510(k) Data Aggregation

    K Number
    K151031

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-06-24

    (68 days)

    Product Code
    Regulation Number
    888.3075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Virage OCT Spinal Fixation System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1-C7) and the thoracic spine from T1-T3; traumatic spinal fractures and/or traumatic dislocations; instability of deformity; failed previous fusions (e.g., pseudoarthorsis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The Virage OCT Spinal Fixation System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advance stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

    In order to achieve additional levels of fixation, The Virage OCT Spinal Fixation System may be connected to the Instinct Java and Sequoia Spinal Systems offered by Zimmer Spine, using rod connectors and transition rods. Refer to the Instinct Java and Sequoia Spinal System package insert for a list of the system specific indications of use.

    The titanium SONGER® Spinal Cable System to be used with the Virage OCT Spinal Fixation System allows for cable attachment to the posterior cervical or thoracic spine.

    Device Description

    The Zimmer Spine Virage® OCT Spinal Fixation System intended for the Occipital-Cervical-Thoracic spine (Occiput-T3). The system consists of a variety of rods, anchors, screws and connectors to build a spinal construct as necessary for the individual patient. The system also includes the instruments necessary for inserting and securing the implants are intended to be removed after solid fusion has occurred.

    The Virage System implants are fabricated from medical grade titanium alloy or medical grade cobalt chromium allov. Implants made from medical grade titanium, medical grade titanium alloy, and medical grade cobalt chromium may be used together. Never use titanium alloy, and/or cobalt chromium with stainless steel in the same construct. All implants are single use only and should not be reused under any circumstances.

    AI/ML Overview

    This document is a 510(k) summary for the "Virage OCT Spinal Fixation System," a medical device. It does not describe a study involving algorithms, AI, or human readers. Instead, it focuses on demonstrating substantial equivalence to a predicate device through mechanical testing, sterilization, and biocompatibility assessments. Therefore, I cannot provide information for most of your requested points, as they are related to studies on AI/algorithm performance.

    However, I can extract the information related to performance testing and acceptance criteria as described in the document.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (What the device must demonstrate)Reported Device Performance (How the device performed)
    Axial and Torsional Gripping per ASTM F1798Device performed as intended.
    Static Torsion, Static and Dynamic Compression Bending per ASTM F1717Device performed as intended.
    SterilizationAssessed and determined to be substantially equivalent to predicate.
    BiocompatibilityAssessed and determined to be substantially equivalent to predicate.

    Summary of the study conducted to prove the device meets acceptance criteria:

    The study was a series of non-clinical, mechanical performance tests, along with assessments for sterilization and biocompatibility.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not specify the exact number of units or samples used for each test (e.g., how many open rod connectors were tested for axial gripping). It states "Performance mechanical testing concluded the Virage open rod connector performed as intended."
    • Data Provenance: Not applicable as this is not a study involving human data or retrospective/prospective clinical data. The tests are laboratory-based mechanical testing and material assessments.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable. The "ground truth" in this context is defined by the standards (ASTM F1798, ASTM F1717) themselves and the predefined pass/fail criteria within those standards for mechanical performance, and established methods for assessing sterilization and biocompatibility. There are no "experts" establishing ground truth in the sense of medical diagnosis or interpretation for this type of device submission.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This is not applicable as there is no human interpretation or subjective assessment being adjudicated. The tests are objective mechanical and material properties.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The document describes a spinal fixation system, not an AI or imaging device with readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This is not applicable. The device is a physical spinal implant system, not an algorithm.

    7. The type of ground truth used:

    The "ground truth" for the performance testing was based on:

    • Engineering standards and recognized test methodologies (ASTM F1798 for axial and torsional gripping, ASTM F1717 for static torsion, static and dynamic compression bending).
    • Established scientific principles and regulatory requirements for sterilization and biocompatibility.

    8. The sample size for the training set:

    This is not applicable. There is no concept of a "training set" for this type of mechanical and material performance testing of a physical medical device.

    9. How the ground truth for the training set was established:

    This is not applicable. The concept of "ground truth for a training set" is relevant to machine learning or AI models, which are not involved in this device's submission.