Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K150842

    Validate with FDA (Live)

    Date Cleared
    2015-04-28

    (29 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    JUSHA-M53 Medical Display is intended to be used in displaying and viewing digital images including those of digital mammography for review and analysis by trained medical practitioners.

    Device Description

    JUSHA-M53 Medical Display is a display system with the high resolution (2560 x 2048), high luminance (1000 cd/m²), and 1024 simultaneous shades of gray out of a palette of 4096, 8 DICOM look up table with 12 bit inside. It is compliant with DICOM Part 14, and it has front sensor, presents induction system and ambient brightness adapting function. JUSHA-M53 employs energy efficient LED as backlight, it can achieve a high brightness and low power consumption at the same time. LED backlights deteriorate more slowly and thus the monitor offers a long service time.JUSHA-M53 with 10 bit (1024 tones) simultaneous grayscale display extends graysacle fidelity to the boundaries of human visual perception abilities and helps radiologists discern the finest nuances within an image.

    The product is consisted of the following components:

    • 21.3 inch, mono-TFT Liquid Crystal Display
    • -JUSHA-45T Motherboard/FR-4/REV1.0
    • JUSHA-M53 Medical Display software
    • Power Adapter
    • Data Cable.
    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information based on the provided text, focusing on what is present and what is not explicitly detailed:

    Device: JUSHA-M53 Medical Display

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state formal "acceptance criteria" for the JUSHA-M53 beyond general compliance with industry standards and substantial equivalence to the predicate device. However, a detailed comparison table of technological characteristics to the predicate device (Eizo Nanao Corporation's RadiForce GX540, K130336) is provided, implying these are the performance metrics against which "acceptance" for substantial equivalence is being measured.

    Feature / MetricPredicate Device (RADIFORCE GX540)Proposed Device (JUSHA-M53 Medical Display)Performance (as reported for JUSHA-M53)Acceptance Criteria (Implied from Predicate Comparison)Met (Y/N)
    Screen technology21.3" TFT Monochrome LCD Panel21.3" TFT Monochrome LCD Panel21.3" TFT Monochrome LCD PanelSame as predicateY
    Viewing angle (H,V)Horizontal 176°, Vertical 176°Horizontal 176°, Vertical 176°Horizontal 176°, Vertical 176°Same as predicateY
    Resolution2048 x 25602048 x 25602048 x 2560Same as predicateY
    Display area337.9 x 422.4 mm337.9 x 422.4 mm337.9 x 422.4 mmSame as predicateY
    Aspect ratio4:54:54:5Same as predicateY
    Contrast Ratio (typ)1200:11200:11200:1Same as predicateY
    Recommended brightness500cd/m²500cd/m²500cd/m²Same as predicateY
    Pixel Pitch0.165 x 0.165 mm0.165 x 0.165 mm0.165 x 0.165 mmSame as predicateY
    BacklightingLED backlightLED backlightLED backlightSame as predicateY
    Grayscale102410241024Same as predicateY
    Response time25ms25ms25msSame as predicateY
    SensorsBacklight, Integrated front, Presence, Ambient lightBacklight, Integrated front, Presence, Ambient lightBacklight, Integrated front, Presence, Ambient lightSame as predicateY
    Input signalsDVI standard 1.0DVI standard 1.0DVI standard 1.0Same as predicateY
    Input termination1 DVI-D Dual Link, 1 Display Port1 DVI-D Dual Link, 1 Display Port1 DVI-D Dual Link, 1 Display PortSame as predicateY
    Display controllerOff the shelfOff the shelfOff the shelfSame as predicateY
    Scanning Freq. (H,V) (Predicate)31-135 kHz / 24-61 Hz123 KHz / 60 Hz123 KHz / 60 HzTo be demonstrated as acceptable for medical imagingY (Implied)
    Power Requirement100-120 Vac, 50/60 Hz, 1.1-0.9A; 200-240 Vac, 50/60 Hz, 0.6-0.5A100-120 Vac, 50/60 Hz, 1.1-0.9A; 200-240 Vac, 50/60 Hz, 0.6-0.5A100-120 Vac, 50/60 Hz, 1.1-0.9A; 200-240 Vac, 50/60 Hz, 0.6-0.5ASame as predicateY
    Power Consumption/Save Mode (Predicate)108W / less than 0.7W48.9W / less than 1.5W48.9W / less than 1.5WLower/comparable consumption is generally favorableY (Improved efficiency)
    Power ManagementDVI DMPMDVI DMPMDVI DMPMSame as predicateY
    USB Ports/standard1 upstream (endpoint), 2 downstream1 upstream (endpoint), 2 downstream1 upstream (endpoint), 2 downstreamSame as predicateY
    Indications for useDisplaying & viewing digital images, including digital mammography, for review & analysis by trained medical practitionersDisplaying & viewing digital images, including digital mammography, for review & analysis by trained medical practitionersSame as predicateSame as predicateY
    Applicable StandardsIEC 60601-1, IEC 60601-1-2IEC 60601-1, IEC 60601-1-2Complies with both standardsCompliance with standardsY

    The key acceptance criterion seems to be "substantial equivalence" to the predicate device, demonstrated by matching or comparable technological characteristics and compliance with relevant safety and EMC standards.

    2. Sample size used for the test set and the data provenance

    The document specifies "Bench testing" was conducted on the JUSHA-M53. However, it does not provide details on the sample size (e.g., number of displays tested) nor the data provenance. It describes performance data as being obtained from the device itself.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable as the testing described is bench testing of a medical display, not a clinical study involving human interpretation of images and ground truth established by experts.

    4. Adjudication method for the test set

    This information is not applicable for the same reason as point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The device is a medical display, not an AI-powered diagnostic or assistive tool. No MRMC study was mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable. The device is a medical display, not an algorithm.

    7. The type of ground truth used

    The document describes "Bench testing" for physical and electrical performance (luminance response, uniformity, contrast ratio, temporal response, stability, electrical safety, EMC). The "ground truth" for these tests would be the measurement standards and technical specifications (e.g., TG18 guideline for display characteristics, IEC 60601 series for safety/EMC). There is no "pathology" or "outcomes data" ground truth involved, as it is a display device, not an image interpretation algorithm or a treatment.

    8. The sample size for the training set

    This information is not applicable. The device is a medical display, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1