Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K151499

    Validate with FDA (Live)

    Date Cleared
    2016-03-18

    (288 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CareTaker is intended to noninvasively and continuously measure a patient's blood pressure ("BP") and heart rate ("HR"), which are derived from the pulse pressure waveform using the scientific method of Pulse Decomposition Analysis ("PDA") for use on adult patients at rest. CareTaker is calibrated using a manual sphygmomanometer. All parameters derived by CareTaker are reported to a remote display monitor via standard radio transmission. CareTaker does not provide any physiological alarm functions. The device is intended for use by clinicians or other properly trained medical personnel in a hospital or other appropriate clinical settings.

    Device Description

    CareTaker is a cardiovascular monitoring device that noninvasively measures continuous blood pressure and heart rate via a finger cuff based on the scientific method of Pulse Decomposition Analysis ("PDA")

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the CareTaker Wireless Vital Signs Monitor, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Parameter / CriteriaAcceptance Criteria (Standard)Reported Device Performance
    Blood PressureAAMI SP-10 (AAMI/ANSI/ISO 81060) guidelinesCompliant with ISO 81060-2 standard
    SafetyIEC 60601-1 (general medical electrical equipment safety)Verified safe
    ElectromagneticIEC 60601-1-2 (electromagnetic compatibility of medical electrical equipment)Verified electromagnetically compatible

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated in the document for the clinical study. It only mentions "a clinical study."
    • Data Provenance: Not explicitly stated regarding country of origin or whether it was retrospective or prospective. It is described as "a clinical study to compare with an arterial catheter."

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not explicitly stated in the document. The ground truth was established by comparison to an "arterial catheter," which is a direct and invasive measurement of blood pressure, generally considered the gold standard.

    4. Adjudication Method for the Test Set

    Not applicable. The ground truth was based on arterial catheter measurements, not expert adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the Effect Size

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed or described in this document. This device is a measurement tool, not an interpretive diagnostic system typically requiring MRMC studies for human reader performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the device's performance was evaluated in a standalone manner. The clinical study compared the device's measurements directly against an arterial catheter, without human-in-the-loop interpretation impacting the primary performance metrics (blood pressure accuracy). The device does require trained personnel to operate it, but its core measurement algorithm's accuracy was tested independently.

    7. The Type of Ground Truth Used

    The ground truth for the clinical study was established by direct comparison to an arterial catheter. This is considered a gold standard for blood pressure measurement, moving beyond expert consensus to direct physiological measurement.

    8. The Sample Size for the Training Set

    Not applicable/not stated. The document describes a post-market notification (510(k)) where the device's performance is being validated, not the development or training of its underlying algorithm. The "Pulse Decomposition Analysis ('PDA')" is a scientific method, implying it's a pre-established analytical approach, not necessarily a machine learning model that requires a distinct training set in the context of this regulatory document.

    9. How the Ground Truth for the Training Set was Established

    Not applicable/not stated, for the same reasons as #8. The PDA method itself would have been developed and validated through scientific research, but this document focuses on the device's compliance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1