510(k) Data Aggregation

    K Number
    K150377

    Validate with FDA (Live)

    Date Cleared
    2015-07-29

    (166 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LCD Monitor is intended to provide 4K 2D color video displays of images from endoscopic / laparoscopic camera systems and other compatible medical imaging systems. The LCD Monitor is a wide-screen, high-definition, medical grade monitor for real-time use during minimally invasive surgical procedures and is suitable for use in hospital operating rooms, surgical centers, clinics, doctors' offices and similar medical environments.

    Device Description

    The Sony LMD-X310S. LMD-X310MD and LMD-X310NB are 31 inches LCD monitors, and LMD-X550S. LMD-X550MD and LMD-X550NB are 55 inches LCD monitors with an advanced display technology, designed for use in medical environments. These monitors provide 4K resolution 2D color video displays of images from a surgical endoscope, laparoscopic camera system, and other compatible medical imaging systems. They are suitable for environments such as operating rooms. surgical centers, clinics, doctor's offices, and similar locations.

    LMD-X310S、LMD-X310MD and LMD-X310NB are operated by DC power supply and supplied with a dedicated AC adaptor, AC-300MD. LMD-X310MD is SONY brand model, LMD-X310S is a model for a specified customer and LMD-X310NB is for unspecified customers.

    MD-X550S, LMD-X550MD and LMD-X550NB are operated by AC power supply. LMD-X550MD is SONY brand model, LMD-X550S is a model for a specified customer and LMD-X550NB is for unspecified customers.

    Key features include:

    • A Large and 4K Screen
    • Wide color qamut
    • Clear and High-contrast View
    • A.I.M.E.™ (Advanced Image Multiple Enhancer)
    • Narrow Bezel
    • Ergonomic Design
    • User-friendly Control Panel
    • A Variety of Display Format
    • Easy to Clean
    • Installation-friendly Cabling
    AI/ML Overview

    This document describes the regulatory submission for Sony LMD-X310S, LMD-X310MD, LMD-X310NB, LMD-X550S, LMD-X550MD, and LMD-X550NB LCD Monitors. The submission focuses on demonstrating substantial equivalence to a predicate device, the SONY LMD-2451MT LCD MONITOR.

    Here's an analysis of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not present a formal table of "acceptance criteria" with specific numerical values for performance metrics that the device had to meet. Instead, it states that the devices conformed to recognized standards and that bench testing "met pre-defined design and performance acceptance criteria."

    The performance characteristics are largely presented in a comparative table with the predicate device, focusing on technological characteristics rather than specific "acceptance criteria" against which performance was measured.

    Functional Performance Comparison (derived from Section 5.6):

    Feature/CharacteristicSubject Device Reported Performance (LMD-X310S/MD/NB and LMD-X550S/MD/NB)Predicate Device Reported Performance (SONY LMD-2451MT LCD MONITOR)
    Intended EnvironmentHospital operating rooms, surgical centers, clinics, doctors' offices and similar medical environmentsHospital operating rooms, surgical centers, clinics, doctors' offices and similar medical environments
    Intended UsersDoctors and assistantsDoctors and assistants
    CompatibilitySurgical endoscopic/laparoscopic camera systems and other compatible medical imaging systemsSurgical endoscopic/laparoscopic camera systems and other compatible medical imaging systems
    PowerAC 100-240 V/50-60 HzAC 100-240 V/50-60 Hz
    Display Dimension2D2D / 3D
    Input SignalsDVI-D, HDMI, 3G/HD/SD-SDIDVI-D, HD15, RGB/Component, Y/C, Composite SDI 4:2:2 (optional) HD/D1-SDI (optional) NTSC/PAL :Y/C, Composite (optional)
    Output SignalsDVI-D, 3G/HD/SD-SDIComposite output connector Y/C output connector RGB/component output connectors External synchronized output connector
    Display DeviceLCD panel (IPS)LCD panel (IPS)
    Backlight DeviceLEDCCFL
    Viewing AngleRight>89[deg] (CR>10) Left>89[deg] (CR>10) Up>89[deg] (CR>10) Down>89[deg] (CR>10)Right>89[deg] (CR>10) Left>89[deg] (CR>10) Up>89[deg] (CR>10) Down>89[deg] (CR>10)
    Active Screen Size (LMD-X310 models)697.958(H)×368.064(V) mm-
    Active Screen Size (LMD-X550 models)1209.573(H)×680.385(V) mm-
    Active Screen Size (Predicate)-518.4(H)×324.0(V) mm
    Resolution (LMD-X310 models)4096×2160 pixels-
    Resolution (LMD-X550 models)3840×2160 pixels-
    Resolution (Predicate)-1920×1200 pixels
    Maximum Luminance (LMD-X310 models)≥550[cd/m2]-
    Maximum Luminance (LMD-X550 models)≥330[cd/m2]-
    Maximum Luminance (Predicate)-≥200[cd/m2]
    Color GamutAvg. 135% of sRGBAvg. 102% of sRGB
    Picture EnhancementA.I.M.E. functionNA
    Refresh Rate (LMD-X310 models)50/60 [Hz]-
    Refresh Rate (LMD-X550 models)100/120 [Hz]-
    Refresh Rate (Predicate)-50/60 [Hz]
    Frame Rate50/60 [fps]50/60 [fps]
    Display FormatNormal, Multi display, Flip displayNormal, Multi display, Mirror image
    Performance StandardsANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 62304, ISO 14971IEC 60601-1, EN 60601-1-2, ISO 14971

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "bench testing as part of required performance verification and validation activities" (Section 5.7) and states that "Results show that the subject device has met pre-defined design and performance acceptance criteria." However, it does not specify any sample size for the test set or the provenance of any data. The testing appears to be functional and safety testing of the device itself, rather than testing on a dataset of medical images.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. This submission focuses on the safety and performance of a medical display monitor and its substantial equivalence to a predicate device, not on diagnostic accuracy based on expert interpretation of medical images. Therefore, no experts were used to establish ground truth for a test set of medical images, as such a test set is not described.

    4. Adjudication Method for the Test Set

    Not applicable. As no test set involving expert interpretation is described, there is no mention of an adjudication method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. The document explicitly states: "No clinical data was collected to support substantial equivalence between the subject and predicate devices" (Section 5.8). Therefore, no MRMC study was performed, and no effect size for human reader improvement with AI assistance is provided.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is an LCD monitor, a hardware device for displaying images, not an algorithm or AI system for image interpretation. Therefore, a standalone algorithm performance study is not relevant.

    7. The Type of Ground Truth Used

    The "ground truth" for the device's performance is implicitly defined by its ability to meet recognized performance and safety standards (e.g., ANSI/AAMI ES60601-1, IEC 60601-1-2) and pre-defined design specifications through functional bench testing. It's not related to medical image interpretations, pathology, or outcomes data, as this is a display device.

    8. The Sample Size for the Training Set

    Not applicable. As the device is a medical monitor and not an AI/ML algorithm, there is no training set involved.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. Since there is no training set, this question is not relevant.