Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K150776

    Validate with FDA (Live)

    Date Cleared
    2015-07-17

    (115 days)

    Product Code
    Regulation Number
    874.4760
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medrobotics Flex® System is a device that is intended for robot-assisted visualization and surgical site access to the oropharynx, hypopharynx, and larynx in adults (≥ 22 years of age). The Flex System also provides accessory channels for compatible flexible instruments used in surgery.

    Device Description

    The Flex System is an operator-controlled flexible endoscope that provides the benefits of both a rigid endoscope and a computer assisted controller. The Flex System allows for the endoscope to be introduced via an operator-controlled user interface, providing visualization and surgical site access to structures within the oropharynx, and larynx. Like some other video endoscopes, visualization is provided by a digital camera incorporated in the distal end of the endoscope. The Flex System's endoscope also provides two accessory channels for use of varied flexible instruments.

    AI/ML Overview

    The Medrobotics Flex System acceptance criteria and performance are as follows:

    1. Table of Acceptance Criteria & Reported Device Performance:

    Acceptance Criteria (Study Type)Target PerformanceReported Device Performance
    Clinical Study: Five-Point Visualization & Access Procedure
    Visualization Achieved - Palatine tonsil area>90%100% (45/45 subjects)
    Access Gained - Palatine tonsil area>90%98% (44/45 subjects)
    Visualization Achieved - Base of tongue area>90%100% (45/45 subjects)
    Access Gained - Base of tongue area>90%98% (44/45 subjects)
    Visualization Achieved - Epiglottis>90%100% (45/45 subjects)
    Access Gained - Epiglottis>90%98% (44/45 subjects)
    Visualization Achieved - Posterior pharyngeal wall>90%100% (45/45 subjects)
    Access Gained - Posterior pharyngeal wall>90%98% (44/45 subjects)
    Visualization Achieved - False vocal cords>90%93% (42/45 subjects)
    Access Gained - False vocal cords>90%91% (41/45 subjects)
    Clinical Study: Surgical Procedures
    Surgical procedures successfully completed with surgical site access and visualization provided by Flex SystemNot explicitly defined as a numerical percentage, but generally implied to be high for effectiveness.89% of 46 surgical procedures were successfully completed.
    Bench TestingMet performance specificationsMet performance specifications (Various tests listed)
    Transportation TestingWithstand anticipated shipping conditionsWithstood anticipated shipping conditions
    Usability/Human Factors TestingMeets intended user requirements and facilitates safe and effective user interactions with little chance of committing dangerous user errorsMeets intended user requirements and facilitates safe and effective user interactions with little chance of committing dangerous user errors
    Animal Testing (Abrasion and blunt force trauma)No increased level of abrasion and/or blunt force trauma compared to intubation controlNo increased level of abrasion and/or blunt force trauma compared to intubation control
    Electrical SafetyCompliance with UL 60601-1, CSA C22.2#601.1, ANSI/AAMI ES60601-1:2005, CAN/CSA-C22.2 No. 60601-1, IEC 60601-1-6, IEC 62366, IEC 60601-1-4Compliant
    Electromagnetic Compatibility TestingCompliance with EN 60601-1-2:2007/AC:2010, IEC 60601-1-2 Ed. 3.0Compliant
    Biocompatibility (Flex Scope patient-contacting materials)Pass Cytotoxicity, ISO Systemic Toxicity in Mice, ISO Intracutaneous Study, ISO Maximization Sensitization Study, ISO Oral Mucosal Irritation TestPassed all studies
    Sterilization (Flex Scope)Sterility assurance level (SAL) of 10-6Validated to SAL of 10-6
    Packaging and Shelf Life (Flex Scope)Packaging system stable over labeled shelf lifeValidated
    Cleaning and Sterilization of Reusable System Components (Flex Instrument Support)Compliance with ANSI/AAMI ST79, AAMI TIR12, AAMI TIR30, ANSI/AAMI/ISO 17665-1Validated
    Cleaning and Disinfection (Flex Base, Flex Cart, Stand, Flex Console, Monitor)Compliance with AAMI TIR12, AAMI TIR30Validated
    SoftwareVerified and Validated according to FDA guidance for "moderate level of concern" softwareVerified and Validated

    2. Sample size used for the test set and the data provenance:

    • Clinical Study: 45 subjects.
    • Data Provenance: European post-market clinical follow-up study (prospective, originating from Europe).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document does not explicitly state the number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience") used to establish the ground truth for the clinical study. It mentions that "investigator visualized and accessed" the anatomical locations, implying the assessment was made by the medical professionals conducting the study.

    4. Adjudication method for the test set:

    • The document does not specify an adjudication method like 2+1 or 3+1 for the clinical study results. The reported percentages (e.g., "45 (100%)" or "44 (98%)") suggest direct observation and recording by the study investigators.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was done to compare human readers with and without AI assistance described in this document. The Flex System is a robot-assisted visualization and surgical access device, not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • The Flex System is an "operator-controlled flexible endoscope" and "robot-assisted visualization and surgical site access" device, explicitly designed for human-in-the-loop operation. Therefore, a standalone algorithm-only performance study would not be applicable or conducted for this type of device.

    7. The type of ground truth used:

    • Clinical Study: The ground truth was based on direct observation and assessment by the study investigators (medical professionals) during the "Five-Point Visualization & Access Procedure" and "Surgical Procedures." This is best described as expert consensus/direct observation by medical professionals.
    • Bench, Animal, Electrical Safety, EMC, Biocompatibility, Sterilization, Packaging, Cleaning, Software Testing: Ground truth established through compliance with recognized standards, specifications, and validated test methodologies.

    8. The sample size for the training set:

    • The document does not provide details about a "training set" in the context of machine learning, as the device primarily involves robotic assistance and control rather than a learning algorithm. The software was verified and validated, but no specific training set size is mentioned.

    9. How the ground truth for the training set was established:

    • This question is not applicable as there is no mention or indication of a machine learning "training set" or corresponding ground truth establishment process for the Flex System in the provided text. The software validation refers to standard engineering verification and validation processes against requirements and specifications.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1