510(k) Data Aggregation

    K Number
    K150801

    Validate with FDA (Live)

    Date Cleared
    2015-05-22

    (57 days)

    Product Code
    Regulation Number
    882.4370
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stryker® Consolidated Operating Room Equipment (CORE™) Maestro Air Pneumatically operated surgical instrument system. The pneumatic motor provides power to operate removable rotating surgical cutting. tools and their accessories intended for use in neurosurgery, including craniotomy and spinal surgery; as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial and sternotomy surgeries.

    Device Description

    The Stryker® Maestro Air™ Motor(s) is a pneumatic motor powered by a regulated gas source. When connected to a gas source via tubing, the pneumatic motor operates at a normal operating pressure up to 150-psi (per square inch). The motor speed is controlled by a handswitch or a footpedal that connects to the motor and to a regulated gas source.

    AI/ML Overview

    This 510(k) summary describes a medical device, the Stryker® Maestro™ Air Motors, which is a pneumatic cranial drill motor. It does not involve any AI/ML components or describe a study related to AI/ML device performance. Therefore, the specific questions regarding AI/ML acceptance criteria, sample sizes for test and training sets, expert ground truth, adjudication methods, MRMC studies, or standalone algorithm performance cannot be answered from the provided text.

    The document primarily focuses on demonstrating substantial equivalence to a predicate device (Stryker Maestro Pneumatic System, K041754) based on non-clinical performance testing.

    Here's an analysis of the provided text, addressing what can be answered:

    1. Table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" with numerical targets for each performance metric in a pass/fail sense. Instead, it states that "The results of the performance testing demonstrate that the functionality, integrity and safety and effectiveness of the Stryker Maestro Air Motors are sufficient for their intended use and support a determination of substantial equivalence to the predicate device."

    However, we can infer performance comparisons from Table 6-1, which compares the subject device to the predicate device. These comparisons implicitly serve as "performance criteria" for demonstrating substantial equivalence.

    DESCRIPTIONPredicted Acceptance Criteria (Based on Predicate)Reported Device Performance (Stryker® Maestro Air™ Motors)
    Classification of DeviceClass IIClass II
    Regulation21 CFR 882.4370; Pneumatic Cranial Drill Motor21 CFR 882.4370; Pneumatic Cranial Drill Motor
    Product CodeHBBHBB
    Power sourceDry, Filtered, Compressed air or nitrogenDry, Filtered, Compressed air or nitrogen
    Patient PopulationGeneralGeneral
    ContraindicationsNone knownNone known
    Motor TypeHigh Speed, High Torque, Pneumatic VaneHigh Speed, High Torque, Pneumatic Vane
    Mode of ActionFootpedal and HandswitchFootpedal and Handswitch
    Operating SpeedUp to 75,000 rotations per minuteUp to 75,000 rotations per minute
    Stall Torque5.6 in-oz. @ 150 psi (per square inch)6.7 in-oz. @ 150 psi (per square inch)
    Maximum Operating Pressure120 psi (per square inch)Up to 150 psi (per square inch)
    Grip DesignSmoothKnurled (Different, but likely considered acceptable as a design feature, not functional)
    Router Retention MechanismFrictionSpring Collar (Different, but likely considered acceptable as a design feature)
    Method of SterilizationMoist Heat (Steam)Moist Heat (Steam)
    Sterility Assurance Level (SAL)10-e (likely 10^-6)10-e (likely 10^-6)
    Method of PackagingPackaged in a polybag case configurationPackaged in a sealed air Korrvu retention Insert configuration (Different packaging method)
    Cleaning MethodManualManual and Mechanical (automated) (Expanded cleaning method)
    Weight of the Motor and Hose2.05 lb.1.65 lb. (Lighter)
    Housing MaterialStainless SteelStainless Steel and Aluminum (Different materials)
    Pneumatic Hose AssemblyExhaust Hose: Flurosilicone Inlet Hose: Nitrile rubberExhaust Hose: Flurosilicone with silicone jacket and Slick Sil coating Inlet Hose: Nitrile rubber and neoprene (Different hose materials/coatings)

    Study that proves the device meets the acceptance criteria:

    The study performed was a set of non-clinical verification tests. These tests included:

    • Reliability Testing (Motor body, Handswitch mounting, Router Retention, Pneumatic Hose assembly and automated washing)
    • Packaging Testing
    • Cleaning Testing
    • Sterilization Testing

    The results of these tests demonstrated that the functionality, integrity, and safety and effectiveness of the Stryker Maestro Air Motors are sufficient for their intended use and support a determination of substantial equivalence.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify exact sample sizes for the non-clinical tests (e.g., how many motors were tested for reliability). It only lists the types of tests performed. It also does not provide information on data provenance in terms of country of origin or retrospective/prospective nature, as these are non-clinical engineering tests, not patient studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is a submission for a mechanical device based on non-clinical engineering tests, not an AI/ML device requiring expert-established ground truth from medical images or clinical data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as ground truth establishment by experts for a test set is not relevant to this type of device submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device and no MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. For this mechanical device, the "ground truth" for performance is established through engineering specifications, validated test methods (e.g., measuring torque, speed, pressure resistance), and established industry standards for sterilization, packaging, and cleaning.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML device.