510(k) Data Aggregation

    K Number
    K150523

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-10-29

    (241 days)

    Product Code
    Regulation Number
    888.1100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine for kyphoplasty (for use with CareFusion Radiopaque Bone Cement).

    Device Description

    The Inflatable Bone Tamp (IBT) was designed for use in balloon kyphoplasty. The balloon serves to create a cavity in the vertebral body, thereby reducing the fracture and preventing cement leakage, while still allowing for cement interdigitation. The balloon catheter is the functional part of the device that creates a cavity and reduces the fracture. The balloon catheter provides a conduit through which the physician can inflate the balloon at the distal end of the catheter.

    AI/ML Overview

    The provided document is a 510(k) summary for the CareFusion AVAmax Vertebral Balloon. It describes the device, its intended use, comparison to a predicate device, and performance data from non-clinical tests.

    Here's an analysis of the document to address your request:

    1. A table of acceptance criteria and the reported device performance

    The document provides "Performance Test Summary - New Device" which lists the characteristic and results summary. It doesn't explicitly state "acceptance criteria" with numerical targets, but rather describes the performance in relation to requirements. I will interpret "acceptance criteria" as the implied requirements based on the "Results Summary."

    CharacteristicAcceptance Criteria (Implied)Reported Device Performance
    Inflation pressureExceed minimum burst pressure in a constrained environment"The balloon catheters exceeded the requirements for the minimum burst pressure in a constrained environment"
    Inflation volumeExceed minimum burst volume in an unconstrained environment"The balloon catheters exceeded the requirements for the minimum burst volume in an unconstrained environment"
    Balloon double wall thicknessSubstantially equivalent to predicate devices"The double wall thickness of the balloons was substantially equivalent to that of the predicate devices"
    Simulated UseSubstantially equivalent performance"Cadaveric simulated use testing demonstrated substantial equivalence in performance."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily describes non-clinical performance and does not detail specific sample sizes for tests beyond implying multiple "balloon catheters" were tested for inflation pressure and volume. For the "Simulated Use" test, it mentions "Cadaveric simulated use testing," but does not specify the number of cadavers.

    • Test Set Sample Size: Not explicitly stated, beyond implying multiple "balloon catheters" and "cadaveric" in the simulated use.
    • Data Provenance: Not specified, but given it's a 510(k) by CareFusion (a US-based company), it's likely conducted in the US. The testing is prospective in nature as it's part of premarket notification for a new/modified device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. The tests described are primarily engineering/performance tests (inflation pressure, volume, thickness) and a simulated use on cadavers. These types of tests do not typically involve human "experts" establishing a diagnostic ground truth in the way a clinical study would (e.g., radiologists reviewing images). The assessment of performance would be based on objective measurements and observations by the test engineers/personnel.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This document describes non-clinical performance testing, not a study involving human readers or subjective assessments that would require an adjudication method.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader, multi-case comparative effectiveness study was not done. The document explicitly states "N/A - No clinical tests were conducted for this submission." This device is a medical device (vertebral balloon), not an AI diagnostic tool, so such a study is irrelevant here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical medical device (vertebral balloon), not an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical tests described:

    • Inflation Pressure & Volume: The "ground truth" is based on objective physical measurements against predefined engineering requirements or industry standards.
    • Balloon double wall thickness: Objective physical measurement.
    • Simulated Use (Cadaver): "Ground truth" would be the observed mechanical performance and ability to perform the intended function (create a cavity, reduce fracture, prevent cement leakage simulation) in the cadaveric model, assessed against the performance of the predicate device. This is a functional assessment rather than a diagnostic one.

    8. The sample size for the training set

    Not applicable. This is not an AI/algorithm-based device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.