510(k) Data Aggregation

    K Number
    K151557

    Validate with FDA (Live)

    Date Cleared
    2015-07-16

    (36 days)

    Product Code
    Regulation Number
    862.3640
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EGENS Urine Test Cup Morphine - Methamphetamine is a rapid test for the qualitative detection of Morphine and Methamphetamine in human urine at a cutoff concentration of 300 ng/mL and 1000 ng/mL, respectively.

    EGENS Urine Test Cup Morphine - Methamphetamine test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

    For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

    The EGENS Urine Test DipCard Morphine - Methamphetamine is a rapid test for the qualitative detection of Morphine and Methamphetamine in human urine at a cutoff concentration of 300 ng/mL and 1000 ng/mL, respectively.

    EGENS Urine Test DipCard Morphine - Methamphetamine test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

    Device Description

    EGENS Urine Test Morphine-Methamphetamine test uses immunochromatographic assays for Morphine and Methamphetamine. The test is a lateral flow system for the qualitative detection of Morphine and Methamphetamine in human urine. The test is the first step in a two-step process. The second step is to send the sample for laboratory testing if preliminary positive results are obtained.

    AI/ML Overview

    The provided document describes the performance characteristics of the EGENS Urine Test Cup Morphine - Methamphetamine and EGENS Urine Test DipCard Morphine - Methamphetamine devices.

    Here's an analysis based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state formal "acceptance criteria" but rather presents performance characteristics from various studies. For the purpose of this response, I infer the reported performance to be the intended criteria and thus list the results.

    Morphine (MOP) and Methamphetamine (MET) Detection
    Cut-off for Morphine: 300 ng/mL
    Cut-off for Methamphetamine: 1000 ng/mL

    Performance CharacteristicAcceptance Criteria (Inferred from data)Reported Device Performance - Morphine DipCardReported Device Performance - Methamphetamine DipCardReported Device Performance - Morphine CupReported Device Performance - Methamphetamine Cup
    Precision
    At -100% cut-off100% negative100% negative (50-/0+)100% negative (50-/0+)100% negative (50-/0+)100% negative (50-/0+)
    At -75% cut-off100% negative100% negative (50-/0+)100% negative (50-/0+)100% negative (50-/0+)100% negative (50-/0+)
    At -50% cut-off100% negative100% negative (50-/0+)100% negative (50-/0+)100% negative (50-/0+)100% negative (50-/0+)
    At -25% cut-off100% negative100% negative (50-/0+)100% negative (50-/0+)100% negative (50-/0+)100% negative (50-/0+)
    At cut-offMixed positive/negative (e.g., ~50% positive)62-66% positive (31-33+/17-19-) across 3 lots60-64% positive (30-32+/18-20-) across 3 lots62-66% positive (31-33+/17-19-) across 3 lots60-62% positive (30-31+/19-20-) across 3 lots
    At +25% cut-off100% positive100% positive (50+/0-)100% positive (50+/0-)100% positive (50+/0-)100% positive (50+/0-)
    At +50% cut-off100% positive100% positive (50+/0-)100% positive (50+/0-)100% positive (50+/0-)100% positive (50+/0-)
    At +75% cut-off100% positive100% positive (50+/0-)100% positive (50+/0-)100% positive (50+/0-)100% positive (50+/0-)
    At +100% cut-off100% positive100% positive (50+/0-)100% positive (50+/0-)100% positive (50+/0-)100% positive (50+/0-)
    Cut-off Studies
    All negative at -25% & -50% cut-off100% negative100% negative100% negative100% negative100% negative
    All positive at +25% & +50% cut-off100% positive100% positive100% positive100% positive100% positive
    InterferenceNo interference from listed compoundsNo interference at 100ug/mL for listed compoundsNo interference at 100ug/mL for listed compoundsNo interference at 100ug/mL for listed compoundsNo interference at 100ug/mL for listed compounds
    Cross-Reactivity (Morphine)Various cross-reactivity percentages listedAchieved for listed compoundsN/AAchieved for listed compoundsN/A
    Cross-Reactivity (Methamphetamine)Various cross-reactivity percentages listedN/AAchieved for listed compoundsN/AAchieved for listed compounds
    Effect of Specific Gravity and Urine pHNo effect on device performanceNo effect observedNo effect observedNo effect observedNo effect observed
    Laboratory Comparison Study (agreement with GC/MS)Not explicitly stated, but high agreement expected for samples not near cut-offHigh agreement (e.g., 100% outside near-cutoff range)High agreement (e.g., 100% outside near-cutoff range)High agreement (e.g., 100% outside near-cutoff range)High agreement (e.g., 100% outside near-cutoff range)
    Lay-User Study (% Agreement with GC/MS)High agreement for samples outside near-cutoff rangeMorphine Dipcard: 90-100%Methamphetamine Dipcard: 90-100%Morphine Cup: 85-100% Methamphetamine Cup: 85-100%

