Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K150472

    Validate with FDA (Live)

    Date Cleared
    2015-06-04

    (101 days)

    Product Code
    Regulation Number
    876.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NxStage System One is indicated for the treatment of acute and chronic renal failure, or fluid overload using hemofiltration, hemodialysis, and/or ultrafiltration, in an acute or chronic care facility.

    The System is also indicated for home hemodialysis, including home nocturnal hemodialysis.

    All treatments must be administered under physician's prescription, and must be performed by a trained and qualified person, considered to be competent in the use of this device by the prescribing physician.

    Device Description

    The NxStage System One is comprised of the NxStage Cycler, an electromechanical control unit; the NxStage Cartridge, a sterile, single-use extracorporeal blood and fluid management circuit (with or without a preattached high permeability filter) that mounts integrally within the NxStage Cycler. The combined system is designed to deliver hemofiltration, hemodialysis and/or ultrafiltration in an acute or chronic care facility. The NxStage System One is also indicated for home hemodialysis, including home nocturnal hemodialysis.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the NxStage System One, a high permeability hemodialysis system. It seeks to demonstrate substantial equivalence to a previously cleared predicate device (K141752).

    Based on the provided document, here's an analysis regarding acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document presents a comparison of technological characteristics between the proposed device and the predicate device in Table 1. However, this table does not directly state acceptance criteria in terms of performance metrics (e.g., minimum accuracy, maximum allowable deviation) against which the proposed device was tested. Instead, it states that the "Proposed Device (NxStage System One)" is "Same" for most parameters compared to the "Predicate Device (NxStage System One)".

    The closest the document comes to defining acceptance criteria is under "Effluent Volume Accuracy" for the predicate device, where specifications for different software versions are provided. However, it only states that the proposed device is "Same," implying it meets these same specifications without explicitly presenting the proposed device's performance against them.

    Therefore, a table of acceptance criteria and reported device performance cannot be fully constructed from the provided text as the explicit acceptance criteria and quantitative performance results for the proposed device are not detailed, beyond stating they were "met."

    Inference of Acceptance Criteria (from Predicate Device Specifications as implied by "Same"):

    ParameterAcceptance Criteria (from Predicate)Reported Device Performance (Implied)
    Flow Rates: Blood10-600 ml/minSame (Implied to meet 10-600 ml/min)
    Flow Rates: Prescription Fluid /Dialysate0-18000 ml/hrSame (Implied to meet 0-18000 ml/hr)
    Flow Rates: Ultrafiltration0-2400 ml/hrSame (Implied to meet 0-2400 ml/hr)
    Transmembrane PressureYes (Implies capability)Same (Implied to have capability)
    Monitoring: Venous Pressure0 to 400 mmHgSame (Implied to monitor 0 to 400 mmHg)
    Monitoring: Effluent Fluid Pressure0 to 500 mmHgSame (Implied to monitor 0 to 500 mmHg)
    Monitoring: Air DetectorReduction of detector signal lasting 6 ms minimum (Approximates a 60 microliter bubble at 400 mmHg venous pressure and 600 ml/min blood flow)Same (Implied to meet this specification)
    Monitoring: Blood Leak Detector15% reduction in detector signal lasting 20 seconds minimum. Signal reduction % based on a 0.35 ml/min leak rate of 32 Hct blood.Same (Implied to meet this specification)
    Effluent Volume Accuracy (Software versions 4.7 and below)Greater of 300 ml/12 hr or 3% of exchange volumeSame (Implied to meet this specification)
    Effluent Volume Accuracy (Software versions 4.8 and higher)Therapy Fluid Flow Rate (L/hr) > 3 ml/hr: +100 or +5% UF; ≤ 3 ml/hr: +25 ml/hr* (*Representing 95/90 tolerance interval established under controlled laboratory testing conditions)Same (Implied to meet this specification)
    IV Prescription FluidOff-line, sterile- physician prescribed, indicated for infusionSame (Implied to handle this type)
    DialysateAvailable as pre-packaged pre-mixed sterile fluids or via the PureFlow SLSame (Implied to be compatible with these types)
    Compatible Blood Tubing SetNxStage Standard CartridgeSame (Implied to be compatible)
    SoftwareSoftware version 4.10Same (Implied to use or be compatible with 4.10)

    2. Sample Size Used for the Test Set and Data Provenance

    The document states: "No Clinical testing was conducted." All testing mentioned is "Non-Clinical Test/Performance Testing - Bench."

    • Sample Size: Not explicitly stated for bench testing. The document refers to "Simulated treatments" but does not give a number of simulations or samples used.
    • Data Provenance: Not applicable as no clinical data was used. All testing appears to be in-house bench testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. Since no clinical testing was conducted and all testing was "Non-Clinical Test/Performance Testing - Bench," there was no need for expert ground truth establishment in a clinical context. The "ground truth" for bench testing would be established by the engineering specifications and validated test methods.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical testing with human subjects or expert review was performed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This is a 510(k) for a medical device (hemodialysis system), not an AI-powered diagnostic or interpretive device that would involve human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Not applicable in the context of an "algorithm only" performance that would typically compare to human clinicians. This device is a treatment system. The "Non-Clinical Test/Performance Testing - Bench" can be considered a form of standalone performance assessment against engineering specifications, but it's not "algorithm only" in the sense of a software-as-a-medical-device (SaMD) context. It's the performance of the integrated electromechanical control unit with its software and cartridge.

    7. The Type of Ground Truth Used

    For the non-clinical bench testing, the "ground truth" would be established by:

    • Engineering specifications and design requirements: The device's performance was measured against these predefined technical standards and parameters (e.g., flow rates, pressure monitoring ranges, accuracy for effluent volume).
    • Physical measurements and simulated conditions: The testing simulates treatment conditions and measures the device's output against expected physical and chemical outcomes.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/ML model that requires a "training set" in the conventional sense. The "training" for a traditional medical device involves its design, development, and iterative testing/refinement against engineering principles and specifications.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for an AI/ML model. The "ground truth" for the device's design and development (analogous to a training process for an AI) comes from established medical standards for hemodialysis, regulatory requirements, risk analysis, and engineering principles.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1