510(k) Data Aggregation

    K Number
    K150573

    Validate with FDA (Live)

    Date Cleared
    2015-03-25

    (19 days)

    Product Code
    Regulation Number
    884.4530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ebb Complete Tamponade System is indicated for use in providing temporary control or reduction of postpartum uterine bleeding. Inflation of the Vaginal Balloon anchors the Uterine Balloon and provides vaginal tamponade if vaginal bleeding is present. The ebb Complete Tamponade System should only be used in the setting of post-partum uterine bleeding when conservative management is warranted.

    Device Description

    Both devices are disposable, multiple lumen catheters attached to an inflatable balloon system designed to provide tamponade for controlling hemorrhage from the uterus and vagina. The devices consist of two inflatable balloons: The upper uterine balloon is inflated inside the uterus and the lower vaginal balloon is inflated inside the vagina. Inflation is accomplished with isotonic intravenous fluid such as normal saline or Ringers Lactate. The uterine balloon catheter has separate lumens to enable inflation/deflation, irrigation and drainage. The vaginal balloon catheter has a lumen to enable inflation. The uterine and vaginal balloons are permanently assembled and are not to be separated. The devices may be retained in position for up to 24 hours in the post-operative mode of treatment, and are supplied sterile in peel open pouches for one time use to a single patient. There are no differences in the material, chemical composition or energy source of the subject and predicate devices. The change in design of the subject device is limited to the modification to the bond of the distal end of the uterine balloon from a heat bond to a UV bond, to increase the bond strength.

    AI/ML Overview

    This document, a 510(k) summary for the ebb Complete Tamponade System, describes modifications made to a previously cleared device (K091958 Belfort-Dildy Obstetric Tamponade System). The key modification is changing the bond at the distal end of the uterine balloon from a heat bond to a UV bond to minimize potential leaks and increase bond strength.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Implicit)Reported Device Performance
    No leaks at the distal end of the uterine balloon under physiological conditions."None of the modified devices failed under those conditions [physiologic pressures under worst-case conditions]."
    Increased bond strength at the distal end of the uterine balloon."Testing performed on the modified devices demonstrated significant improvement in strength over the previous version."
    Withstand 7.5 times the pressure indicated in published literature without failure."...we subjected the modified devices to 7.5 times the pressure indicated in the published literature, and no failures were encountered."
    Maintain substantial equivalence to the predicate device in terms of safety and effectiveness."Performance data demonstrates that the ebb Complete Tamponade System is as safe and effective as the Belfort-Dilty Obstetric Tamponade System."

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not explicitly stated. The document mentions "none of the modified devices failed" and "we subjected the modified devices," implying a test set was used, but the exact number of devices tested is not provided.
    • Data Provenance: Not explicitly stated, but based on the context of a 510(k) submission, this would be prospective testing conducted by the manufacturer, Clinical Innovations, LLC, to support the device modification. There's no indication of clinical data or patient data being used for this specific performance testing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

    • Not Applicable. This study is focused on the mechanical performance of the device (bond strength, leakage under pressure) rather than diagnostic accuracy or clinical outcomes that would require expert-established ground truth. The "ground truth" here is the physical failure or non-failure of the device under specific test conditions.

    4. Adjudication Method for the Test Set:

    • Not Applicable. As this is a mechanical performance study, there's no "adjudication method" in the typical sense of expert review for medical imaging or clinical diagnosis. The outcome (failure/non-failure) is objectively measurable based on the test protocol.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No. This is a mechanical performance study of a physical medical device (uterine tamponade system), not a diagnostic algorithm or imaging system that would typically involve an MRMC study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This device is a physical medical device, not an algorithm. Therefore, the concept of "standalone algorithm performance" does not apply. The performance described is the standalone function of the device itself.

    7. The Type of Ground Truth Used:

    • Engineering/Mechanical Performance Data: The "ground truth" for this study is derived from controlled laboratory testing measuring physical properties such as bond strength and resistance to leakage under pressure. This is objective experimental data rather than expert consensus, pathology, or outcomes data typically associated with diagnostic device ground truth.

    8. The Sample Size for the Training Set:

    • Not Applicable. This is not an AI/machine learning study that would involve a "training set." The testing performed is to verify a physical modification to a medical device.

    9. How the Ground Truth for the Training Set Was Established:

    • Not Applicable. There is no training set involved in this mechanical performance study.