510(k) Data Aggregation

    K Number
    K151690

    Validate with FDA (Live)

    Date Cleared
    2015-07-16

    (23 days)

    Product Code
    Regulation Number
    862.2570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Alere™ i Influenza A & B:

    The Alere™ i Influenza A & B assay performed on the Alere™ i Instrument is a rapid molecular in vitro diagnostic test utilizing an isothermal nucleic acid amplification technology for the qualitative detection and discrimination of influenza A and B viral RNA in nasal swabs from patients with signs and symptoms of respiratory infection. It is intended for use as an aid in the differential diagnosis of influenza A and B viral infections in humans in conjunction with clinical and epidemiological risk factors. The assay is not intended to detect the presence of influenza C virus.

    Negative results do not preclude influenza virus infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

    Performance characteristics for influenza A were established during the 2012-2013 influenza A/H3 and A/H1N1 pandemic were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary.

    If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

    Alere™ i Strep A:

    Alere™ i Strep A is a rapid, instrument-based, molecular in vitro diagnostic test utilizing isothermal nucleic acid amplification technology for the qualitative detection of Streptococcus pyogenes, Group A Streptococcus bacterial nucleic acid in throat swab specimens obtained from patients with signs and symptoms of pharyngitis. It is intended to aid in the rapid diagnosis of Group A Streptococcus bacterial infections.

    All negative test results should be confirmed by bacterial culture because negative results do not prection with Group A Streptococcus and should not be used as the sole basis for treatment.

    Device Description

    Alere™ i Influenza A & B is a rapid, instrument-based isothermal tests for the qualitative detection and differentiation of influenza A and influenza B from nasal swabs collected from patients presenting with signs and symptoms of respiratory infection. Alere™ i Strep A is a rapid, instrument-based isothermal test for the qualitative detection of Group A Strep from throat swab specimens. Both Alere™ i Influenza A & B and Alere™ i Strep A Systems utilize isothermal nucleic acid amplification technology and are comprised of:

    • . Sample Receiver – single use, disposable containing the elution buffer
    • . Test Base – single use, disposable comprising two sealed reaction tubes, each containing a lyophilized pellet
    • Transfer Cartridge single use, disposable for transfer of the eluted sample to the Test Base, and ●
    • Alere™ i Instrument – repeat use reader

    The reaction tubes in the Alere™ i Influenza A & B Test Base contain the reagents required for amplification of the target nucleic acid and an internal control. Alere™ i Influenza A & B utilizes a pair of templates (similar to primers) for the specific amplification of RNA from influenza A and B and a fluorescently labeled molecular beacon designed to specifically identify the amplified RNA targets.

    The reaction tubes in the Alere™ i Strep A Test Base contain the reagents required for Group A Strep bacterial lysis and the subsequent amplification of the target nucleic acid and an internal control. Alere™ i Strep A utilizes a pair of templates (similar to primers) for the specific amplification of DNA from Group A Strep and a fluorescently labeled molecular beacon designed to specifically identify the amplified nucleic acid target.

    Both assays are performed within the confinement of the Test Base, and no other part of the Alere™ i Instrument has contact with the sample during the amplification process. This reduces the risk of instrument contamination and sample carry-over between measurements.

    To perform the assay, the Sample Receiver and Test Base are inserted into the Alere™ i Instrument and the elution buffer is automatically heated by the instrument. The sample is added to the Sample Receiver and transferred via the Transfer Cartridge to the Test Base, resuspending the lyophilized pellets contained within the Test Base and initiating bacterial lysis (for Alere™ i Strep A) and target amplification. Heating, mixing and detection by fluorescence is provided by the instrument, with results automatically reported.

    Results are displayed by the Alere™ i Instrument and are also stored in an on-board archive and are assigned to a sample ID that has been entered into the Alere™ i Instrument by the operator, and the date/time the test was performed. Data can be retrieved and downloaded by the operator at any time after testing. An external Alere™ Universal Printer can be attached via USB to the Alere™ i Instrument to print test results.

