    Intended Use

    The Tenodesis screws are intended for soft tissue reattachment procedures in foot surgery.

    The system is indicated for soft tissue reattachment procedures in foot surgery, such as tendon reconstructions and tendon transfers.

    The system can be used in adult patients. Additional information is provided in the corresponding surgical techniques.

    Device Description

    The Tenodesis screw system is intended for soft tissue reattachment procedures in foot surgery. The screws are made of titanium alloy Ti-6AI-4V (ASTM F136). They are threaded, fully cannulated anchors.

    The implants are offered in various sizes. The implants are provided non-sterile and single-use only.

    The Tenodesis instruments are non-sterile and reusable.

    AI/ML Overview

    Based on the provided text, the device in question is a "Tenodesis screw system." This document is a 510(k) submission, which focuses on demonstrating substantial equivalence to pre-existing predicate devices rather than proving performance against specific acceptance criteria in a traditional clinical study format for a novel device. Therefore, many of the requested categories for a study proving acceptance criteria will not be directly applicable or available in this type of submission.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a 510(k) submission, formal "acceptance criteria" for a novel device performance study are not explicitly stated. Instead, the submission relies on demonstrating substantial equivalence to predicate devices through mechanical testing and biocompatibility.

    Acceptance Criteria CategoryReference/MethodReported Device Performance
    Intended UseComparison to predicate devices (G-Force Tenodesis Screw, Arthrex Tenodesis Screw Family, Arthrex Resorbable Interference Screw)The Tenodesis screw system has the same intended use for soft tissue reattachment procedures in foot surgery (tendon reconstructions and tendon transfers in adult patients).
    Technological CharacteristicsComparison to predicate devices regarding design, components, materials, and principles of operation.The system has similar performance characteristics, is manufactured from similar materials (titanium alloy Ti-6AI-4V), and is similar in design to the selected predicate devices.
    BiocompatibilityISO 10993-1Biocompatibility testing on the plates was conducted and evaluated per ISO 10993-1. All testing passed.
    Mechanical PropertiesSelf-tapping performance, driving, torsional properties, and axial pullout strength tests. (Implied comparison to predicate devices, though specific benchmarks are not provided).Self-tapping performance, driving, torsional properties, and axial pullout strength of the Tenodesis screw system and the predicate devices resulted in similar mechanical properties and performance. The subject and predicate devices are substantially equivalent.

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for the mechanical and biocompatibility tests. It states that "Biocompatibility testing on the plates was conducted and evaluated per ISO 10993-1" and "Self-tapping performance, driving, torsional properties and axial pullout strength of the Tenodesis screw system and the predicate devices resulted in similar mechanical properties and performance." The data provenance is from the manufacturer's internal testing as part of the 510(k) submission. No information on country of origin of data or retrospective/prospective status is available for these non-clinical tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable. The evaluations (biocompatibility and mechanical properties) rely on established international standards (ISO 10993-1) and direct physical measurements, not expert interpretation of data points that require a "ground truth" established by human experts in the sense of a clinical trial.

    4. Adjudication method for the test set

    Not applicable, as this is not a human-adjudicated clinical study. The evaluations are based on objective test results against established mechanical and biological performance parameters.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device (screw system), not an AI diagnostic or assistance tool. Therefore, an MRMC study and AI improvement metrics are not relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For biocompatibility, the ground truth is compliance with the requirements of ISO 10993-1. For mechanical properties, the "ground truth" is measured physical properties that are comparable to those of the predicate devices, implying that the established performance of the predicate serves as the benchmark. No clinical "ground truth" (pathology, outcomes data) was used or required for this 510(k) submission.

    8. The sample size for the training set

    Not applicable. There is no "training set" as this is not a machine learning or AI-driven device.

    9. How the ground truth for the training set was established

    Not applicable, for the same reason as point 8.