    The Stryker Spine AVS® AL and AVS® ALign PEEK Spacers are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1.

    DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). These patients should be skeletally mature and have six months of nonoperative therapy.

    Additionally, the AVS® AL and AVS® ALign PEEK Spacers can be used as an adjunct to fusion in patients diagnosed with degenerative scoliosis.

    The AVS® AL and AVS® ALign PEEK Spacers are to be implanted via anterior or anterolateral approach.

    The AVS® AL and AVS® ALign PEEK Spacers are intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine.

    The Stryker Spine AVS® PL and AVS® UniLIF™ PEEK Spacers are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1.

    DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). These patients should be skeletally mature and have six months of nonoperative therapy.

    Additionally, the AVS® PL and UniLIF™ PEEK Spacers can be used as an adjunct to fusion in patients diagnosed with degenerative scoliosis.

    The AVS® PL PEEK Spacers and AVS® UniLIF™ PEEK Spacers are to be implanted via posterior approach.

    The AVS® PL PEEK Spacers and AVS® UniLIF™ PEEK Spacers are intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine (i.e., posterior pedicle screw and rod systems).

    The Stryker Spine AVS® TL PEEK Spacers are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1.

    DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). These patients should be skeletally mature and have six months of nonoperative therapy.

    Additionally. the AVS® TL PEEK Spacers can be used as an adjunct to fusion in patients diagnosed with degenerative scoliosis.

    The AVS® TL Peek Spacers are to be implanted via posterior approach.

    The AVS® TL PEEK Spacers are intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine (i.e., posterior pedicle screw and rod systems).

    The Stryker Spine AVS® Navigator PEEK Spacers are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1.

    DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). These patients should be skeletally mature and have six months of nonoperative therapy.

    Additionally, the AVS® Navigator PEEK Spacers can be used as an adjunct to fusion in patients diagnosed with degenerative scoliosis.

    The AVS® Navigator PEEK Spacers are to be implanted via a posterolateral approach.

    The AVS® Navigator PEEK Spacers are intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine.

    The Stryker Spine AVS® ARIA PEEK Spacers are intervertebral body fusion devices indicated for use with autograff and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1.

    DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). These patients should be skeletally mature and have six months of nonoperative therapy.

    Additionally, the AVS® ARIA PEEK Spacers can be used as an adjunct to fusion in patients diagnosed with degenerative scoliosis.

    The AVS® ARIA PEEK Spacers are intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine.

    Device Description
    1. AVS® AL and AVS® ALign PEEK Spacers
      The AVS® AL (Anterior Large) and AVS® ALign PEEK Spacers are intended for use as interbody fusion devices. They are offered in a variety of lengths, heights and lordotic angles. The hollow, ring shaped implant has serrations on the top and bottom for fixation. The spacers are manufactured from PEEK OPTIMA LT1 and include tantalum markers for visualization.
    2. AVS® PL and AVS® UniLIF PEEK Spacers
      The AVS® Partial Lumbar (PL) PEEK Spacers and AVS® UniLIF™ PEEK Spacers are intended for use as an aid in spinal fixation. This hollow, rectangular implant is offered in a variety of lengths, heights and lordotic angles to adapt to a variety of patient anatomies. It has serrations on the superior and inferior surfaces of the implant designed to help with fixation, an ergonomically shaped anterior edge, and a flat posterior edge. Radiopaque markers have been embedded within the implant to help allow for visualization in radiographic images. The spacers are manufactured from PEEK OPTIMA LT1 and include tantalum markers for visualization.
    3. AVS® TL PEEK Spacer
      The AVS® TL PEEK Spacer is intended for use as an aid in spinal fixation. It is offered in both parallel and wedge shapes. The hollow implant has serrations on the top and bottom which are designed to help with fixation. The spacers are manufactured from PEEK OPTIMA LT1 and include tantalum markers for visualization.
    4. AVS® Navigator PEEK Spacer
      The AVS® Navigator PEEK Spacer is intended for use as an interbody fusion device. It is offered in a variety of lengths, heights and lordotic angles. The hollow implant has serrations on the top and bottom for fixation. Radiopaque markers have been embedded within the implant to help allow for visualization in radiographic images. The spacers are manufactured from PEEK OPTIMA LT1 and include tantalum markers for visualization.
    5. AVS® ARIA PEEK Spacer
      The AVS® ARIA PEEK Spacer is intended for use as an interbody fusion device. It is offered in a variety of lengths, heights and lordotic angles. The hollow, oblong-shaped implant has serrations on the top and bottom for fixation. The spacers are manufactured from PEEK OPTIMA LT1 and include tantalum markers for visualization.
    AI/ML Overview

    This document is a 510(k) premarket notification for several intervertebral body fusion devices from Stryker Corporation. The purpose of the submission is to expand the indications for use of these devices to include patients diagnosed with degenerative scoliosis.

    Here's the breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not specify quantitative acceptance criteria or a direct performance measurement for a device's ability to diagnose degenerative scoliosis. Instead, it relies on demonstrating substantial equivalence to predicate devices and leveraging existing clinical data for similar devices to support the expanded indication.

    Acceptance Criterion (Implicit)Reported Device Performance
    Substantial Equivalence to Predicate Devices for new indication (Degenerative Scoliosis)"The design features, materials used, manufacturing, and sterilization methods are identical to the previously cleared Stryker Spine lumbar intervertebral body fusion devices with the exception of broadening the indications to include the use Stryker Spine lumbar intervertebral body fusion devices to treat patients diagnosed with degenerative scoliosis."
    Expanded indication does not adversely affect system performance or represent a new worst-case scenario."Published clinical data for lumbar interbody fusion devices similar to the Stryker Spine lumbar intervertebral body fusion devices that are the subject of this submission was provided in support of this application. The published clinical outcomes demonstrated that the use of the lumbar interbody fusion procedures to treat patients diagnosed with degenerative scoliosis above does not adversely affect performance of the system and does not represent a new worst case scenario."
    No additional implant testing required due to no changes to existing devices."No changes were made to the existing devices; therefore, no additional implant testing was required or performed."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    The document does not describe a specific "test set" in the context of an algorithm or diagnostic device. Instead, it refers to "published clinical data for lumbar interbody fusion devices similar to the Stryker Spine lumbar intervertebral body fusion devices."

    • Sample Size: Not specified. The document refers generally to "published clinical data."
    • Data Provenance: Not specified (e.g., country of origin, retrospective/prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This information is not provided in the document as it's not a study evaluating a diagnostic algorithm. The "ground truth" for the original indication of DDD is defined as "back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies." For degenerative scoliosis, it is implied to be a clinical diagnosis, but the process of establishing this "ground truth" in the context of the referenced clinical data is not detailed.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not applicable/not provided as there is no described test set with human readers making judgments on the expanded indication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable/not provided. The submission is for an expanded indication for a physical intervertebral body fusion device, not an AI-powered diagnostic tool. Therefore, no MRMC study or AI assistance evaluation was conducted or is relevant here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This is not applicable/not provided. The device is a physical implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    For the expanded indication of degenerative scoliosis, the ground truth is implied to be clinical diagnosis as established in the "published clinical data." For the original DDD indication, diagnosis is "confirmed by history and radiographic studies." The document strongly emphasizes "clinical outcomes" from the published data.

    8. The sample size for the training set:

    This is not applicable/not provided. There is no "training set" in the context of an AI/algorithm being developed or tested. The submission relies on existing clinical data and the equivalence of the physical devices.

    9. How the ground truth for the training set was established:

    This is not applicable/not provided for the same reasons as point 8.