510(k) Data Aggregation

    K Number
    K150866

    Validate with FDA (Live)

    Date Cleared
    2015-07-17

    (107 days)

    Product Code
    Regulation Number
    878.4010
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SURGISEAL Topical Skin Adhesive is intended for topical applications only to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, simple, thoroughly cleansed, trauma induced lacerations.

    SURGISEAL may be used in conjunction with, but not in place of, deep dermal sutures.

    Device Description

    Adhesive containing a monomeric (2-octyl cyanoacrylate) formulation and the colorant D&C Violet #2. It incorporates the Teardrop applicator that consists of a thermoformed blister tray with a heat sealed lid with an attached applicator sponge tip. When SURGISEAL is applied to the skin with the applicator, it polymerizes in minutes.

    AI/ML Overview

    This document describes the 510(k) premarket notification for the SURGISEAL® Topical Skin Adhesive. The submission primarily addresses a device modification due to a shift in supplier for an equivalent patient-contacting material. Therefore, the "study" mainly focuses on demonstrating that the new material does not negatively impact the device's performance or safety compared to the predicate device.

    Here's an analysis based on the provided text, focusing on acceptance criteria and the supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific TestAcceptance Criteria (Implied/Explicit)Reported Device Performance
    BiocompatibilityCytotoxicity (ISO 10993, Part 5)Non-cytotoxicNew applicator material tested per ISO-10993, Part 5. Results are implicitly acceptable as the conclusion states substantial equivalence.
    Acute Systemic Injection (ISO 10993, Part 11)No adverse systemic reactionsNew applicator material tested per ISO-10993, Part 11. Results are implicitly acceptable as the conclusion states substantial equivalence.
    Guinea Pig Maximization Sensitization (ISO 10993)Not a sensitizerNew applicator material tested per ISO-10993. Results are implicitly acceptable as the conclusion states substantial equivalence.
    Intracutaneous Irritation Test (ISO 10993, Part 10)Non-irritatingNew applicator material tested per ISO-10993, Part 10. Results are implicitly acceptable as the conclusion states substantial equivalence.
    Performance TestingFilm Thickness determinationsConsistent film thickness with the predicate device (implied, as the modification is for an "equivalent patient contacting material")Conducted for the new material component. Results are implicitly acceptable as the conclusion states substantial equivalence.
    FlexibilityEquivalent flexibility to the predicate device (implied, as the modification is for an "equivalent patient contacting material")Conducted. Results are implicitly acceptable as the conclusion states substantial equivalence.
    Linear and Surface CoverageEquivalent linear and surface coverage to the predicate device (implied, as the modification is for an "equivalent patient contacting material")Conducted. Results are implicitly acceptable as the conclusion states substantial equivalence.
    Set-TimeEquivalent set-time to the predicate device (implied, as the modification is for an "equivalent patient contacting material")Conducted. Results are implicitly acceptable as the conclusion states substantial equivalence.
    ViscosityEquivalent viscosity to the predicate device (implied, as the modification is for an "equivalent patient contacting material")Conducted. Results are implicitly acceptable as the conclusion states substantial equivalence.
    In vitro Wound Closure StrengthEquivalent wound closure strength to the predicate device (implied, as the modification is for an "equivalent patient contacting material")Conducted. Results are implicitly acceptable as the conclusion states substantial equivalence.
    Sterilization & Shelf-LifeSterilization validationSterilized in accordance with ISO 11137-2:2006 (EO and gamma irradiation validated)Sterilized by ethylene oxide and gamma irradiation in accordance with ISO 11137-2:2006. Results are implicitly acceptable as the conclusion states substantial equivalence.
    Stability and Shelf LifeSame acceptance criteria as the predicate device (implies no significant degradation or loss of function over the stated shelf life)Verified with the same acceptance criteria as the predicate. Results are implicitly acceptable as the conclusion states substantial equivalence.

    Note: The document states that the new material is "equivalent" and that the modified device is "substantially equivalent to the predicate." Therefore, the acceptance criteria for most tests are implicitly that the performance of the modified device (with the new material) must be comparable to or within acceptable limits of the predicate device. Specific numerical targets are not provided in this summary.


    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not specify the sample sizes used for the various biocompatibility and performance bench tests.

    The data provenance is from bench testing (in vitro) rather than clinical studies, and the location of this testing is not specified, though it would typically be conducted by the manufacturer or a contract research organization. There is no mention of country of origin for data or whether it was retrospective or prospective, as these are typically considerations for clinical data.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable as the studies described are bench tests (biocompatibility and performance) and not studies requiring expert interpretation or ground truth establishment based on medical assessment. The "ground truth" for these types of tests is generally defined by standardized methods and analytical measurements.


    4. Adjudication Method for the Test Set

    This information is not applicable as the studies described are bench tests and do not involve human interpretation or adjudication in the medical sense.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not performed. This submission is for a medical device (topical skin adhesive), not an AI/imaging device.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    This information is not applicable as the device is a topical skin adhesive, not an algorithm or AI system.


    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The concept of "ground truth" as typically applied to clinical or diagnostic studies (e.g., pathology, outcomes data, expert consensus) is not applicable here. For the bench tests described:

    • Biocompatibility: Ground truth is established by adherence to ISO 10993 standards and accepted biological responses in specific in vitro and in vivo models.
    • Performance Testing: Ground truth is established by standardized material testing methods and measurements (e.g., precise measurement of film thickness, set-time, viscosity, wound closure strength).
    • Sterilization and Shelf-Life: Ground truth is established by adherence to ISO 11137-2:2006 and established stability testing protocols.

    8. The Sample Size for the Training Set

    This information is not applicable as the device is not an AI/machine learning system that requires a "training set."


    9. How the Ground Truth for the Training Set was Established

    This information is not applicable as the device is not an AI/machine learning system.