): Minimum of seven human serum samples, analyzed in replicates of 5, one run per day over 5 days.
    • Linearity: Serum samples.
    • Analytical Specificity: Two human serum samples (single donors, native) spiked with potential cross-reactant compounds.
    • Endogenous Interferences: Three human serum samples (single donors, native as well as spiked) for each interfering substance.
    • Exogenous Interferences - Drugs: Two human serum samples (single donors, native) spiked with 16 pharmaceutical compounds.
    • Exogenous Interferences - Anticoagulants: Minimum of 90 serum/plasma pairs per sample material (single donors, native as well as spiked).
    • Method Comparison: 120 human serum samples (all single donors, native as well as spiked) covering the entire measuring range.
    • Reference Range Study: 120 apparently healthy male & female subjects. Data provenance: USA for sample collection, evaluation done at one site in Germany. Samples were native human serum samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is an in-vitro diagnostic (IVD) device for quantitative measurement of a biomarker (Vitamin B12). The "ground truth" for such devices is typically established through analytical methods and highly controlled reference materials, not through expert consensus on qualitative interpretation of images or clinical findings. The document does not describe the use of human experts to establish ground truth for this type of test.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As an IVD device, the performance is assessed through analytical validation studies (precision, linearity, accuracy against reference methods/materials) rather than human expert adjudication of results.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is an in-vitro diagnostic device for measuring Vitamin B12, not an imaging device or AI-assisted diagnostic tool that relies on human readers/interpreters.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device is a standalone in-vitro diagnostic assay (Elecsys Vitamin B12 II assay) intended to quantitatively measure Vitamin B12. Its performance is evaluated analytically, separate from human interpretation of the results for direct diagnostic decision-making, though the results contribute to a physician's overall diagnosis and treatment plan. The studies described (precision, linearity, LoB/LoD/LoQ, interference, method comparison) are all "standalone" evaluations of the device's analytical performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for the analytical and method comparison studies is established based on:

    • Reference methods/materials (e.g., specific calibrators, spiked samples with known concentrations, comparison to the legally marketed predicate device).
    • Defined analytical standards and statistical methods (e.g., CLSI guidelines).
    • The "value assignment" for calibrators is a key part of establishing the reference for the assay.

    8. The sample size for the training set

    This document does not describe a "training set" in the context of machine learning. The studies described are validation and verification studies for an analytical assay. The "Elecsys Vitamin B12 II CalSet" serves as a calibrator for the assay, analogous to the concept of a "calibration set" for an analytical instrument, but not a training set for an AI/ML algorithm.

    9. How the ground truth for the training set was established

    Not applicable as there is no "training set" in the machine learning sense. However, for the calibrator (Elecsys Vitamin B12 II CalSet), its "target values" are assigned to obtain the best fit with the Master Calibration Curve, using multiple analyzers (at least 3 cobas e 411 and at least 3 cobas e 601/MODULAR ANALYTICS E170 analyzers) and multiple reagent lots. The assigned value for each calibrator is the mean value obtained over at least six runs on at least three analyzers. PreciControl Varia is also used to monitor accuracy and precision of analytes, and acceptance criteria for these controls must be met to release the assigned values for the CalSet. This process establishes the "ground truth" for the calibrators.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1