    AI/ML Overview

    The provided document describes the Alere™ i Instrument, Alere™ i Influenza A & B, and Alere™ i Strep A devices. This submission focuses on a software modification to the Alere™ i Instrument to allow it to run both the Influenza A & B and Strep A assays, rather than changes to the assays themselves. Therefore, the "acceptance criteria" and "device performance" in this context refer to demonstrating that the software modification does not negatively impact the performance of the previously cleared assays.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Rationale for the Study: The primary goal of this 510(k) submission (K151690) is to demonstrate that a software modification to the Alere™ i Instrument, which enables it to run both the Alere™ i Influenza A & B and Alere™ i Strep A assays, does not alter the established safety and effectiveness of these assays. The previous versions of these tests (Alere™ i Influenza A & B, K141520 and Alere™ i Strep A, K141757) were already cleared. Therefore, the study focuses on validating the software modification.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of numerical "acceptance criteria" for clinical performance that needed to be met by a de novo study of the device. Instead, the acceptance criterion for this submission is implicitly that the software modification does not change the performance of the previously cleared assays. The reported "performance" for the modified device is that its functionality remains unchanged compared to the predicate devices.

    Acceptance Criterion (Implicit)Reported Device Performance (with software modification)
    No change in the established assay functionality for Alere™ i Influenza A & B due to the software modification.Alere™ i Influenza A & B assay functionality remains unchanged.
    No change in the established assay functionality for Alere™ i Strep A due to the software modification.Alere™ i Strep A assay functionality remains unchanged.
    The Alere™ i Instrument with modified software performs equivalently to the predicate devices for both assays.The Alere™ i Influenza A & B and Alere™ i Strep A tests performed on the Alere™ i Instrument containing modified software are substantially equivalent to the current legally marketed devices (Alere™ i Influenza A & B and Alere™ i Strep A performed on the Alere™ i Instrument).
    All technological characteristics (e.g., FDA Product Code, Assay Target, Intended Use, Instrumentation, Sample Type, Technology, etc.) remain identical to the predicate device.All listed technological characteristics are "Same" as the predicate devices, with the exception of the "Instrument software" which is the subject of the modification.

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "Software verification and validation studies performed demonstrated that Alere™ i Influenza A & B and Alere™ i Strep A assay functionality remains unchanged due to this change."

    However, no specific sample sizes for a "test set" (e.g., number of patient samples, artificial samples) used in these software validation studies are provided in the excerpt. The data provenance (country of origin, retrospective/prospective) is also not detailed for these specific software validation studies. Given that the submission asserts "no changes to the Alere™ i Influenza A & B or Alere™ i Strep A tests," it's highly probable that the software validation primarily involved testing with internal samples or simulated data to ensure proper functionality and integration, rather than a new full-scale clinical study with patient samples. The information indicates the tests themselves had established performance characteristics from previous submissions.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Since the study described here is focused on software validation and demonstrating unchanged functionality of existing assays, rather than determining the initial clinical accuracy of the assays against a clinical ground truth, there is no mention of experts establishing a ground truth for a test set in this context.


    4. Adjudication Method for the Test Set

    As there is no mention of a clinical test set requiring expert ground truth, no adjudication method is described.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was mentioned. The device is an in vitro diagnostic test, not an AI software intended to assist human readers in image interpretation or a similar task. It is a standalone instrument that provides qualitative diagnostic results.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the performance of the Alere™ i Instrument is inherently standalone. The device processes the sample and provides an automated qualitative result ("Influenza A Detect", "Influenza B Detect", "Strep A Detect", or "Not Detected"). The "Software verification and validation studies" described would have assessed this standalone performance with the modified software.


    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the original clinical performance studies of the Alere™ i Influenza A & B (K141520) and Alere™ i Strep A (K141757) assays, the ground truth would typically involve:

    • For Influenza A & B: Viral culture as the gold standard, or potentially a highly multiplexed PCR reference method, on patient nasal swab samples.
    • For Strep A: Bacterial culture as the gold standard on patient throat swab samples.

    However, for this specific submission (K151690), the "software validation" would likely use pre-characterized samples (e.g., positive and negative control samples, spiked samples with known concentrations of nucleic acid targets) or previously run clinical samples with known results from predicate studies, to verify that the software processes them identically and yields the same results. The document does not specify the ground truth for these software validation studies, but it would not be a new clinical ground truth establishment.


    8. The Sample Size for the Training Set

    No training set is mentioned in the context of this software modification. The assays themselves are isothermal nucleic acid amplification tests, not machine learning algorithms that require a training set in the typical sense. Any "training" would have been part of the initial development and optimization of the assays prior to the K141520 and K141757 submissions.


    9. How the Ground Truth for the Training Set was Established

    Not applicable, as no training set for a machine learning model is described.