Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K150920

    Validate with FDA (Live)

    Device Name
    truFreeze System
    Manufacturer
    Date Cleared
    2015-06-17

    (72 days)

    Product Code
    Regulation Number
    878.4350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The truFreeze® System is intended for cryogenic destruction of tissue requiring either active or passive venting during surgical procedures.

    The truFreeze® System is Indicated for use as a cryosurgical tool in the fields of dermatology, gynecology, and general surgery, to ablate benign and malignant

    Device Description

    The truFreeze system is a cryosurgical tool that applies medical-grade liquid nitrogen to the treatment area via a small, low pressure, open tipped catheter. The truFreeze System consists of a console and a disposable spray kit.

    Console:
    The console is the central interface of the system and is comprised of a touch panel computer (TPC) and cryogen, suction, and electronics modules packaged in a mobile cart, Users interact with the console through a dual foot pedal and a touch panel. An off-the-shelf controller and associated software manage the cryogen level sensing, filling, pressure, cooling, defrost, suction, timing and data management functions. A wireless remote control provides alternative timer control from a distance in the treatment room. A fill kit, stored on the rear of the console, allows for liquid nitrogen transfer from the source tank to the console. Safety features include indicators, tank pressure relief valves, an isolated low voltage power system, and an emergency button to be used in the event of user or technical malfunction.

    Disposable spray kit:
    There are no proposed changes to the disposable spray kit. The truFreeze disposable spray kit continues to consist of five (5) individually packaged sterile single-use catheters (7 Fr Straight Tip Catheter and one Catheter Introducer) and 5 individually packaged cryogen decompression tubes (CDTs) (each containing one Dual Lumen 20 Fr CryoDecompression Tube (CDT), Connector, and Suction Tubing). The catheter is flexible and capable of retroflex in a scope. The CDT and accessory tubes are included for use with the onboard suction system.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for a medical device called the truFreeze® System (K150920). This submission is for a software modification to an existing device (K143625). The document focuses on demonstrating that the updated device is substantially equivalent to its predicate.

    Here's an analysis based on your request, extracting the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a quantitative manner (e.g., minimum sensitivity or specificity). Instead, it describes various tests performed to ensure the software modification did not negatively impact the device and added new safety features. The "acceptance criteria" can be inferred as successful completion of these tests without introducing new safety or performance concerns.

    Test TypeAcceptance Criteria (Inferred from study description)Reported Device Performance
    Software Regression TestsNo negative impact on existing code functionality. All previous functionalities maintained.All tests passed, verifying that updates did not negatively impact the code. No additional safety or functionality concerns were observed.
    Software Integration TestsAbility of the software to correctly detect and respond to cDAQ related hardware and communication errors.All integration tests completed and considered acceptable. Demonstrated ability to correctly detect (or not) cDAQ related hardware and communication errors and respond accordingly.
    System Functional TestsNo negative impact on system performance and functionality. System functions as intended per software specifications.All tests successfully performed and passed. Confirmed software updates did not negatively impact system performance and functionality.
    Specific Software FunctionsSoftware functions as intended per software performance specifications. All input and output software values are as specified. System functions as intended. Software functions as intended when a cDAQ hardware error is present or not present.All items successfully verified and validated.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states:

    • "Two sets of testing were performed for each of the load packages available to customers."
    • "Three consoles were successfully loaded with the respective load packages."

    This implies a very small sample size for the "test set" (e.g., 3 consoles). The data provenance is prospective in the sense that the testing was conducted specifically for this 510(k) submission on the modified device. The country of origin of the data is not explicitly stated, but the company is based in Lexington, Massachusetts, USA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The testing described is primarily focused on software and system functionality, not on diagnostic accuracy requiring expert interpretation of clinical data.

    4. Adjudication Method for the Test Set

    This information is not provided and is not relevant for the type of software/system functional testing described. The "ground truth" for these tests would be the expected software behavior and system response to various inputs and error conditions.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. The device is a cryosurgical tool, and the submission concerns a software update to improve safety by detecting communication errors, not an AI-assisted diagnostic or interpretive tool that would involve human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is a standalone assessment of the software's functionality and its impact on the system. The "algorithm" (software control) was tested for its ability to detect and respond to communication interruptions. The performance is assessed without a human-in-the-loop in the context of the software's detection capability. However, the overall device (truFreeze® System) is a tool used by a clinician, so there is always a human-in-the-loop for its clinical application.

    7. The Type of Ground Truth Used

    The ground truth used for the software and system testing was based on:

    • Software performance specifications: The expected design and functionality of the software.
    • Intended system behavior: How the system is designed to operate, especially its response to error conditions (e.g., closing valves upon error detection).
    • Predicate device's established performance: Ensuring the modified device maintained equivalence to the predicate for unchanged aspects.

    It is not expert consensus, pathology, or outcomes data, as these are not relevant to the described software modification.

    8. The Sample Size for the Training Set

    The document describes software updates and validation, not a machine learning model that requires a "training set." Therefore, a training set sample size is not applicable.

    9. How the Ground Truth for the Training Set Was Established

    As there is no training set for a machine learning model, this question is not applicable. The "ground truth" for the software development and testing was established through engineering specifications and design requirements.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1