    2. Sample Size and Data Provenance

    • Precision Studies: For each concentration level (-100%, -75%, -50%, -25%, cut-off, +25%, +50%, +75%, +100% cut-off), 50 tests were performed per lot. With 3 lots and 9 concentrations, this is 50 * 3 * 9 = 1350 tests per drug for each device format (Cup/DipCard).
      • Data Provenance: Not explicitly stated, but the studies were conducted in-house. This typically implies controlled lab conditions rather than real-world patient samples, but the samples themselves are referenced as "samples with concentrations."
    • Cut-off Studies: A total of 125 morphine samples and 125 methamphetamine samples were used, equally distributed across 5 concentration levels (-50%, -25%, cut-off, +25%, +50%).
      • Data Provenance: Not explicitly stated, but implied to be controlled, spiked samples.
    • Interference/Specificity Studies: Samples were prepared by adding potential interfering substances or drug metabolites to urine containing target drugs. The quantity of samples is not explicitly given but implied to be sufficient for testing (e.g., "These samples were tested using three batches...").
      • Data Provenance: Not explicitly stated, likely controlled lab conditions.
    • Effect of Specific Gravity and Urine pH: 12 urine samples covering normal, high, and low specific gravity ranges, and urine pH adjusted to 4.00-9.00 in 1 pH unit increments. These were spiked samples.
      • Data Provenance: Not explicitly stated, likely controlled lab conditions.
    • Laboratory Comparison Studies: 80 "unaltered clinical samples" (40 negative and 40 positive) for each drug (Morphine and Methamphetamine) per device format (DipCard, Cup). This means 80 * 2 (drugs) * 2 (formats) = 320 samples in total for this part of the study.
      • Data Provenance: "Clinical samples" are explicitly stated, implying these are from actual patients rather than purely synthetic or spiked lab samples. The country of origin is not specified but the study was performed "in-house."
    • Lay-User Study: 560 lay persons. The study used 80 "drug-free samples", 240 "morphine samples", and 240 "methamphetamine samples". Within these, the document describes 20 samples per concentration level (-100%, -75%, -50%, -25%, +25%, +50%, +75% of cut-off) per drug.
      • Data Provenance: Urine samples were "spiked drugs into drug free-pooled urine specimens." So, these were controlled, spiked samples, not unalterted clinical samples.

    3. Number of Experts and their Qualifications for Ground Truth

    • Laboratory Comparison Studies: "Three laboratory assistants" were involved in running the tests. Their qualifications are not specified beyond being "laboratory assistants."
    • Precision, Cut-off, Interference, Specific Gravity/pH Studies: "Three operators" repeatedly tested the samples. Their qualifications are not specified.
    • Lay-User Study: The "lay persons" were the "users" in this study, not "experts" establishing ground truth. There were 560 lay users. Their qualifications were described as having "diverse educational and professional backgrounds and ranged in age from 21 to >50 years."

    4. Adjudication Method for the Test Set

    • Laboratory Comparison Studies: The collected results from the three viewers/operators were directly compared to the GC/MS results. There is no explicit mention of an adjudication process (e.g., 2+1, 3+1) among the viewers to establish a consensual device output. Each viewer's result was compared to GC/MS independently.
    • Other Studies (Precision, Cut-off, Interference, etc.): The results were aggregated from multiple operators/lots, but the tables present aggregated counts (e.g., 50-/0+, 33+/17-), not individual operator adjudication.
    • Lay-User Study: Each participant was given one blind-labeled sample and a device. Their individual results were recorded and compared to the GC/MS ground truth. No adjudication between lay users is mentioned.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    There is no mention of a formal MRMC comparative effectiveness study or analysis of how much human readers improve with AI vs. without AI assistance. The study involves human readers (operators/lay users) interpreting the results of the device, but it does not compare their performance with and without an AI or advanced assistance system. It is a standalone device performance comparison against GC/MS.

    6. Standalone Performance Study (Algorithm only without Human-in-the-loop)

    The devices described (EGENS Urine Test Cup and DipCard) are rapid, qualitative immunoassay tests that rely on visual interpretation of results (presence or absence of colored lines). They are not "algorithms" in the computational sense, and there is no "algorithm only" or "standalone" performance without human-in-the-loop, as human interpretation of the visual result is integral to the device's function. The "laboratory comparison study" and "lay-user study" assess the device's performance when interpreted by humans.

    7. Type of Ground Truth Used

    • Analytical Performance Studies (Precision, Cut-off, Interference, Specificity, Specific Gravity/pH): The ground truth was established by precise spiking of known concentrations of the target drug or interfering substances into drug-free urine.
    • Laboratory Comparison Studies: GC/MS (Gas Chromatography/Mass Spectrometry) was used as the preferred confirmatory method and thus, the primary ground truth reference.
    • Lay-User Study: GC/MS was used to confirm the concentrations of the spiked samples, serving as the ground truth.

    8. Sample Size for the Training Set

    The document describes premarket notification (510(k)) studies to demonstrate substantial equivalence, not the development of a predictive model or algorithm that would typically involve a "training set." Therefore, no training set size is mentioned or applicable in the context of these device types and studies. The tests are immunoassays, not machine learning models.

    9. How the Ground Truth for the Training Set Was Established

    As stated above, this document does not describe the development of a device requiring a "training set" for a machine learning algorithm. The "training set" concept is not applicable here.